Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Calibration analysis, calculation

EXAFS Data Analysis. A key aspect of the analysis outlined above is knowledge of the correct (k) and for a particular absorber-scatterer interaction. These parameters can either be calculated ab initio (6) or can be determined by measuring the EXAFS of structurally characterized model compounds (7). The ab initio method has the advantage that one need not prepare appropriate models for all possible unknowns. Unfortunately however, the ab initio parameters must be adjusted by a scaling factor and an assignment of E, (8). For this reason, one typically calibrates the calculated... [Pg.31]

The output of electronic integrators can be readily fed into a computer system. Here the raw area data can be compared to calibration data, calculations made, and complete analysis printed out. An alternate is to feed the detector output voltage directly into a computer using the proper interface. The capability of the computer can be used to integrate and calculate results. [Pg.175]

External standard calibration is used if the changes in the analytical system that may occur from the time the instrument has been calibrated to the time the sample analysis is completed are negligible. These changes are assumed to produce an insignificant error that is built into the daily calibration verification acceptance criteria. The response (calibration) factor of the external standard calibration is calculated according to Equation 1, Appendix 22, and it is measured in concentration or mass units (or their inverse). [Pg.249]

Table 6.3 Calibration and calculated data, obtained by GC-MS analysis, used for the quantification of cocaine in a drug sample... Table 6.3 Calibration and calculated data, obtained by GC-MS analysis, used for the quantification of cocaine in a drug sample...
Thus, by substitution in this equation for the peak areas from the chromatogram, the relative response factors, derived from the calibration analysis, and the concentration of the internal standard added to the sample, the concentration of the components in the sample can be calculated. Since this method involves ratios of peak areas rather than absolute values, it should be noted that the precision of analysis is not dependent on the injection of an accurately known amount of sample. However, the accuracy does depend on the accurate measurement of peak area. Assay and quantitation by the internal standard is often the preferred method as it takes account of variable compound response and removes potential errors due to variation in sample injection. The 80 and 200 mg% (mg per 100 ml) standard solutions are used to confirm the linearity of response over this concentration range. [Pg.461]

Torr. Extracted hydrogen is transferred by a mercury diffusion pump to an evacuated, calibrated analysis volume. After an initial pressure measurement, hydrogen is removed from the analysis volume through a heated palladium/silver osmosis tube, and the pressure of the residual gases is measured. The volume of hydrogen extracted from the sample is calculated from the differential pressure measurements. [Pg.120]

Figure 4. Calculation of prediction intervals in calibration analysis is the observed value of Y is the predicted value of X corresponding to y and are... Figure 4. Calculation of prediction intervals in calibration analysis is the observed value of Y is the predicted value of X corresponding to y and are...
In addition to laboratoiy glassware and equipment necessary fOT cleanup of the extract, traditional pesticide residue methods require expensive chromatogrsqihic instrumentation for identification and quantitation of residues. EIA methods require minimal amounts of glassware, disposable plasticware, or other supplies. Quantitative EIAs often make use of 96-well microtiter plates fOT multiple simultaneous assays of about a dozen extracts and associated reference standard. Major equipment consists of a plate reader, which automatically measures the absorbance of each well. Plate readers can be used alone or in conjunction with a personal computer, which can correlate replicate measurements, construct the calibration curve, calculate results, and provide a complete statistical analysis. Such an EIA workstation can be obtained for roughly half the cost of the GC or HPLC system typically used for pesticide residue analysis. [Pg.53]

Include all calibration plots calculate the correlation coefficient for both calibration plots. Determine the precision and accuracy for triplicate injections of the ICV for both calibrations. Report on the concentration in ppm for all unknown samples analyzed. This is an example in which two different methods are used to conduct analysis of the same unknown sample. Use the statistical evaluation for comparing two dependent averages (paired data) to determine whether the two different methods give the same result. You may want to include a representative chromatogram obtained from the gasoline-contaminated sample. Ask your lab instructor to assist you in obtaining a hardcopy from the laboratory printers. [Pg.517]

Include all calibration data, ICVs, and sample unknowns for both instrumental methods. Perform a statistical evaluation in a manner that is similar to previous experiments. Use EXCEL or LSQUARES (refer to Appendix C) or other computer programs to conduct a least squares regression analysis of the calibration data. Calculate the accuracy (expressed as a percent relative error for the ICV) and the precision (relative standard deviation for the ICV) from both instrumental methods. Calculate the percent recovery for the matrix spike and matrix spike duplicate. Report on the concentration of Cr in the unknown soil samples. Be aware of all dflution factors and concentrations as you perform calculations ... [Pg.527]

