Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Pulse microwave detection

Fig. 9. ODMR investigations at T = 1.4 K of Pd(2-thpy)2 dissolved in an n-octane Shpol skii matrix. Concentration = 10 mol/1 cw excitation Ag c = 330 nm (30.3 x 10 cm 0- Detection of the emission at 18418 cm (Tj —> Sq transition), (a) Zero-field ODMR (optically detected magnetic resonance) spectrum (b) Zero-field microwave recovery ODMR signal after pulsed microwave excitation with a microwave frequency of 2886 MHz. The best fit of the recovery signal is obtained with Eq. (4). (Compare Ref. [61])... Fig. 9. ODMR investigations at T = 1.4 K of Pd(2-thpy)2 dissolved in an n-octane Shpol skii matrix. Concentration = 10 mol/1 cw excitation Ag c = 330 nm (30.3 x 10 cm 0- Detection of the emission at 18418 cm (Tj —> Sq transition), (a) Zero-field ODMR (optically detected magnetic resonance) spectrum (b) Zero-field microwave recovery ODMR signal after pulsed microwave excitation with a microwave frequency of 2886 MHz. The best fit of the recovery signal is obtained with Eq. (4). (Compare Ref. [61])...
One very recently published result must be mentioned. Petukhov et al. have reported the detection of the EPR spectrum of micron-sized crystals of Fe8 via their magnetization response as a function of apphed magnetic field, using a Hall-probe magnetometer under either continuous wave or pulsed microwave irradiation at 118 GHz and between 1.4 and 50 K [53]. Dips are observed in the magnetization vs. field curves corresponding to resonant absorption - that is, EPR transitions. This method offers potentially extraordinary sensitivity and, furthermore, manipulation of the magnetization data in the absence and presence of the microwave radiation allows determination of the spin temperature. [Pg.82]

The microwave detected MODR scheme closely resembles pulsed nuclear magnetic resonance (Hahn, 1950), optical coherent transients by Stark switching (Brewer and Shoemaker, 1971) and laser frequency switching (Brewer and Genack, 1976). The on-resonance microwave radiation field, ojq = ( 2 — Ei)/H, creates an oscillating bulk electric dipole polarization (off-diagonal element of the density matrix, pi2(t)). The oscillation is at u>o u>r, where ojr is the (Mj-dependent) Rabi frequency,... [Pg.435]

Microwave detection can be performed from the penetrating side (pulse-echo mode) or from the opposite side (through-transmission mode). The through-transmission technique can determine the slate of resin cure in boron fiber laminates. However, it is not suitable for laminates with 0 /60 filament orientation [87]. The propagation of micro-waves (wavelength 10 10 m) is governed by Maxwell s equations of the magnetic field [88]. Maxwell s equations are... [Pg.800]

A microwave pulse from a tunable oscillator is injected into the cavity by an anteima, and creates a coherent superposition of rotational states. In the absence of collisions, this superposition emits a free-mduction decay signal, which is detected with an anteima-coupled microwave mixer similar to those used in molecular astrophysics. The data are collected in the time domain and Fourier transfomied to yield the spectrum whose bandwidth is detemimed by the quality factor of the cavity. Hence, such instruments are called Fourier transfomi microwave (FTMW) spectrometers (or Flygare-Balle spectrometers, after the inventors). FTMW instruments are extraordinarily sensitive, and can be used to examine a wide range of stable molecules as well as highly transient or reactive species such as hydrogen-bonded or refractory clusters [29, 30]. [Pg.1244]

An alternative approach to obtaining microwave spectroscopy is Fourier transfonn microwave (FTMW) spectroscopy in a molecular beam [10], This may be considered as the microwave analogue of Fourier transfonn NMR spectroscopy. The molecular beam passes into a Fabry-Perot cavity, where it is subjected to a short microwave pulse (of a few milliseconds duration). This creates a macroscopic polarization of the molecules. After the microwave pulse, the time-domain signal due to coherent emission by the polarized molecules is detected and Fourier transfonned to obtain the microwave spectmm. [Pg.2441]

Time-resolved microwave conductivity measurements with electrodes in electrochemical cells can conveniently be made with pulsed lasers (e.g., an Nd-YAG laser) using either normal or frequency-doubled radiation. Instead of a lock-in amplifier, a transient recorder is used to detect the pulse-induced microwave reflection. While transient microwave experiments with semiconducting crystals or powders have been performed... [Pg.447]

HYSCORE, is a 2D four-pulse ESEEM technique which provides correlation between nuclear frequencies originating from different electron manifolds. The sequence of four microwave pulses is tx/2—x—tx/2—/tx— t2-nl2-x-echo where the echo amplitude is measured as a function of tx and t2 at fixed x. The a-proton anisotropic couplings can be detected by this technique (Konovalova et al. 2001a, Focsan et al. 2008). [Pg.168]

