Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Signal suppression, measuring

In isotope dilution inductively coupled plasma-mass spectrometry (ID-ICP-MS) the spike, the unspiked and a spiked sample are measured by ICP-MS in order to determine the isotope ratio. Using this technique, more precise and accurate results can be obtained than by using a calibration graph or by standard addition. This is due to elimination of various systematic errors. Isotopes behave identically in most chemical and physical processes. Signal suppression and enhancement due to the matrix in ICP-MS affects both isotopes equally. The same holds for most long-term instrumental fluctuations and drift. Accuracy and precision obtained with ID-ICP-QMS are better than with other ICP-QMS calibration... [Pg.660]

Figure 34 Excerpts of two-dimensional HMBC spectra of cholesteryl acetate recorded on a Bruker Avancell 400 MHz spectrometer (A) with the standard HMBC pulse sequence (Figure 1), and (B) with the IMPACT-HMBC experiment depicted in Figure 30. The same contour levels are used for all spectra. In (A), F, ridges are still visible (indicated by a vertical arrow), while they are very efficiently suppressed in (B). The proposed sequence results in signals with no coupling structure, as a result of the incorporation of a constant-time period. The improved peak dispersion is shown for the correlation between C-3 and H-2 (expanded in the small boxes). Asterix and the dashed box indicate residual Vch signals. The measurement duration was 22 min for both experiments. Figure 34 Excerpts of two-dimensional HMBC spectra of cholesteryl acetate recorded on a Bruker Avancell 400 MHz spectrometer (A) with the standard HMBC pulse sequence (Figure 1), and (B) with the IMPACT-HMBC experiment depicted in Figure 30. The same contour levels are used for all spectra. In (A), F, ridges are still visible (indicated by a vertical arrow), while they are very efficiently suppressed in (B). The proposed sequence results in signals with no coupling structure, as a result of the incorporation of a constant-time period. The improved peak dispersion is shown for the correlation between C-3 and H-2 (expanded in the small boxes). Asterix and the dashed box indicate residual Vch signals. The measurement duration was 22 min for both experiments.
It is also important to note that matrix-related effects, either signal enhancement or more commonly signal suppression, can have a pronounced effect on quantitative measurements. Based on these observations, the use of isotope-labeled standards is helpful to achieve accurate analytical measurement data on the diastereoisomers. Several methods found in the open literature include use of both 13C-labeled and d18-labeled surrogates as recovery and/or instrument standards [118],... [Pg.56]

In PCL systems care must be taken to ensure that the signal received directly from the transmitter does not compete with and swamp that from the target. Typically this will be the case unless measures are taken to suppress the direct signal occurring in the indirect channel. We can formulate a simple expression for the amount of direct signal suppression required by calculating the ratio of the indirect received sig-... [Pg.8]

Since active transport mechanisms require energy, the incubation temperature during the assay plays a crucial role. At 4°C, the fluidity of the cell membrane is reduced, the metabolism of the cell is downregulated, and energy-dependent transport processes are suppressed. Consequently, the amount of cell-associated target system refers mainly to the cytoadhesive fraction. In contrast, incubation at 37°C increases the fluidity of the cell membrane and the metabolic activity to an optimum, so both cytoadhesion and cytoinvasion occur at the same time. Thus, the uptake rate can be calculated from the difference in signal intensity measured upon incubation at both respective temperatures. [Pg.648]

The study showed that at concentrations of 100 mg l-1 fluoride and 1.2 percent KI (as the prereducing agent), the formation of stibine from Sb(V) was suppressed, while TMeSbO showed no signal depression. On the other hand, the use of 100 mg 1 1 fluoride in the absence of KI led to complete signal suppression for Sb(V) with apparent no influence on the signal intensity of Sb(III) and TMeSbO. While the concentration of TMeSbO was measured directly, Sb(III) and Sb(V) concentrations were calculated on the basis of the three analyzing steps, giving LoDs of 1.1, 1.2, and 1.4 pg l-1, respectively... [Pg.468]

The circular-shaped domains can be attributed to the globular PBD particles. Hence, the much more irregular shaped and smaller islands can be identified as the cross-sections of the Sb2S3 particles. For reasons of comparison, the mean values of the signal magnitudes measured on each of the constituents are displayed in histograms where the offset suppressed in the images was taken into account. [Pg.146]