When the assumption of error-free x-values is not valid, either in method comparisons or, in a conventional calibration analysis, because the standards are unreliable (this problem sometimes arises with solid reference materials), an alternative comparison method is available. This technique is known as the functional relationship by maximum likelihood (FREML) method, and seeks to minimize and estimate both x- and y-dlrection errors. (The conventional least squares approach can be regarded as a special and simple case of FREML.) FREML involves an iterative numerical calculation, but a macro for Minitab now offers this facility (see Bibliography), and provides standard errors for the slope and intercept of the calculated line. The method is reversible (i.e. in a method comparison it does not matter which method is plotted on the x-axis and which on the y-axis), and can also be used in weighted regression calculations (see Section 5.10). [Pg.130]

In particular, the neutron activation analyst, to whom efficiency curves may be an irrelevance, is not weU served by most spectrum analysis packages. As far as activation analysis is concerned, there is much evidence to show that absolute analysis, calculating concentrations from first principles, is much less accurate than comparative analysis. Apart from aU of the problems which derive from having to use efficiency calibration curves, there are specific problems associated with defining and measuring neutron fluxes and cross-sections which make absolute analysis not worthwhile, in my opinion (although there are those who have devoted a considerable amount of effort into developing absolute neutron activation analysis procedures who would dispute that). For that reason, almost every activation analysis involves irradiation of samples and standards. A direct comparison between them is the simplest solution. [Pg.199]

The calibration and calculation operations in a quantitative TLC analysis are handled automatically by the software in modern computer-controlled densitometers. Chromatographic and measurement errors in scanning densitometry were discussed by Allwohn and Ebel (1989), Ebel and Glaser (1979), and Poliak (1989). [Pg.215]

The sequence in the quantitative evaluation of a chromatogram is raw data acquisition-integration-calibration and calculation of results-generating the analysis report. [Pg.142]

All proportional detectors are designed to give as near a linear response as possible to minimize calibration and calculation procedures when used for quantitative analysis. In many instances the output from the sensing device of the detector may not be linear, in fact, it is often logarithmic or exponential and thus, the associated electronics have to be designed to render the output linear. For example, a sensing device with an exponential output when used with a logarithmic amplifier would provide a linear response. [Pg.8]

Acetoxyquinoline-2-carboxylic acid (1.2 g, 5.2 mmol), benzotri-azole (0.62 g, 5.2 mmol), and DCC (1.07 g, 5.2 mmol) in dry THF (8 mL) were stirred at 0 C (2h), then 25 C (Ih). After filtration and concentration, the residual solid (1.79 g), PVA (MW = 25000 88% hydrolyzed 1.02 g, 20.8 mmol) and triethylamine (0.61 g) in DMSO (17 mL) were heated at 60 C for 48 hr. under nitrogen. The product was precipitated by pouring onto ice water, filtered, and washed with acetonitrile. Yield 1.35 g. Unchanged QA and other low molecular weight species were removed by precipitation from DMF with acetonitrile, then dialysis in 50% aq. ethanol using an Amicon YMIO membrane (10,000 cutoff). The QA incorporation was determined to be 5 mole-% by comparison of the UV absorbance at 258 nm with a calibration curve derived from standard solutions of QA in 50% aq. ethanol this result was verified qualitatively by H-NMR spectroscopy. A satisfactory elemental analysis was not obtained. Analysis calculated for 5 mole-% QA in PVA C, 61.4 H, 6.39 N, 3.58. Found C, 55.27 H, 6.87 N, 1.63. Films of the polymer were solution cast from DMF. [Pg.86]

Determinate measurement errors can be minimized by calibration. A pipet can be calibrated, for example, by determining the mass of water that it delivers and using the density of water to calculate the actual volume delivered by the pipet. Although glassware and instrumentation can be calibrated, it is never safe to assume that the calibration will remain unchanged during an analysis. Many instruments, in particular, drift out of calibration over time. This complication can be minimized by frequent recalibration. [Pg.60]

A linear regression analysis should not be accepted without evaluating the validity of the model on which the calculations were based. Perhaps the simplest way to evaluate a regression analysis is to calculate and plot the residual error for each value of x. The residual error for a single calibration standard, r , is given as... [Pg.124]

