Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Nephelometric Turbidity

The goal of filtration in the modem municipal treatment plant is a maximum of 0.1 ntu (nephelometric turbidity unit), which ensures a sparkling, clear water (8). Freedom from disease organisms is associated with freedom from turbidity, and complete freedom from taste and odor requites no less than such clarity. The National Interim Primary Drinking Water Regulations (NIPDWR) requite that the maximum contaminant level for turbidity at the point of entry into the distribution system be 1.0 ntu unless it can be shown that levels up to 5 ntu do not interfere with disinfection, interfere with the maintenance of a chlorine residual in the distribution system, nor interfere with bacteriological analyses. [Pg.276]

Water from cooling tower pump suction, pH 8.6-8.8, pressure 20-30 psi (140-210 kPa), flow 2-5 ft/s (0.2 to 1.5 m/s). Dispersant, 1-3 ppm tolyltriazole, sodium hypochlorite 2 hr/day to 0.8 free residual chlorine 0.6-0.8 ppm total zinc and 0.1-0.2 ppm soluble zinc. Free chlorine maintained at 1 ppm for 5 consecutive days/month during the summer. Chemical treatment started after 2 years of no treatment. Water conductivity -612 ( imhos/cm), turbidity 27 NTU (nephelometric turbidity units), chloride 110 ppm, sulfate 50 ppm, carbonate alkalinity (CaCOa) 27 ppm, bicarbonate alkalinity (CaCOs) 118 ppm... [Pg.65]

Turbidity, a measure of the colloidal haze present in Nephelometric Turbidity Units (NTU). [Pg.479]

Triple media (MM) filters use anthracite, sand, and garnet, with the garnet providing the final and finest filtering quality, so that water with a turbidity less than 5 nephelometric turbidity units (NTU) and 20p maximum particle size can be expected. Media SG, grain size, and bed depths are given in Table 9.1. [Pg.322]

High concentrations of particulate matter cause water to appear cloudy due to the absorption and scattering of light. The cloudiness of water is operationally measured as the amount of light transmitted through water. This quantity is termed turbidity and is reported in nephelometric turbidity units (NTU). [Pg.208]

Turbidity is caused by any solid material which is dissolved or suspended in a liquid. The intensity of light scattered by a sample is measured and compared with that measured for standard formazin suspensions, and expressed as nephelometric turbidity units (NTU). Colour is determined as the absorbance (measured spectrophotometrically at 400 nm) of the sample filtered through a 0.4S micron pore size membrane filter [12, 13]. [Pg.226]

Note TCU = true color unit NTU = nephelometric turbidity... [Pg.723]

Field meters use the nephelometric method for turbidity measurements and read turbidity in Nephelometric Turbidity Units (NTUs). A nephelometer measures the intensity of light scattered by a sample and compares it to the intensity light scattered by a reference standard. Sample turbidity is proportional to the intensity of the scattered light. The reference standard is a formazin suspension in water, which has a known turbidity value. [Pg.173]

Defining these units of measurement and establishing the formazin standard for turbidity measurement has not prevented the use of other units. For instance, nephelometric turbidity unit (NTU), identical to formazin turbidity unit (FTU) and equal to the FNU, are displayed by many types of commercial turbidimeters used in Romania. Some instruments also indicate Jackson turbidity units (JTU) or the silica unit (mg/1 Si02) [6], that are still found as references today. [Pg.57]

Suspended solids are typically measured using turbidity. Turbidity measures the light-scattering ability of particles in water. The water quality guidelines call for an influent turbidity of less than 1 Nephelometric Turbidity Units (NTU), which also happens to be a warranty requirement of membrane manufacturers. Exceed 1 NTU and the membrane warranty is voided. The lower the turbidity, the less likely the membranes are to foul with suspended solids. RO best practices call for feed water turbidity less than 0.5 NTU. [Pg.125]

Hongve DG (1998) Comparison of nephelometric turbidity measurements using wavelengths 400-600 and 860. Water Res 32(10) 3143-3145... [Pg.403]

