Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Lowest calibrated level

However, when the LOQ of a method is significantly lower than the actual concentrations monitored for compliance with a MRL, it may be more appropriate to carry out the validation experiments based on a lowest calibrated level (LCL), typically 0.5 x the MRL. For use in a regulatory program, the limits of detection and quantification are important parameters when the method will be applied to estimate exposure to residues, where there may be an interest in monitoring residues at concentrations below the MRL, or when conducting residue analyses for substances that do not have ADIs or MRLs. For monitoring compliance with a MRL, it is important that an LCL be included in the analysis that adequately demonstrates that the MRL concentration may be reliably determined. The LCL of a method used to support a MRL should not be less than the LOQ. [Pg.288]

The chromatograms of the nitrosamines included in this method show a quick elution of the compounds from 7.87 min for NDM A to 12.47 min for NMOR allowing a short cycle time for a high sample throughput. High peak intensities as shown in Figure 4.66 are achieved at the lowest calibration level of 1 ppb. The nitrosamine MDMA can be detected with good S/N values. [Pg.598]

Adequate sensitivity should be demonstrated and estimates of the limit of detection (LOD) and the limit of quantitation (LOQ) should be provided. The slope of the calibration line may indicate the ability of the method to distinguish the tme analyte concentration. The LOD of a method is the lowest analyte concentration that produces a reproducible response detectable above the noise level of the system. The LOQ is the lowest level of analyte that can be accurately and precisely measured. For a regulatory method, quantitation is limited by the lowest calibration standard. The techniques for these estimations should be described. [Pg.85]

Calibration data (e.g., linearity or sensitivity) are not discussed in detail between laboratories, but a typical calibration starts with 50% of the lowest fortification level and requires at least three additional calibration levels. Another point of calibration is the use of appropriate standards. In 1999 a collaborative study tested the effect of matrix residues in final extracts on the GC response of several pesticides.Five sample extracts (prepared for all participants in one laboratory using the German multi-residue procedure) and pure ethyl acetate were fortified with several pesticides. The GC response of all pesticides in all extracts was determined and compared with the response in the pure solvent. In total, 20 laboratories using 47 GC instruments... [Pg.125]

The straight line portion of a calibration curve Jrom the lowest detection level to the highest concentration before the line begins to bend. This usually is given in powers of 10. For example, the LDR of a hydrogen flame ionization detector is about 10. ... [Pg.230]

In some cases with extremely sensitive sensors a zero-point calibration might be difficult to perform. because traces of the analyte would always be present or might easily be carried into the calibration process by solvents or reagents. The lowest measurable level is then called the blank value. The LOD is then defined as three times the standard deviation of the blank value, expressed in concentration units (not in the units of the signal, as is sometimes improperly suggested). [Pg.956]

The diameter of the capillary tube must be known with accuracy and the cross-section must be truly circular. It is not as a rule easy to obtain tubing of uniform circular bore, but in default of time for the tedious process of calibrating tubing the difficulty may be overcome by the following method due to Ferguson Proc. Phys. Soc. XXXVI. 37, 1923) Lengths of capillary tube are examined under a micrometer until one is found whose end cross-section is circular. This tube is then used so that the measured end dips downwards into the liquid the upper end, instead of being open to the atmosphere is connected to a source of pressure and a manometer, and the meniscus is forced down until its lowest point is level with the end of the tube, so as to be observed at the only point where its curvature is accurately known. If then p be the pressure in dynes/cm. recorded by the manometer we have... [Pg.11]

Typically, laboratories choose the analyte PQL value at 2-10 times its MDL. The selection, however, is not entirely arbitrary because the laboratories must use the selected PQL concentration value as the lowest standard in the multipoint calibration curve. This enables the laboratory to assure that even at a low concentration level, the analyte is detected, identified, and quantified correctly. Therefore, the PQL may be also defined as a concentration that is 2-10 times greater than the MDL and that represents the lowest point on the calibration curve during routine laboratory operations. [Pg.241]

Figure 8-1 4 Calibration curve of response, R, vs. concentration, c. The slope of the calibration curve is called the calibration sensitivity, m. The detection limit, DL, designates the lowest concentration that can be measured at a specified confidence level. Figure 8-1 4 Calibration curve of response, R, vs. concentration, c. The slope of the calibration curve is called the calibration sensitivity, m. The detection limit, DL, designates the lowest concentration that can be measured at a specified confidence level.

See other pages where Lowest calibrated level is mentioned: [Pg.118]    [Pg.119]    [Pg.120]    [Pg.124]    [Pg.126]    [Pg.118]    [Pg.119]    [Pg.120]    [Pg.124]    [Pg.126]    [Pg.323]    [Pg.371]    [Pg.82]    [Pg.432]    [Pg.4004]    [Pg.55]    [Pg.104]    [Pg.52]    [Pg.1122]    [Pg.207]    [Pg.867]    [Pg.281]    [Pg.509]    [Pg.91]    [Pg.175]    [Pg.194]    [Pg.147]    [Pg.232]    [Pg.66]    [Pg.91]    [Pg.151]    [Pg.46]    [Pg.201]    [Pg.95]    [Pg.206]    [Pg.144]    [Pg.95]    [Pg.59]    [Pg.635]    [Pg.451]    [Pg.91]    [Pg.1360]    [Pg.3]    [Pg.15]    [Pg.342]   
See also in sourсe #XX -- [ Pg.118 ]




SEARCH



© 2024 chempedia.info