Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Two-electrode system

According to the definition, a passive technique is one for which no appHed signal is required to measure a response that is analytically usehil. Only the potential (the equiHbrium potential) corresponding to zero current is measured. Because no current flows, the auxiHary electrode is no longer needed. The two-electrode system, where the working electrode may or not be an ion-selective electrode, suffices. [Pg.55]

The electrochemical circuitry required for SECM is relatively straightforward. Since the interface is not generally externally polarized in SECM measurements of liquid-liquid interfaces, a simple two-electrode system suffices (Fig. 3). A potential is applied to the tip, with respect to a suitable reference electrode, to drive the process of interest at the tip and the corresponding current that flows is typically amplified by a current-to-voltage converter. [Pg.295]

The kinetic investigation requires, as already stated in Section 5.1, page 252, a three-electrode system in order to programme the magnitude of the potential of the working electrode, which is of interest, or to record its changes caused by flow of controlled current (the ultramicroelectrode is an exception where a two-electrode system is sufficient). [Pg.303]

Figure 2. Cycleability of the PPy/CNTs composite at different cell voltages in 1M H2SO4 Current load 1=2 mA Mass of each electrode 8 mg Two-electrode system. Figure 2. Cycleability of the PPy/CNTs composite at different cell voltages in 1M H2SO4 Current load 1=2 mA Mass of each electrode 8 mg Two-electrode system.
A stable and predictable potential difference between the mobile phase and the working electrode is therefore impossible with a two-electrode system a third electrode is needed to monitor the potential of the mobile phase. [Pg.13]

The counter electrode is the current carrying electrode and it must be inert and larger in dimension. Platinum wire or foil is the most common counter electrode. For work with micro- or ultramicroelectrode where the maximum current demand is of the order of few microamperes, the counter electrode is not necessary. At very low current, a two-electrode system with the reference electrode can function as the current-carrying electrode with very little change in the composition of the reference electrode. Many commercial glucose sensors and on-chip microcells have such electrode configuration. [Pg.668]

In the case of mineral-mineral interactions, a mineral with higher potential acts as a cathode, while a mineral with lower potential acts as an anode. For a multiple mineral/grinding media(steel)system. The galvanic interactions become more complex than the two-electrode systems. The galvanic reactions among multielectrode systems are also governed by the mixed potential principle as shown in an example of polarization curves involving pyrite, pyrrhotite and mild steel in Fig. 1.9 (Pozzo and Iwasaki, 1987). [Pg.16]

To determine an overpotential, however, it is necessary to alter the above two-electrode system by introducing an extra auxiliary electrode, which is termed the auxiliary or counter-electrode. Thus a three-electrode arrangement is set up as shown in Fig. 736. In such a setup, the counter electrode is connected to the test electrode via a polarizing circuit (e.g., a power source) through which a controllable current is made to pass and produce alterations in the potential of the test electrode. Between the nonpolarizable reference electrode and test electrode is connected an instrument that is capable of measuring the potential difference between these electrodes. [Pg.388]

The historically important Daniell cell consists of a copper vessel containing a saturated solution of copper sulphate which forms one of the two electrode systems, and a central amalgamated zinc rod immersed in a zinc sulphate/sulphuric acid solution, which forms the other (Fig. 2.3). Gross mixing of the two solutions is prevented by a membrane or porous pot. It is assumed here that both electrodes are provided with copper terminals. [Pg.23]

The two-electrode system is just fine under the zero-current condition, but this does not allow very much chemistry to be accomplished. Figure 6.2b represents the finite-current case, where A( )w is the same as in Figure 6.2a. If the... [Pg.167]

The counter electrode in the two-electrode system serves two functions. First, it completes the circuit, allowing charge to flow through the cell, second, it is assumed to maintain a constant interfacial potential difference regardless of the current. These two functions are mutually exclusive only under very restricted circumstances. Both needs are better served by two separate electrodes ... [Pg.168]

Before the introduction of potentiostats in the early 1960s, the study of electrode processes was done mainly with two-electrode systems in which the functions of the reference and the counter electrode were unified in one simple electrode. Such an electrode is a non-polarisable electrode with a relatively large surface to be sure that it can conduct a certain amount of current due to occurring electrochemical processes. [Pg.57]

Fig. 16.3. Cyclic voltammogram of 13.6 U PP2A from Upstate (blank) and 0.3 mM catechol monophosphate+13.6 U PP2A from Upstate (CMP+PP) at lOOrnVs-1 using a single-drop configuration on a horizontally supported screen-printed two-electrode system, with graphite as working and Ag/AgCl as reference/auxiliary electrode. Reprinted from Campas et al. [86], with permission from Elsevier. Fig. 16.3. Cyclic voltammogram of 13.6 U PP2A from Upstate (blank) and 0.3 mM catechol monophosphate+13.6 U PP2A from Upstate (CMP+PP) at lOOrnVs-1 using a single-drop configuration on a horizontally supported screen-printed two-electrode system, with graphite as working and Ag/AgCl as reference/auxiliary electrode. Reprinted from Campas et al. [86], with permission from Elsevier.
Positive DEP drives particles towards field maxima and negative DEP moves them to field minima. It is only possible to create field maxima at the surfaces of the electrodes, no maximum can exist elsewhere in an isotropic solution. Trapping or positioning of particles by positive DEP therefore requires the dielectrophoretic force to be balanced by some other force (e.g. sedimentation, buoyancy or flow) and to be subjected to feedback control. This was first done with a two electrode system by Jones and Kaler [33-37]. [Pg.92]

