Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Mass spectrometers fragmentations

One of the key components in the system is the mass spectrometer. Fragments are ionized by a VUV laser pulse between a pair of plane parallel-plate electrodes (6 x 14 cm). One of the plane electrodes has a slit of 1 x 10 cm, which is covered by a metal mesh. The slit is parallel to the VUV laser beam and is the entrance of the mass spectrometer. Ions are accelerated by a pulsed electric field present between the plane parallel-plate electrodes, and then pass through the slit before they enter the mass spectrometer. [Pg.171]

Fig. 7. Protein identification with electrospray tandem mass spectrometry and a triple quadrupole mass spectrometer. Fragment spectra of several peptides are generated during one investigation. From the fragment spectra short sequence stretches can be read. Together with their mass location in the peptide of the measured mass, they can be used to specifically identify a protein in the database. Because the protein identification depends only on one peptide, several proteins can be identified from one sample. Fig. 7. Protein identification with electrospray tandem mass spectrometry and a triple quadrupole mass spectrometer. Fragment spectra of several peptides are generated during one investigation. From the fragment spectra short sequence stretches can be read. Together with their mass location in the peptide of the measured mass, they can be used to specifically identify a protein in the database. Because the protein identification depends only on one peptide, several proteins can be identified from one sample.
Both the liquid and gas products were analyzed by gas chromatography. The column for the liquid analysis was 20% Apiezon L on 60-80 mesh Chromosorb P. The column measured 1/4 inch by 7 feet. The gas analysis utilized a 1/4 inch by 10 foot column of 60-80 mesh Chromosorb 102. Temperature programming was required in both analyses. Identification of the GC peaks was based on retention time of pure compounds when these were available. In addition, two of the samples were analyzed by combined gas chromatography-mass spectrometry. By comparing the observed mass spectrometer fragmentation patterns with tabulated patterns it was possible to identify virtually every component in the product. Further details are available in the theses by Wu (23) and Early (J+). [Pg.74]

In all mass spectrometers, fragmentation continues to occur outside the source (that is, after full acceleration). If, in a sector instrument, an ion (Mi+) decomposes to M2 prior to entering the magnetic field [Mj - M2" -t- (Mj — M2)], the M2 will no longer have the same kinetic energy as the normal ions and, therefore, will not be mass analyzed at M2. Instead, its kinetic energy is M2(M2/Mi) because of energy conservation, and the ion will be analyzed at a lower apparent mass m =... [Pg.470]

Atmospheric pressure chemical ionization. This technique is compatible with the effluent from a chemically suppressed ion chromatography system (30). In this case, field assisted ion evaporation, operated in the negative ion mode, gave easily identified molecular ions for alkyl sulfates and sulfonates. By using a tandem mass spectrometer, fragmentation data were also obtained, corresponding chiefly to loss of a S03 or HSO4" ion. [Pg.466]

In addition to the low accuracy of this method, when organic compounds are pyrolyzed inside the mass spectrometer, fragments are... [Pg.63]

In a mass spectrometer, the molecules, in the gaseous state, are ionized and fragmented. The fragments are detected as a function of their mass-to-charge ratio, m/e. The graphical representation of the ion intensity as a function of m/e makes up the mass spectrogram as illustrated In Figure 3.1. [Pg.44]

In the FFR of the sector mass spectrometer, the unimolecular decomposition fragments, and B, of tire mass selected metastable ion AB will, by the conservation of energy and momentum, have lower translational kinetic energy, T, than their precursor ... [Pg.1335]

Time-of-flight mass spectrometers have been used as detectors in a wider variety of experiments tlian any other mass spectrometer. This is especially true of spectroscopic applications, many of which are discussed in this encyclopedia. Unlike the other instruments described in this chapter, the TOP mass spectrometer is usually used for one purpose, to acquire the mass spectrum of a compound. They caimot generally be used for the kinds of ion-molecule chemistry discussed in this chapter, or structural characterization experiments such as collision-induced dissociation. Plowever, they are easily used as detectors for spectroscopic applications such as multi-photoionization (for the spectroscopy of molecular excited states) [38], zero kinetic energy electron spectroscopy [39] (ZEKE, for the precise measurement of ionization energies) and comcidence measurements (such as photoelectron-photoion coincidence spectroscopy [40] for the measurement of ion fragmentation breakdown diagrams). [Pg.1354]