Some methods take advantage of a difference in a particular property between water and the molecule to be studied. In particular, a macromolecule usually has a shorter value for proton Tt than water and a much lower diffusion coefficient. One of the oldest methods for water signal suppression is WEFT (water elimination Fourier transform), in which an inversion recovery sequence is applied (see Fig. 2.12) with r chosen to be the time that the water signal goes through zero (Tj In 2), just as in the BIRD pulse sequence. Another method makes use of the technique described in Section 9.3 to measure diffusion coefficients. [Pg.241]

Studies in which the voltage applied to the external split cathodes was pulsed have illustrated the formation of the capacitive double-layer which is responsible for signal suppression 35). The maximum LEI signal could be recovered only if the 1 ps laser pulse was delayed a minimum of 1.5 ms after the initiation of the 4 ms high voltage pulse. Figure 9 shows the results of potential measurements made with a tungsten... [Pg.12]

The very low concentrations expected in the analysis of trace elements in offshore and coastal Antarctic sea water can be also detected thanks to the high detection power of spectroscopic techniques such as Electrothermal Atomic Absorption Spectrometry (ETA-AAS) and Inductively Coupled Plasma Atomic Emission Spectrometry (ICP-AES) or ICP-MS. However, the saline matrix which constitutes the ideal medium in which to perform electrochemical measurements poses severe problems to the direct analysis of sea water because of possible signal suppression and/or undesired matrix effects. [Pg.118]

In general, there is not an upper mass limit to the proteins that can be analyzed using the ESI-MS-based methodology. However, successful application of the ESI-MS-based methodology does require that the requisite histidine-containing peptides be successfully detected and identified in the ESI-MS readout. In some cases, signal suppression issues associated with the ESI-MS readout can preclude measurements of specific histidine residues in a protein. Accurate value determina-... [Pg.174]

Consequently, a value higher than 100% means that signal enhancement occurs in the mass spectrometer because of the coextracted and coeluted matrix. A value lower than 100% implies that signal snppression occnrs in the mass spectrometer. This matrix effect was measured twofold and calcnlated for ultrapure water, drinking water, and surface water at a concentration level of 0.2 mg/L for each compound. In order to investigate the effect of the matrix on the response of the mass spectrometer, the signal suppression was calculated for 15 deuterated standards in sample extracts of snrface water and wastewater compared to a standard in drinking water. The results were corrected for the blank values. [Pg.101]

The issue of matrix effects that result in signal suppression (or possibly, enhancement) is a cause for concern, in that analyses may be subject to bias. Tomy et al. [116] report one of the first applications of isotopically labeled internal standards to the determination of HBCD diastereomers in complex matrix samples. The use of and labeled HBCD isomers, in principle, should reduce measurement bias, since quantitation is based on the ratio of coeluting HBCD species (i.e., the analyte and the internal standard) rather than absolute responses of the analytes. Analyte response may vary with instrument conditions and matrix effects however, the ratio remains constant. The use of isotopically labeled internal standards for the individual diastereomers is now common practice for BFRs [117]. HBCD isomers have been determined by LC—MS/MS methods in diverse sample types, including estuary [94] lake, river, and snow water samples [118] styrenic polymers [119] marine sediments [120] fish [121], marine mammal tissues [122—124] seaweed [125] foods [126] and human biological matrices [124,127]. [Pg.358]

In order to overcome the limitations of extraction methods, Schilling et al. [199] used a direct, quantitative procedure that identifies type and quantity of each additive present in stabilised polyolefin samples. High-field (11.7 T), high-resolution NMR and selective signal suppression techniques can discriminate between additives with similar molecular structure and provide a quantitative measure of each compound. [Pg.647]


See other pages where Signal suppression, measuring is mentioned: [Pg.111]    [Pg.438]    [Pg.442]    [Pg.698]    [Pg.701]    [Pg.96]    [Pg.78]    [Pg.345]    [Pg.336]    [Pg.149]    [Pg.927]    [Pg.300]    [Pg.436]    [Pg.381]    [Pg.7]    [Pg.301]    [Pg.172]    [Pg.322]    [Pg.388]    [Pg.707]    [Pg.309]    [Pg.321]    [Pg.109]    [Pg.194]    [Pg.28]    [Pg.248]    [Pg.343]    [Pg.72]    [Pg.193]    [Pg.200]    [Pg.979]    [Pg.363]    [Pg.62]    [Pg.228]    [Pg.752]   
See also in sourсe #XX -- [ Pg.69 ]




SEARCH



Signal measurements

Signal suppression

Signal, measuring

© 2024 chempedia.info