Turbidity is the measure of suspended material in the water. Drinking water turbidity is measured by nephelometric turbidity units (NTU). The standard is based on a monthly average, not to exceed 1.0 mg L1. Also, the level must not exceed 5.0 mg L 1 on two consecutive days. [Pg.479]

Turbidity reflects the amount of suspended matter in water (e.g., clay, silt, organic and inorganic matter, and plankton). The standard measure of turbidity is the turbidity unity (TU), which is based on the optical property of a water sample, causing transmitted light to be limited. As the number of particles increases, turbidity increases. The measuring instrument is called a nephelometer, and the readings are expressed as nephelometric turbidity units (NTU). [Pg.491]

For example, Getcher Fish s turbidimeter is aimed at the sport fishing market, so it does not have to be as accurate as similar instruments used for environmental measurements. At a level of 10 NTUs (nephelometric turbidity units), the measurement variation can be within 2.5 NTUs and still be acceptable. [Pg.293]

The nephelometric method involves illuminating the sample in a turbidimeter and measuring the amount of light scattered at 90 to the incident beam. The higher the intensity of scattered light, the greater the sample turbidity. A formazan polymer suspension is used as the standard and results are reported in nephelometric turbidity units (NTU). The greater precision, sensitivity, and wider applicability of this method make it preferable to the visual method. [Pg.230]

Figure 17. Turbidity of the microbe-enriched solution of pohl ethyienediamine tetraacetic acid-co-lactose) (PEL, 3). Buffer was 0.1 M potassium phosphate buffer (pH 7.2) at 25 consisting of 71.7 mL of 1 M K2HPO4 and 28.3 mL of I M K211PO4 Medium was mineral medium solution. (NTU = Nephelometric Turbidity Unit). Figure 17. Turbidity of the microbe-enriched solution of pohl ethyienediamine tetraacetic acid-co-lactose) (PEL, 3). Buffer was 0.1 M potassium phosphate buffer (pH 7.2) at 25 consisting of 71.7 mL of 1 M K2HPO4 and 28.3 mL of I M K211PO4 Medium was mineral medium solution. (NTU = Nephelometric Turbidity Unit).
Although the above is the internet page title, if you type it in, the page is apparently unavailable. To find it, go to www.lamotte.com, click on product manuals and then click on 2020 Turbidity Meter. This takes you to the 2020 Turbidity Meter Instruction Manual. The image shown in Fig. 2.7 is on page 9 and the specification for Nephelometric turbidity, calibrated in NTU is given on page 5. [Pg.46]

Nowadays, optical instruments, known as turbidimeters, provide objective measurements of light diffused in a given direction. If the measurements are made perpendicularly to the incident light, the apparatus is called a nephelometer. Results, expressed in NTU (nephelometric turbidity units), are correlated with the wine s appearance (Section 11.3.1). These instruments are very sensitive, which is especially useful in assessing the effectiveness of a treatment, e.g. filtration. [Pg.287]

Turbidity is measured by appreciating the disturbance in the diffusion of light caused by contact with particles in a liquid. A turbidimeter measures the intensity of the diffused light. A turbidimeter that makes measurements at a 90° angle is also known as a nephelometer. These apparatus are calibrated in NTU (nephelometric turbidity units). [Pg.336]


See other pages where Nephelometric Turbidity is mentioned: [Pg.444]    [Pg.230]    [Pg.900]    [Pg.987]    [Pg.257]    [Pg.266]    [Pg.806]    [Pg.732]    [Pg.225]    [Pg.226]    [Pg.615]    [Pg.536]    [Pg.289]    [Pg.57]    [Pg.349]    [Pg.282]    [Pg.365]    [Pg.145]    [Pg.203]    [Pg.4855]    [Pg.569]    [Pg.2538]    [Pg.28]    [Pg.29]    [Pg.18]   
See also in sourсe #XX -- [ Pg.72 , Pg.248 ]




SEARCH



Nephelometric turbidity detector

Nephelometric turbidity units

Turbidity

© 2024 chempedia.info