A replaceable C disk electrode (two-electrode system) was used for end-column amperometric detection of biogenic amines. The electrode was inserted through a guide tube and was 30 5 [tm away from the capillary exit. This configuration allowed easy replacement of the electrode, especially after biofouling. The electrode was situated at a central position to maximize coulombic... [Pg.216]

Figure 5.2 Circuits for the measurement of cell potentials (a) two-electrode system with current passing through both working and reference electrodes (b) threePSlectrode system with current passing through the working and counter electrodes but not the reference electrode. Figure 5.2 Circuits for the measurement of cell potentials (a) two-electrode system with current passing through both working and reference electrodes (b) threePSlectrode system with current passing through the working and counter electrodes but not the reference electrode.
In most electrochemical experiments our interested is concentrated on only one of the electrode reactions. Since all measurements must be on a complete cell involving two electrode systems, it is common practice to employ a reference electrode as the other half of the cell. The major requirements of a reference electrode are that it be easy to prepare and maintain, and that its potential be stable. The last requirement essentially means that the concentra-... [Pg.11]

FIGURE 8.26 Influence of the maximum voltage on the specific capacitance (per mass of one electrode) vs. the number of galvanostatic cycles of a symmetric two-electrode system based on a PPy/CNT composite (20wt% of CNT) in lmol L-1 H2S04. (From Khomenko, V., et al., Appl. Phys. A, 82, 567, 2006. With permission.)... [Pg.355]

The simplest version of the atomic interferometer consists of two electrodes with the slits for passing the beam, separated with the variable gap L. For Lamb shift measurement corresponding interferometer is made of two two-electrode systems with longitudinal electric fields, mixing 2S and 2P-states. The systems were separated with a field-free gap of variable length L. This implies, that it is possible to write an exact expression for the probability W(L)e1,e2 of the yield I2P of 2P-atoms from the double system and determine, by processing the experimental dependence I2p(L), the Lamb shift value S. [Pg.358]

From this equation it can been seen that as the ohmic resistance increases, the remaining voltage driving force available to increase the overpotentials of the anode or cathode is diminished. Consequently smaller total current, I, flows through the cell. This is shown in Fig. 2, where in the absence of a finite Ra there is a single galvanic couple potential. A similar argument can be developed for the case of a driven two-electrode system. [Pg.178]

See also - three-electrode system, - two-electrode system. [Pg.533]

Two-electrode system — A system in which the -> working electrode is polarized to a desired - potential vs. the -> reference electrode. As a result, both electrodes pass -> current. To preserve the constant potential of the reference electrode during the measurements, the reference electrode should be of large area and with high activities of the redox forms. This requirement can be... [Pg.686]

Undivided cell — Electrochemical cells where all electrodes (two or three) are placed in the same compartment. Undivided cells are typically used for analytical experiments at small or -> microelectrodes in aqueous solutions when a -> two-electrode system is applied, or in a -> three-electrode measurements in nonaqueous media with a (platinum) -> quasireference electrode. A requirement in the use of an undivided cell is that the reaction products produced at the counter electrode do not reach or perturb the behavior of the working electrode. [Pg.689]

The electrochemical cell is coupled to a three-electrode potentiostatted form of instrumentation. If a two-electrode (working and reference) system were to be used, the current would have to flow through the reference electrode, thus risking instability in the reference potential. Furthermore, in a two-electrode system, the IR drop could be substantial. In contrast, in the three-electrode potentiostatted system, the current is forced to flow through the counter electrode, thereby avoiding problems with the reference electrode. Additionally, much of the IR drop is compensated by the potentiostat circuitry (Macdonald, 1977), which drives the potential between the working and counter electrode to a value which compensates the majority of this potential loss. However, the use of a potentiostat does not remove all of the IRu drop, since uncompensated resistance remains due to solution resistance between the tip of the reference and working electrodes, and... [Pg.13]

Fig. 5 Electrical circuitry for polarography/voltammetry. (A) A simple two-electrode system. (B) Illustrates a modern three-electrode system incorporating a potentiostat circuit. Fig. 5 Electrical circuitry for polarography/voltammetry. (A) A simple two-electrode system. (B) Illustrates a modern three-electrode system incorporating a potentiostat circuit.

See other pages where Two-electrode system is mentioned: [Pg.129]    [Pg.17]    [Pg.109]    [Pg.120]    [Pg.387]    [Pg.388]    [Pg.840]    [Pg.142]    [Pg.494]    [Pg.499]    [Pg.150]    [Pg.298]    [Pg.511]    [Pg.535]    [Pg.577]    [Pg.578]    [Pg.686]    [Pg.1490]    [Pg.1518]   
See also in sourсe #XX -- [ Pg.16 ]

See also in sourсe #XX -- [ Pg.44 ]




SEARCH



Electrode systems

The Pseudo Two-Phase System Type-II Electrode

Two-Dimensional Systems Microdisc Electrodes

© 2024 chempedia.info