Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Calorimeter proper calorimetry

The calorimetry lexicon also includes other frequently used designations of calorimeters. When the calorimeter proper contains a stirred liquid, the calorimeter is called stirred-liquid. When the calorimeter proper is a solid block (usually made of metal, such as copper), the calorimeter is said to be aneroid. For example, both instruments represented in figure 6.1 are stirred-liquid isoperibol calorimeters. The term scanning calorimeter is used to designate an instrument where the temperatures of the calorimeter proper and/or the jacket vary at a programmed rate. [Pg.84]

The energy change associated with the process under study induces an energy change of the calorimeter proper, which can be determined by monitoring a corresponding temperature change or heat flux. In some calorimeters the reaction occurs in a closed vessel whose volume does not vary in the course of the experiment. This happens, for example, in bomb combustion calorimetry, where the reaction takes place inside a pressure vessel called the bomb, and in... [Pg.84]

Electrical calibration has the advantage of being more flexible. It can afford s0 through equation 7.23 ifitisdone on the reference calorimeter proper. Flowever, it can also be performed on the initial or final state of the actual experiment leading to (e0 + ecl) or (e0 + ecf), respectively. Twenty or 30 years ago the electrical calibration required very expensive instrumentation that was not readily available except in very specialized places, such as the national standards laboratories. Although the very accurate electronic instrumentation that is available today at moderate prices may change the situation, most users of combustion calorimetry still prefer to calibrate their apparatus with benzoic acid. [Pg.95]

It seems that immersion calorimetry into liquid nitrogen or liquid argon allows to go one step further in the determination of the internal surface area of micropores. These experiments requested a specially designed calorimeter operating at 77 or 87 K, with the special feature that the brittle end broken to start the immersion is located out of the calorimeter proper and therefore has no effeet on the calorimetric measurement. [Pg.176]

The present theory of calorimetry is concerned mostly with the instruments whose principle of operation is assumed to involve the transfer of heat in the system. This is true for most of the existing calorimeters, whether those with a constant or those with a variable temperature of the shield. It includes calorimeters in which the flow of heat between the calorimeter proper and its surroundings is quite intense, and also those in which this flow of heat is very low. On the other hand, the present theory is concerned to only a minor degree with calorimeters whose principle of operation is based on the assumption that there is no heat transfer (adiabatic calorimeters) or that, by definition, the heat transfer process is stationary (the generated heat effect is compensated). [Pg.177]

Pulse calorimeters pass electrical current through an electrically conducting sample to force a temperature increase, which is measured along with the voltage drop across the sample. If the heat loss from the sample is known (or estimated by calibration), the energy input divided by the temperature increase determines the true heat capacity, if the temperature change is small. Pulse calorimetry eliminates many of the drawbacks of drop calorimetry. It is fast, reproducible, and, with proper calibration, accurate. However, its use is limited to conductive materials. [Pg.762]

The measured quantity can be weight loss (TGA), a mechanical quantity (TMA) or a comparison between the behavior of two specimens (DTA) which, when properly calibrated, yields thermodynamic quantities that compare favorably with older, more conventional calorimetric techniques. Since the measurement time and equipment costs for scanning calorimetry (DSC) are orders of magnitude lower, the DSC has essentially replaced the conventional adiabatic calorimeter and finds a place in nearly every modern analytical laboratory. [Pg.239]

The enthalpy as a function of time is readily available from, for example, drop calorimetry experiments or from adiabatic calorimeters with incremental temperature increases. Scanning calorimeters, however, furnish the heat capacity of the sample. In these cases, the phase transition shows as a peak and the enthalpy of transition is calculated by integration of the peak area. Traditionally, this is done after constructing a proper baseline under the peak between the start and the end of the peak. The definition of the start and the end of the peak and the shape of the baseline under the peak are somehow arbitrary, particularly when the phase transition is accompanied by a heat capacity change. The enthalpy change at the transition temperature trs can be calculated from the heat capacity curve by... [Pg.126]

After the insertion of the sample into the calorimeter, enough time must be given to the instrument to come to a stable state and thermal equilibrium before the measurement can be started. Proper measurement parameters must be chosen in the case of scanning calorimetry, the initial temperature and the scanning rate must be chosen so as to give the calorimeter enough time to come to steady-state conditions before the event to be investigated starts. [Pg.255]

More recently, in addition to the adsorption equilibria studies, calorimetry was employed in examining of the metal oxide aqueous interface. The results deal either with relatively simple situations, such as protonation and deprotonation reactions at the surface or with the heats of the adsorption of organic molecules onto surfaces. In both cases, one encounters two problems. The first one is due to the fact that, upon addition of a reactant, several reactions take place in the calorimeter. In order to evaluate enthalpy of a specific surface reaction, one has to be able to distinguish between different contributions to the heat and also to determine the extents of all reactions taking place in the system. The next problem is related to the electrostatic contribution to the enthalpy. Recently, an experimental design was developed which enabled the proper interpretation of the calorimetric data so that one... [Pg.116]

Differential Scanning Calorimetry. Differential scanning calorimetry is in some respects a modification of differential thermal analysis, and the schematic diagram of a differential scanning calorimeter (dsc) is similar to that of the dta device shown in Figure 12. The essential difference, however, is that the measured quantity here is the differential power supplied to the two wells, rather than the temperature difference. In other words, the dsc device maintains the same programmed temperature in each well and records the power required to achieve this. If a transition takes place in the sample, a characteristic excursion in the measured differential power is observed. The nature of these excursions can be related to the transitional properties of the sample. Furthermore, by proper calibration with a reference material of known thermal properties, the specific heat capacity of the sample may be obtained. [Pg.1172]


See other pages where Calorimeter proper calorimetry is mentioned: [Pg.1907]    [Pg.123]    [Pg.1907]    [Pg.1912]    [Pg.137]    [Pg.1912]    [Pg.289]    [Pg.423]    [Pg.827]    [Pg.227]    [Pg.125]    [Pg.551]    [Pg.48]   
See also in sourсe #XX -- [ Pg.161 ]




SEARCH



Calorimeter Calorimetry

Calorimeters

Proper

© 2024 chempedia.info