Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Signals common

Evacuation signals commonly take the form of a siren or buzzer, which after a ten second or more exposure may be followed by a voice message. The alarm and message should be repeated until the area is evacuated. [Pg.128]

The effects of adsorption on an electrochemical response signal are varied, ranging from complete passivation of the electrode, to catalytic enhancement of a redox process, to the appearance of new electrochemical response signals. Commonly encountered situations are dealt with briefly below. The reader is referred to the works by Bauer [13] and Reilley and Stumm [14] for a more detailed discussion. [Pg.47]

It is very difficult to achieve H NMR spectra of anthocyanins without an intense water peak around 5 ppm, which may overlap with signals commonly representing anomeric protons. After some hours of storage in the acidified deuterated solvent, this peak tends to migrate upfield -0.4 ppm. It is therefore advised to record H NMR spectra immediately after preparation to reveal peaks which may be hidden under the water peak, and to repeat this procedure after several hours as well. [Pg.834]

Comparable infrared (IR) spectra of complex organic solids such as coals (Fig. 7) 23), cherts, kerogens, humic substances and some natural polymers have been presented in various publications. They show a limited number of rather broad bands which are due to well defined chemical groups and can often be interpreted by comparison to less complicated spectra. The signals commonly observed in fossil organic matter are as follows (Robin et al., 1977 21), Tissot Welte, 1978 24), Rouxhet et al., 1980 22>, Friedel Carlson, 1972 25)) ... [Pg.10]

Reverse mirage effect — A deflection signal commonly related to photothermal experiments in which the deflected probe light beam passes on the opposite side of the photoilluminated interface [i]. [Pg.583]

In pulsed NMR, a short rf pulse is applied to all spins at the Larmor frequency, creating a small B field along the v-axis in the rotating frame and causing the nuclear spins to rotate toward the xy-plane as a coherent packet (Fig. 3). Typically, the pulse duration is such that Mq flips into the y-axis where the decaying vy component of the magnetization can be detected as a time domain signal, commonly known as a free... [Pg.1908]

The overall process of phosphorylation and dephosphorylation is a cascade of reactions and consists of an extracellular signal, commonly referred to asfirst messenger (e.g., hormones, such as glucagon) a specific receptor on the cell membrane of the target cell a transducer coupled to the... [Pg.110]

Aspects of signaling common to all known mammalian chemokine receptors include induction of calcium flux and chemotaxis, and marked inhibition of both... [Pg.3]

A FIGURE 1-16 External signals commonly cause a change in the activity of preexisting proteins or in the amounts and types of proteins that cells produce, (a) Binding of a... [Pg.16]

There are several different types of signals commonly encountered. [Pg.614]

Lock or field/ A method for accurately maintaining frequency lock (locking) the static magnetic field at a fixed value. The frequency of a nuclear magnetic resonance signal - commonly deuterium from a deuterated solvent - is held at constant value by a field/fre-quency feedback loop to compensate for any magnetic field drift during an experiment. [Pg.3271]

In standards, it is recommended that the devices, systems, or protection layers be assessedfor independence, and potential for common cause failure. Common cause failure can cause multiple devices, systems, or layers to fail simultaneously. Another term is common mode failure (a subset of common cause failure), which describes the simultaneous failure of two devices in the same mode, for example, two redundant transmitters fail simultaneously and are disabled due to loss of signal (common mode failure) [10,11]. [Pg.79]

The scanning electron microscope (SEM) forms an image by scanning a probe across the specimen, and in the SEM the probe is a focused electron beam. The probe interacts with a thin surface layer of the specimen, a few micrometers thick at most. The detected signal commonly used to form the TV-type image is the number of low energy secondary electrons emitted from the sample surface. Scanning electron microscopy is fully described in several texts [22-26], and its use with polymers has been reviewed by White and Thomas [27]. [Pg.25]

The paragraphs above indicate levels of major spectral ion signals. Many of these cause slight or occasionally severe interferences with M/z ion signals commonly chosen for measurement in the determination of impurity levels. [Pg.387]

The anti reset-windup technique discussed above is known as external reset feedback. For most applications either it, or the modification mentioned below, is our preferred scheme. It has the disadvantage that the controller output signal, commonly labeled valve position, is really different from the actual position. It differs by the product of the error signal times the proportional gain. Lag in the reset circuit may cause further error. A modification therefore is introduced by some vendors, particularly in the newer microprocessor controls. This consists of setting the reset time equal to zero when the controller is overridden. This technique is sometimes called integral tracking. It should not be used with auto overrides. [Pg.201]


See other pages where Signals common is mentioned: [Pg.267]    [Pg.278]    [Pg.135]    [Pg.195]    [Pg.21]    [Pg.30]    [Pg.606]    [Pg.39]    [Pg.183]    [Pg.183]    [Pg.258]    [Pg.324]    [Pg.500]    [Pg.1242]    [Pg.1700]    [Pg.258]    [Pg.2]    [Pg.285]    [Pg.206]    [Pg.127]    [Pg.654]    [Pg.97]    [Pg.269]    [Pg.30]    [Pg.606]    [Pg.143]    [Pg.290]    [Pg.277]    [Pg.295]    [Pg.50]   
See also in sourсe #XX -- [ Pg.126 ]

See also in sourсe #XX -- [ Pg.126 ]




SEARCH



Common Spurious Signals Observed in Mass

Common Spurious Signals Observed in Mass Spectrometers

RESIDUAL SOLVENT SIGNALS OF COMMON NMR SOLVENTS

© 2024 chempedia.info