Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Level zero tests

ASTM E 1421-94, Standard Practice for Describing and Measuring Performance of Fourier Transform Infrared (FT-IR) Spectrometers Level Zero and Level One Tests, 1994. [Pg.107]

The variable X3 takes 3 levels, so it was chosen to correspond to the stirring rate. The normal rate for the dissolution testing of the formulation was 75 rpm, and the extremes to be tested were 75 25 rpm. The coded levels 0.866 correspond therefore to 50 and 100 rpm. (It would also have been possible to set the levels 1 to 50 and 100 rpm, in which case the levels coded 0.866 would have corresponded to 53 and 97 rpm.) If each experiment were carried out twice, these experiments could be done in two dissolution runs. We assume here that the run to run reproducibility is sufficiently good. There remain 7 experiments at the level zero of stirring speed (75 rpm). These experiments also were carried out in duplicate. This allowed the whole experimental design to be carried out using only 4 dissolution runs as shown in table 5.23. [Pg.244]

Often we have data from several populations that we believe follow the same parametric distribution (such as the normal distribution), but may have different values of the parameter (such as the mean). The classical frequentist approach would be to analyze each population separately. The maximum likelihood estimate of the parameter for each population would be estimated from the sample from that population. Simultaneous confidence intervals such as Bonferroni, Tiikey, or Scheff6 intervals would be used for the difference between different population parameter values. These wider intervals would control the overall confidence level, and the overall significance level for testing the hypothesis that the differences between all the population parameters are zero. However, these intervals don t do anything about the parameter estimates themselves. [Pg.244]

Filtration experiments are typically conducted in pilot scale equipment and generally tests are conducted either at constant pressure or constant rate to determine axo, as well as s and Rf, for a given sludge and filter medium. Such tests provide empirical information that will enable the time required tor the pressure drop to reach the desired level for a specified set of operating conditions to be determined. In the initial stages of filtration, the filter medium has no cake. Furthermore, Ap is not zero, but has a value that is a function of the resistance of the medium for a given flowrate. This initial condition can be stated as ... [Pg.169]

Therefore, a different approach was followed in the present paper in order to improve the understanding of the relationship between the structure and the behavior of crosslinked polymers. A series of directly comparable model polymers were prepared with crosslink densities varying from high (thermoset) to zero (thermoplastic). Five polymers with well defined crosslink densities [11] were tested at various levels of deformation. This approach produced a small but assessable and fairly consistant body of results. Basic relationships derived from these results were related to corresponding results from the literature. [Pg.317]

The test for the significance of a slope b is formally the same as a t-test (Section 1.5.2) if the confidence interval CI( ) includes zero, b cannot significantly differ from zero, thus ( = 0. If a horizontal line can be fitted between the plotted CL, the same interpretation applies, cf. Figures 2.6a-c. Note that si, corresponds to fx ean). that is, the standard deviation of a mean. In the above example the confidence interval clearly does not include zero this remains so even if a higher confidence level with t(f = 3, p = 0.001) = 12.92 is used. [Pg.102]

Parameter estimates are tested on whether they differ significantly from zero at a certain probability level. If not, the parameter should be skipped from the model and the model should be redefined even if the test for model adequacy was positive. When the errors have constant variance, the random variable... [Pg.547]

Optimal Salinities Phase inversions at optimal salinity were assessed routinely by salt titrations into systems maintained at constant temperature. For the Leonox IOS surfactant system, increasing levels of salinity were necessary to cause the emulsion state to phase invert as the alkane molecular weight increased (Figure 11). Ihe initial conductivity value at the condition where zero salt had been added may in part reflect the salt contamination naturally present within the supplied formulation. Ihe internal olefin sulphonate species again revealed a linear relationship between EACN and optimal salinity as did all ionic formulations under test (see Figures 12 and 13, plus Table III). Ihe estimation of EACN values for both toluene... [Pg.315]

The level of filler was measured using TGA. A portion of each sample was heated from room temperature to 900°C at a rate of 15°C/min in air. The two PC samples yielded nearly identical TGA traces, as seen in Figure 42. In both cases, there appears to be negligible amount of filler, based on the fact that the mass remaining at the conclusion of the test was near zero. This implies that a filler component was not a possible contributing factor to the cracking. [Pg.647]

In the various sections of this article, it has been attempted to show that heat-flow calorimetry does not present some of the theoretical or practical limitations which restrain the use of other calorimetric techniques in adsorption or heterogeneous catalysis studies. Provided that some relatively simple calibration tests and preliminary experiments, which have been described, are carefully made, the heat evolved during fast or slow adsorptions or surface interactions may be measured with precision in heat-flow calorimeters which are, moreover, particularly suitable for investigating surface phenomena on solids with a poor heat conductivity, as most industrial catalysts indeed are. The excellent stability of the zero reading, the high sensitivity level, and the remarkable fidelity which characterize many heat-flow microcalorimeters, and especially the Calvet microcalorimeters, permit, in most cases, the correct determination of the Q-0 curve—the energy spectrum of the adsorbent surface with respect to... [Pg.259]

In our tests, the furnace was occasionally deficient of oxygen. When oxygen level went to zero, a significant amount of CO was produced as well as visible and sooty smoke. These are indications of incomplete combustion. The only product of incomplete combustion monitored was CO. Incomplete combustion products other than CO were not accounted for. [Pg.419]

The MP2/TZDP optimized structures were then used to calculate the stationary state geometry force constants and harmonic vibrational frequencies, also at the MP2 level. These results serve several purposes. Firstly, they test that the calculated geometry is really an energy minimum by showing all real frequencies in the normal coordinate analysis. Secondly, they provide values of the zero-point energy (ZPE) that can be used... [Pg.18]


See other pages where Level zero tests is mentioned: [Pg.186]    [Pg.186]    [Pg.470]    [Pg.84]    [Pg.273]    [Pg.146]    [Pg.401]    [Pg.404]    [Pg.194]    [Pg.69]    [Pg.150]    [Pg.368]    [Pg.110]    [Pg.321]    [Pg.18]    [Pg.18]    [Pg.88]    [Pg.681]    [Pg.121]    [Pg.249]    [Pg.588]    [Pg.412]    [Pg.227]    [Pg.67]    [Pg.167]    [Pg.75]    [Pg.607]    [Pg.879]    [Pg.544]    [Pg.184]    [Pg.147]    [Pg.102]    [Pg.89]    [Pg.203]    [Pg.368]    [Pg.63]   
See also in sourсe #XX -- [ Pg.186 ]




SEARCH



© 2024 chempedia.info