A major limitation of CW double resonance methods is the sensitivity of the intensities of the transitions to the relative rates of spin relaxation processes. For that reason the peak intensities often convey little quantitative information about the numbers of spins involved and, in extreme cases, may be undetectable. This limitation can be especially severe for liquid samples where several relaxation pathways may have about the same rates. The situation is somewhat better in solids, especially at low temperatures, where some pathways are effectively frozen out. Fortunately, fewer limitations occur when pulsed radio and microwave fields are employed. In that case one can better adapt the excitation and detection timing to the rates of relaxation that are intrinsic to the sample.50 There are now several versions of pulsed ENDOR and other double resonance methods. Some of these methods also make it possible to separate in the time domain overlapping transitions that have different relaxation behavior, thereby improving the resolution of the spectrum. [Pg.162]

The recent advent of the ability to apply short and very intense microwave pulses to samples and detect the fast response to the excitation has made it... [Pg.162]

Double-resonance spectroscopy involves the use of two different sources of radiation. In the context of EPR, these usually are a microwave and a radiowave or (less common) a microwave and another microwave. The two combinations were originally called ENDOR (electron nuclear double resonance) and ELDOR (electron electron double resonance), but the development of many variations on this theme has led to a wide spectrum of derived techniques and associated acronyms, such as ESEEM (electron spin echo envelope modulation), which is a pulsed variant of ENDOR, or DEER (double electron electron spin resonance), which is a pulsed variant of ELDOR. The basic principle involves the saturation (partially or wholly) of an EPR absorption and the subsequent transfer of spin energy to a different absorption by means of the second radiation, leading to the detection of the difference signal. The requirement of saturability implies operation at close to liquid helium, or even lower, temperatures, which, combined with long experimentation times, produces a... [Pg.226]

Radiations outside the ultraviolet, visible and infrared regions cannot be detected by conventional photoelectric devices. X-rays and y-rays are detected by gas ionization, solid-state ionization, or scintillation effects in crystals. Non-dispersive scintillation or solid-state detectors combine the functions of monochromator and detector by generating signals which are proportional in size to the energy of the incident radiation. These signals are converted into electrical pulses of directly proportional sizes and thence processed to produce a spectrum. For radiowaves and microwaves, the radiation is essentially monochromatic, and detection is by a radio receiver tuned to the source frequency or by a crystal detector. [Pg.283]

The two techniques, ENDOR and ESE envelope modulation, supplement each other. ESE envelope modulation seems to be more sensitive in detecting nuclear transitions at very low frequencies but is limited in the frequency range by yeB , where ye denotes the gyromagnetic ratio of the electron and Bj the microwave pulse amplitude. ENDOR, whose sensitivity increases with frequency, suffers on the other hand from the small transition probability at low frequencies. [Pg.47]

Electron-nuclear double resonance (ENDOR) spectroscopy A magnetic resonance spectroscopic technique for the determination of hyperfine interactions between electrons and nuclear spins. There are two principal techniques. In continuous-wave ENDOR the intensity of an electron paramagnetic resonance signal, partially saturated with microwave power, is measured as radio frequency is applied. In pulsed ENDOR the radio frequency is applied as pulses and the EPR signal is detected as a spin-echo. In each case an enhancement of the EPR signal is observed when the radiofrequency is in resonance with the coupled nuclei. [Pg.250]

The second approach is to use thermal beams of alkali atoms as shown in Fig. 10.2.4 A beam of alkali atoms passes into a microwave cavity where the atoms are excited by pulsed dye lasers to a Rydberg state. A1 /zs pulse of microwave power is then injected into the cavity. After the microwave pulse a high voltage pulse is applied to the septum, or plate, inside the cavity to analyze the final states after interaction with the microwaves. By adjusting the voltage pulse it is possible to detect separately atoms which have and have not been ionized or to analyze by selective field ionization the final states of atoms which have made transitions to other bound states. [Pg.163]


See other pages where Pulse microwave detection is mentioned: [Pg.596]    [Pg.145]    [Pg.115]    [Pg.130]    [Pg.224]    [Pg.186]    [Pg.1803]    [Pg.551]    [Pg.47]    [Pg.302]    [Pg.332]    [Pg.1548]    [Pg.1576]    [Pg.215]    [Pg.129]    [Pg.130]    [Pg.672]    [Pg.31]    [Pg.32]    [Pg.158]    [Pg.163]    [Pg.218]    [Pg.111]    [Pg.75]    [Pg.7]    [Pg.46]    [Pg.164]    [Pg.345]    [Pg.347]    [Pg.373]    [Pg.374]    [Pg.482]    [Pg.496]    [Pg.6]    [Pg.259]   
See also in sourсe #XX -- [ Pg.544 ]




SEARCH



Detection pulsed

Microwave pulses

© 2024 chempedia.info