Liquid mass spectrum

Kovat s retention index (p. 575) liquid-solid adsorption chromatography (p. 590) longitudinal diffusion (p. 560) loop injector (p. 584) mass spectrum (p. 571) mass transfer (p. 561) micellar electrokinetic capillary chromatography (p. 606) micelle (p. 606) mobile phase (p. 546) normal-phase chromatography (p. 580) on-column injection (p. 568) open tubular column (p. 564) packed column (p. 564) peak capacity (p. 554)... [Pg.609]

In many applications in mass spectrometry (MS), the sample to be analyzed is present as a solution in a solvent, such as methanol or acetonitrile, or an aqueous one, as with body fluids. The solution may be an effluent from a liquid chromatography (LC) column. In any case, a solution flows into the front end of a mass spectrometer, but before it can provide a mass spectrum, the bulk of the solvent must be removed without losing the sample (solute). If the solvent is not removed, then its vaporization as it enters the ion source would produce a large increase in pressure and stop the spectrometer from working. At the same time that the solvent is removed, the dissolved sample must be retained so that its mass spectrum can be measured. There are several means of effecting this differentiation between carrier solvent and the solute of interest, and thermospray is just one of them. Plasmaspray is a variant of thermospray in which the basic method of solvent removal is the same, but the number of ions obtained is enhanced (see below). [Pg.71]

By passing a continuous flow of solvent (admixed with a matrix material) from an LC column to a target area on the end of a probe tip and then bombarding the target with fast atoms or ions, secondary positive or negative ions are ejected from the surface of the liquid. These ions are then extracted into the analyzer of a mass spectrometer for measurement of a mass spectrum. As mixture components emerge from the LC column, their mass spectra are obtained. [Pg.86]

Schematic diagram of an orthogonal Q/TOF instrument. In this example, an ion beam is produced by electrospray ionization. The solution can be an effluent from a liquid chromatography column or simply a solution of an analyte. The sampling cone and the skimmer help to separate analyte ions from solvent, The RF hexapoles cannot separate ions according to m/z values and are instead used to help confine the ions into a narrow beam. The quadrupole can be made to operate in two modes. In one (wide band-pass mode), all of the ion beam passes through. In the other (narrow band-pass mode), only ions selected according to m/z value are allowed through. In narrow band-pass mode, the gas pressure in the middle hexapole is increased so that ions selected in the quadrupole are caused to fragment following collisions with gas molecules. In both modes, the TOF analyzer is used to produce the final mass spectrum. Schematic diagram of an orthogonal Q/TOF instrument. In this example, an ion beam is produced by electrospray ionization. The solution can be an effluent from a liquid chromatography column or simply a solution of an analyte. The sampling cone and the skimmer help to separate analyte ions from solvent, The RF hexapoles cannot separate ions according to m/z values and are instead used to help confine the ions into a narrow beam. The quadrupole can be made to operate in two modes. In one (wide band-pass mode), all of the ion beam passes through. In the other (narrow band-pass mode), only ions selected according to m/z value are allowed through. In narrow band-pass mode, the gas pressure in the middle hexapole is increased so that ions selected in the quadrupole are caused to fragment following collisions with gas molecules. In both modes, the TOF analyzer is used to produce the final mass spectrum.
The mass spectrum of polymeric sulfur S, prepared from either liquid sulfur or by extraction of commercial flowers of sulfur , has been measured and interpreted in terms of Ss, Sy, and Ss molecules leaving the polymer on heating and depolymerization [203]. This result is in agreement with depolymerization studies in solution which also show Ss and Sy as the major thermal degradation products [174]. [Pg.89]

Fig. 34 Mass spectrum of sulfur clusters obtained by evaporation of liquid sulfur followed by cooling and adiabatic expansion of the vapor [209]. The figures at the peaks give the number of Sg molecules which represent this particular mass... Fig. 34 Mass spectrum of sulfur clusters obtained by evaporation of liquid sulfur followed by cooling and adiabatic expansion of the vapor [209]. The figures at the peaks give the number of Sg molecules which represent this particular mass...
The liquid jet from the DLI probe has to be initiated at atmospheric pressure, i.e. before insertion of the interface into the mass spectrometer, and, for best performance, the spray direction has to be coaxial to the probe. Any deviation from this, however slight, tends to produce changes in the mass spectrum obtained. [Pg.141]

The flow rate of liquid in the HPLC-electrospray system is paramount in determining performance both from chromatographic and mass spectrometric perspectives. The flow rate affects both the size and size distribution of the droplets formed during the electrospray process (not all droplets are the same size) and, consequently, the number of charges on each droplet. This, as we will see later, has an effect on the appearance of the mass spectrum which is generated. It should also be noted that the smaller the diameter of the spraying capillary, then... [Pg.159]