Although GGMS is the most widely used ana lytical method that combines a chromatographic sep aration with the identification power of mass spectrometry it is not the only one Chemists have coupled mass spectrometers to most of the mstru ments that are used to separate mixtures Perhaps the ultimate is mass spectrometry/mass spectrome try (MS/MS) m which one mass spectrometer gener ates and separates the molecular ions of the components of a mixture and a second mass spec trometer examines their fragmentation patterns ... [Pg.573]

Exact Mass Differences. If the exact mass of the parent or fragment ions are ascertained with a high-resolution mass spectrometer, this relationship is often useful for combinations of C, H, N, and O (Table 1.15b) ... [Pg.813]

In GC-MS effluent from the column is introduced directly into the mass spectrometer s ionization chamber in a manner that eliminates the majority of the carrier gas. In the ionization chamber all molecules (remaining carrier gas, solvent, and solutes) are ionized, and the ions are separated by their mass-to-charge ratio. Because each solute undergoes a characteristic fragmentation into smaller ions, its mass spectrum of ion intensity as a function of mass-to-charge ratio provides qualitative information that can be used to identify the solute. [Pg.571]

Instead of the fast-atom beam, a primary ion-beam gun can be used in just the same way. Generally, such an ion gun emits a stream of cesium ions (Cs ), which are cheaper to use than xenon but still have large mass (atomic masses Cs, 139 Xe, 131). Although ion guns produce no fragment ions in the primary beam, they can contaminate the mass spectrometer by deposition with continued use. [Pg.18]

If a sample solution is introduced into the center of the plasma, the constituent molecules are bombarded by the energetic atoms, ions, electrons, and even photons from the plasma itself. Under these vigorous conditions, sample molecules are both ionized and fragmented repeatedly until only their constituent elemental atoms or ions survive. The ions are drawn off into a mass analyzer for measurement of abundances and mJz values. Plasma torches provide a powerful method for introducing and ionizing a wide range of sample types into a mass spectrometer (inductively coupled plasma mass spectrometry, ICP/MS). [Pg.87]

By being able to obtain an unequivocal relative molecular mass, or even a molecular formula derived from that mass, the hybrid mass spectrometer becomes a powerful tool for investigating single substances or mixtures of substances. With an APCI inlet, fragmentation can be induced to obtain structural information (see Chapter 9). [Pg.167]

In one instrument, ions produced from an atmospheric-pressure ion source can be measured. If these are molecular ions, their relative molecular mass is obtained and often their elemental compositions. Fragment ions can be produced by suitable operation of an APCI inlet to obtain a full mass spectrum for each eluting substrate. The system can be used with the effluent from an LC column or with a solution from a static solution supply. When used with an LC column, any detectors generally used with the LC instrument itself can still be included, as with a UV/visible diode array detector sited in front of the mass spectrometer inlet. [Pg.167]

A suitable mass spectrometer can be set to measure relative atomic, molecular, or fragment ion mass. [Pg.270]

Prior separation of mixtures into individual components may not be needed. If the mass spectrometer is capable of MS/MS operation, one of the mass spectrometers is used to isolate individual ions according to m/z value (mass-to-charge ratio), and the other is used to examine their fragmentation products to obtain structural information. [Pg.277]


See other pages where Mass spectrometers fragmentations is mentioned: [Pg.177]    [Pg.164]    [Pg.225]    [Pg.42]    [Pg.964]    [Pg.38]    [Pg.438]    [Pg.282]    [Pg.108]    [Pg.209]    [Pg.438]    [Pg.191]    [Pg.2912]    [Pg.60]    [Pg.177]    [Pg.164]    [Pg.225]    [Pg.42]    [Pg.964]    [Pg.38]    [Pg.438]    [Pg.282]    [Pg.108]    [Pg.209]    [Pg.438]    [Pg.191]    [Pg.2912]    [Pg.60]    [Pg.252]    [Pg.2060]    [Pg.550]    [Pg.568]    [Pg.14]    [Pg.57]    [Pg.74]    [Pg.105]    [Pg.134]    [Pg.160]    [Pg.173]    [Pg.225]    [Pg.225]    [Pg.228]    [Pg.228]    [Pg.238]    [Pg.243]    [Pg.277]   
See also in sourсe #XX -- [ Pg.248 , Pg.249 , Pg.250 , Pg.251 , Pg.252 , Pg.253 , Pg.254 , Pg.255 , Pg.256 ]




SEARCH



Fragmentation in the Time-of-Flight Mass Spectrometer

Mass fragmentation

© 2024 chempedia.info