Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Z value

A parameter closely related to the energy of activation is the Z value, the temperature dependence of the decimal reduction time (D). The Z value is the temperature increase required for a one-logjo reduction (90% decrease) in the D value. The Z value can be determined from a plot of logjo D versus temperature (Fig. 12.5). The temperature dependence of the decimal reduction time can be expressed in linear and nonlinear forms  [Pg.145]

It can be shown that the Z value is inversely related to the energy of activation (Ea)  [Pg.146]


Knowing what to look for with respect to isotopic clusters can aid in interpreting mass spectra How many peaks would you expect to see for the molecular ion in each of the following compounds At what m/z values would these peaks appear (Disregard the small peaks due to and )... [Pg.570]

Some classes of compounds are so prone to fragmentation that the molecular ion peak IS very weak The base peak m most unbranched alkanes for example is m/z 43 which IS followed by peaks of decreasing intensity at m/z values of 57 71 85 and so on These peaks correspond to cleavage of each possible carbon-carbon bond m the mol ecule This pattern is evident m the mass spectrum of decane depicted m Figure 13 42 The points of cleavage are indicated m the following diagram... [Pg.570]

First the peak for the molecular ion M for all compounds that contain only car bon hydrogen and oxygen has an m z value that is an even number The presence of a nitrogen atom m the molecule requires that the m z value for the molecular ion be odd An odd number of nitrogens corresponds to an odd value of the molecular weight an even number of nitrogens corresponds to an even molecular weight... [Pg.953]

Positive Z values indicate a downfield shift, and an arrow indicates the point of attachment of the substituent group to the double bond. [Pg.784]

Positive ions are obtained from a sample by placing it in contact with the filament, which can be done by directing a gas or vapor over the hot filament but usually the sample is placed directly onto a cold filament, which is then inserted into the instrument and heated. The positive ions are accelerated from the filament by a negative electrode and then passed into a mass analyzer, where their m/z values are measured (Figure 7.1). The use of a suppressor grid in the ion source assembly reduces background ion effects to a very low level. Many types of mass analyzer could be used, but since very high resolutions are normally not needed and the masses involved are quite low, the mass analyzer can be a simple quadrupole. [Pg.45]

The ion current resulting from collection of the mass-separated ions provides a measure of the numbers of ions at each m/z value (the ion abundances). Note that for this ionization method, all ions have only a single positive charge, z = 1, so that m/z = m, which means that masses are obtained directly from the measured m/z values. Thus, after the thermal ionization process, m/z values and abundances of ions are measured. The accurate measurement of relative ion abundances provides highly accurate isotope ratios. This aspect is developed more fully below. [Pg.46]

Thus the m/z value for such ions is [M -i- n-l]/n, if the mass of hydrogen is taken to be one. As a particular example, suppose M = 10,000. Under straightforward Cl conditions, [M + H]+ ions will give an m/z value of 10,001/1 = 10,001, a mass that is difficult to measure with any accuracy. In electrospray, the sample substance can be associated with, for example, 20 hydrogens. Now the ion has a mass-to-change ratio of [M -t 20-H] and therefore m/z = 10,020/20 = 501. This mass is easy to measure accurately with a wide range of instruments. [Pg.57]

A major advantage of this hydride approach lies in the separation of the remaining elements of the analyte solution from the element to be determined. Because the volatile hydrides are swept out of the analyte solution, the latter can be simply diverted to waste and not sent through the plasma flame Itself. Consequently potential interference from. sample-preparation constituents and by-products is reduced to very low levels. For example, a major interference for arsenic analysis arises from ions ArCE having m/z 75,77, which have the same integral m/z value as that of As+ ions themselves. Thus, any chlorides in the analyte solution (for example, from sea water) could produce serious interference in the accurate analysis of arsenic. The option of diverting the used analyte solution away from the plasma flame facilitates accurate, sensitive analysis of isotope concentrations. Inlet systems for generation of volatile hydrides can operate continuously or batchwise. [Pg.99]

For either the in-line or hybrid analyzers, the ions injected into the TOF section must all begin their flight down the TOF tube at the same instant if arrival times of ions at a detector are to be used to measure m/z values (see Chapter 26, TOF Ion Optics ). For the hybrid TOF instruments, the ion detector is usually a microchannel plate ion counter (see Chapter 30, Comparison of Multipoint Collectors (Detectors) of Ions Arrays and MicroChannel Plates ). [Pg.153]

In general terms, the main function of the magnetic/electric-sector section of the hybrid is to be able to resolve m/z values differing by only a few parts per million. Such accuracy allows highly accurate measurement of m/z values and therefore affords excellent elemental compositions of ions if these are molecular ions, the resulting compositions are in fact molecular formulae, which is the usual MS mode. Apart from accurate mass measurement, full mass spectra can also be obtained. The high-resolution separation of ions also allows ions having only small mass differences to be carefully selected for MS/MS studies. [Pg.157]