ESI-MS has emerged as a powerful technique for the characterization of biomolecules, and is the most versatile ionization technique in existence today. This highly sensitive and soft ionization technique allows mass spectrometric analysis of thermolabile, non-volatile, and polar compounds and produces intact ions from large and complex species in solution. In addition, it has the ability to introduce liquid samples to a mass detector with minimum manipulation. Volatile acids (such as formic acid and acetic acid) are often added to the mobile phase as well to protonate anthocyanins. A chromatogram with only the base peak for every mass spectrum provides more readily interpretable data because of fewer interference peaks. Cleaner mass spectra are achieved if anthocyanins are isolated from other phenolics by the use of C18 solid phase purification. - ... [Pg.493]

Fig. 3. (Top left) Chemical methods used to depolymerize the polyesters. (Top right) Thin-layer and gas-liquid chromatograms (as trimethylsilyl derivatives) of the monomer mixture obtained from the cutin of peach fruits by LiAlD4 treatment. In the thin-layer chromatogram the five major spots are, from the bottom, C18 tetraol, C16 triol, and C18 triol (unresolved), diols, and primary alcohol. Nx = C16 alcohol N2= C18 alcohol Mj = C16 diol M2 = C18 diol D = C16 triol D2 and D3 = unsaturated and saturated C18 triol, respectively, T4 and T2, unsaturated and saturated C18 tetraol, respectively. (Bottom) Mass spectrum of component D3 in the gas chromatogram. BSA = bis-N,O-trimethylsilyl acetamide... Fig. 3. (Top left) Chemical methods used to depolymerize the polyesters. (Top right) Thin-layer and gas-liquid chromatograms (as trimethylsilyl derivatives) of the monomer mixture obtained from the cutin of peach fruits by LiAlD4 treatment. In the thin-layer chromatogram the five major spots are, from the bottom, C18 tetraol, C16 triol, and C18 triol (unresolved), diols, and primary alcohol. Nx = C16 alcohol N2= C18 alcohol Mj = C16 diol M2 = C18 diol D = C16 triol D2 and D3 = unsaturated and saturated C18 triol, respectively, T4 and T2, unsaturated and saturated C18 tetraol, respectively. (Bottom) Mass spectrum of component D3 in the gas chromatogram. BSA = bis-N,O-trimethylsilyl acetamide...
Figure 6. FD Mass Spectrum of HTE Liquid Polymer (Water as modifier)... Figure 6. FD Mass Spectrum of HTE Liquid Polymer (Water as modifier)...
The MC-ICP-MS consists of four main parts 1) a sample introduction system that inlets the sample into the instrument as either a liquid (most common), gas, or solid (e.g., laser ablation), 2) an inductively coupled Ar plasma in which the sample is evaporated, vaporized, atomized, and ionized, 3) an ion transfer mechanism (the mass spectrometer interface) that separates the atmospheric pressure of the plasma from the vacuum of the analyzer, and 4) a mass analyzer that deals with the ion kinetic energy spread and produces a mass spectrum with flat topped peaks suitable for isotope ratio measurements. [Pg.118]

Mass spectrometry is a sensitive analytical technique which is able to quantify known analytes and to identify unknown molecules at the picomoles or femto-moles level. A fundamental requirement is that atoms or molecules are ionized and analyzed as gas phase ions which are characterized by their mass (m) and charge (z). A mass spectrometer is an instrument which measures precisely the abundance of molecules which have been converted to ions. In a mass spectrum m/z is used as the dimensionless quantity that is an independent variable. There is still some ambiguity how the x-axis of the mass spectrum should be defined. Mass to charge ratio should not lo longer be used because the quantity measured is not the quotient of the ion s mass to its electric charge. Also, the use of the Thomson unit (Th) is considered obsolete [15, 16]. Typically, a mass spectrometer is formed by the following components (i) a sample introduction device (direct probe inlet, liquid interface), (ii) a source to produce ions, (iii) one or several mass analyzers, (iv) a detector to measure the abundance of ions, (v) a computerized system for data treatment (Fig. 1.1). [Pg.4]

Dass C (1996) The Role of a Liquid Matrix in Controlling FAB-Induced Fragmentation. J Mass Spectrum 31 77... [Pg.152]

It is now common to couple an instrument for separating a mixture of organic compounds e.g. using gas chromatography (GC) or high performance liquid chromatography (HPLC), directly to the input of a mass spectrometer. In this way, as each individual compound is separated from the mixture, its mass spectrum can be recorded and compared automatically with the library of known compounds and identified immediately if it is a known compound. [Pg.28]


See other pages where Liquid mass spectrum is mentioned: [Pg.155]    [Pg.155]    [Pg.20]    [Pg.55]    [Pg.56]    [Pg.72]    [Pg.84]    [Pg.163]    [Pg.257]    [Pg.265]    [Pg.403]    [Pg.69]    [Pg.195]    [Pg.234]    [Pg.368]    [Pg.498]    [Pg.47]    [Pg.103]    [Pg.117]    [Pg.65]    [Pg.16]    [Pg.327]    [Pg.327]    [Pg.328]    [Pg.338]    [Pg.368]    [Pg.71]    [Pg.214]    [Pg.568]    [Pg.220]    [Pg.218]    [Pg.159]   