Calcium, D. of - continued in limestone or dolomite, (fl) 813 in presence of barium, (ti) 333 with CDTA, (ti) 333 with lead by EDTA, (ti) 333 with magnesium by EDTA, 328 by EGTA, (ti) 331 by flame emission, (aa) 804 Calcium oxalate, thermal analysis 498 Calcon 318 Calculators 133 Calibration of apparatus, 87 of burettes, 88 of graduated flasks, 88 of pipettes, 88 of weights, 74... [Pg.858]

Recognizing the difficulty satisfying the requirements for successful CLS, you may wonder why anyone would ever use CLS. There are a number of applications where CLS is particularly appropriate. One of the best examples is the case where a library of quantitative spectra is available, and the application requires the analysis of one or more components that suffer little or no interference other than that caused by the components themselves. In such cases, we do not need to use equation [33] to calculate the pure component spectra if we already have them in a library. We can simply construct a K matrix containing the required library spectra and proceed directly to equation [34] to calculate the calibration matrix K., . [Pg.68]

Figure 4.31. Key statistical indicators for validation experiments. The individual data files are marked in the first panels with the numbers 1, 2, and 3, and are in the same sequence for all groups. The lin/lin respectively log/log evaluation formats are indicated by the letters a and b. Limits of detection/quantitation cannot be calculated for the log/log format. The slopes, in percent of the average, are very similar for all three laboratories. The precision of the slopes is given as 100 t CW b)/b in [%]. The residual standard deviation follows a similar pattern as does the precision of the slope b. The LOD conforms nicely with the evaluation as required by the FDA. The calibration-design sensitive LOQ puts an upper bound on the estimates. The XI5% analysis can be high, particularly if the intercept should be negative. Figure 4.31. Key statistical indicators for validation experiments. The individual data files are marked in the first panels with the numbers 1, 2, and 3, and are in the same sequence for all groups. The lin/lin respectively log/log evaluation formats are indicated by the letters a and b. Limits of detection/quantitation cannot be calculated for the log/log format. The slopes, in percent of the average, are very similar for all three laboratories. The precision of the slopes is given as 100 t CW b)/b in [%]. The residual standard deviation follows a similar pattern as does the precision of the slope b. The LOD conforms nicely with the evaluation as required by the FDA. The calibration-design sensitive LOQ puts an upper bound on the estimates. The XI5% analysis can be high, particularly if the intercept should be negative.
One point, which is often disregarded when nsing AFM, is that accurate cantilever stiffness calibration is essential, in order to calculate accurate pull-off forces from measured displacements. Althongh many researchers take values quoted by cantilever manufacturers, which are usually calculated from approximate dimensions, more accurate methods include direct measurement with known springs [31], thermal resonant frequency curve fitting [32], temporary addition of known masses [33], and finite element analysis [34]. [Pg.30]

The validity of a practical method of direct viscosity calculation from size exclusion chromatographic analysis is demonstrated. The method is convenient to use and is not limited by the availability of narrow MWD standards. It is possible to accurately measure polymer Mark-Houwink constants using the suggested broad standard SEC-[n] and SEC-MW calibration procedure. [Pg.106]

The aim of all the foregoing methods of factor analysis is to decompose a data-set into physically meaningful factors, for instance pure spectra from a HPLC-DAD data-set. After those factors have been obtained, quantitation should be possible by calculating the contribution of each factor in the rows of the data matrix. By ITTFA (see Section 34.2.6) for example, one estimates the elution profiles of each individual compound. However, for quantitation the peak areas have to be correlated to the concentration by a calibration step. This is particularly important when using a diode array detector because the response factors (absorptivity) may considerably vary with the compound considered. Some methods of factor analysis require the presence of a pure variable for each factor. In that case quantitation becomes straightforward and does not need a multivariate approach because full selectivity is available. [Pg.298]


See other pages where Calibration analysis, calculation is mentioned: [Pg.259]    [Pg.289]    [Pg.256]    [Pg.357]    [Pg.298]    [Pg.76]    [Pg.32]    [Pg.93]    [Pg.3611]    [Pg.294]    [Pg.131]    [Pg.104]    [Pg.72]    [Pg.342]    [Pg.13]    [Pg.484]    [Pg.47]    [Pg.294]    [Pg.286]    [Pg.8]    [Pg.24]    [Pg.86]   


SEARCH



Calibration analysis

© 2024 chempedia.info