A further important property of the two instruments concerns the nature of any ion sources used with them. Magnetic-sector instruments work best with a continuous ion beam produced with an electron ionization or chemical ionization source. Sources that produce pulses of ions, such as with laser desorption or radioactive (Californium) sources, are not compatible with the need for a continuous beam. However, these pulsed sources are ideal for the TOF analyzer because, in such a system, ions of all m/z values must begin their flight to the ion detector at the same instant in... [Pg.157]

Alternatively, ions of any one selected m/z value can be chosen by holding the magnetic field steady at the correct strength required to pass only the desired ions any other ions are lost to the walls of the instrument. The selected ions pass through the gas cell and are detected in the singlepoint ion collector. If there is a pressure of a neutral gas such as argon or helium in the gas cell, then ion-molecule collisions occur, with decomposition of some of the selected incident ions. This is the MS/MS mode. However, without the orthogonal TOF section, since there is no further separation by m/z value, the new ions produced in the gas cell would not be separated into individual m/z values before they reached the detector. Before the MS/MS mode can be used, the instrument must be operated in its hybrid state, as discussed below. [Pg.159]

The hexapole cannot act as a mass filter by applying a DC field and is used only in its all-RF mode, in which it allows all ions in a beam to pass through, whatever their m/z values. In doing so, the ion beam is constrained, so it leaves the hexapole as a narrow beam. This constraint is important because the ion beam from the inlet system tends to spread due to mutual ion repulsion and collision with residual air and solvent molecules. By injecting this divergent beam into a hexapole unit, it can be refocused. At the same time, vacuum pumps reduce the background pressure to about 10 mbar (Figure 22.1). The pressure needed in the TOF analyzer is about 10 ... [Pg.164]

Schematic diagram of an orthogonal Q/TOF instrument. In this example, an ion beam is produced by electrospray ionization. The solution can be an effluent from a liquid chromatography column or simply a solution of an analyte. The sampling cone and the skimmer help to separate analyte ions from solvent, The RF hexapoles cannot separate ions according to m/z values and are instead used to help confine the ions into a narrow beam. The quadrupole can be made to operate in two modes. In one (wide band-pass mode), all of the ion beam passes through. In the other (narrow band-pass mode), only ions selected according to m/z value are allowed through. In narrow band-pass mode, the gas pressure in the middle hexapole is increased so that ions selected in the quadrupole are caused to fragment following collisions with gas molecules. In both modes, the TOF analyzer is used to produce the final mass spectrum. Schematic diagram of an orthogonal Q/TOF instrument. In this example, an ion beam is produced by electrospray ionization. The solution can be an effluent from a liquid chromatography column or simply a solution of an analyte. The sampling cone and the skimmer help to separate analyte ions from solvent, The RF hexapoles cannot separate ions according to m/z values and are instead used to help confine the ions into a narrow beam. The quadrupole can be made to operate in two modes. In one (wide band-pass mode), all of the ion beam passes through. In the other (narrow band-pass mode), only ions selected according to m/z value are allowed through. In narrow band-pass mode, the gas pressure in the middle hexapole is increased so that ions selected in the quadrupole are caused to fragment following collisions with gas molecules. In both modes, the TOF analyzer is used to produce the final mass spectrum.
For the sake of illustration, a TOF analyzer could be likened to a camera taking snapshots of the m/z values of an assembly (beam) of ions the faster the repetition rate at which the camera shutter is clicked, the greater is the number of mass spectra that can be taken in a very short time. For TOF analyzers, it is not uncommon to measure several thousand mass spectra in one second All such spectra can be added to each other digitally, a process that improves the signal-to-noise ratio in the final accumulated total. [Pg.171]


See other pages where Z value is mentioned: [Pg.1342]    [Pg.117]    [Pg.17]    [Pg.562]    [Pg.571]    [Pg.201]    [Pg.20]    [Pg.29]    [Pg.86]    [Pg.96]    [Pg.101]    [Pg.115]    [Pg.151]    [Pg.153]    [Pg.154]    [Pg.155]    [Pg.155]    [Pg.156]    [Pg.158]    [Pg.158]    [Pg.159]    [Pg.159]    [Pg.159]    [Pg.160]    [Pg.160]    [Pg.160]    [Pg.161]    [Pg.163]    [Pg.164]    [Pg.165]    [Pg.165]    [Pg.166]    [Pg.166]    [Pg.169]    [Pg.170]    [Pg.171]   
See also in sourсe #XX -- [ Pg.13 ]

See also in sourсe #XX -- [ Pg.287 ]

See also in sourсe #XX -- [ Pg.202 , Pg.294 ]




SEARCH



Kosower Z-value

M/z-value

Z value and

© 2024 chempedia.info