Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Instrumentation probe mounting

The main disadvantages of surface-mount components stem from the fact that there are no leads to easily grip with instrumentation probes and that there may not be access to the leads... [Pg.293]

The. v-y rastering of the surface relative to the tip needs to be driven by actuators which displace either the sample stage (usually the case for SFM instruments) or the probe mount (usually the case for STM in.struments). The raster drive and the fine r-positioning always use one or other types of piezoelectric device. [Pg.412]

Electric, acoustic, and vibration isolation is essential for high-resolution current measurements and positional control. All instruments are mounted on vibration isolation tables (8 in Figure 19.2d), within a Faraday cage (custom made) with acoustic foam to reduce vibrations (9 in Figure 19.2d). Cameras or an optical microscope (not shown) can be used to aid positioning of the probe. [Pg.662]

When solution must be pumped, consideration should be given to use of holding tanks between the dry feed system and feed pumps, and the solution water supply should be controlled to prevent excessive dilution. The dry feeders may be started and stopped by tank level probes. Variable-control metering pumps can then transfer the alum stock solution to the point of application without further dilution. Means should be provided for calibration of the chemical feeders. Volumetric feeders may be mounted on platform scales. Belt feeders should include a sample chute and box to catch samples for checking actual delivery with set delivery. Gravimetric feeders are usually furnished with totalizers only. Remote instrumentation is frequently used with gravimetric equipment, but seldom used with volumetric equipment. [Pg.95]

Multimode instruments with an IR sensor mounted in the cavity side wall (see Section 3.4) certainly need larger volumes for precise temperature monitoring. Immersion temperature probes require a well-defined minimum volume for accurate measurement, depending on the total vessel volume. It must be ensured that the temperature probe has extensive contact with the reaction mixture, even when the mixture is stirred, in order to obtain reliable, reproducible results. [Pg.105]

FIAs can be based on steady-state intensity measurements without probe amplification, owing to the sensitivity of detection that is possible with fluorescence instrumentation, which exceeds that of spectrophotometers by two or three orders of magnitude. A sensitive fluorometer has been described for an estradiol assay(36) in which the limit of estradiol detection is 3 x KT11 M. Estradiol antibody labeled with rhodamine B is reacted with estradiol samples. Unreacted labeled antibody is removed with Sepharose-estradiol-casein beads, and the remaining fluorescence is directly proportional to the analyte concentration. The detection limit of rhodamine B on the same fluorometer is 5 x 1(T12 M. This instrument uses a 0.75 mW green helium-neon (HeNe) laser to irradiate the sample from above, at the air-liquid interface, to increase the light path and to decrease surface reflections. The sample compartment has a top-mounted photon trap, and a mirror mounted on the side of the sample compartment opposite the PMT to enhance detection. [Pg.460]

Probe cleaners should be mounted in sight-flow glasses for good visibility. (Courtesy of Aimco Instruments, Inc.)... [Pg.333]

Force/Displacement-Temperature Experiments. Force as a function of temperature was obtained by placing a sample in a vertical glass oven and gripping both ends, with one end attached to a load cell and the other to an adjustable mount. A slow nitrogen flow was introduced through the bottom and an RTD probe was placed near the middle of the sample. Output from the probe and load cell was recorded on a Bascom-Turner Instrument model 4000. With a sample in place several heating and cooling cycles were performed at various loads. [Pg.271]


See other pages where Instrumentation probe mounting is mentioned: [Pg.116]    [Pg.18]    [Pg.473]    [Pg.262]    [Pg.262]    [Pg.15]    [Pg.472]    [Pg.180]    [Pg.1141]    [Pg.89]    [Pg.29]    [Pg.425]    [Pg.219]    [Pg.122]    [Pg.102]    [Pg.329]    [Pg.196]    [Pg.76]    [Pg.267]    [Pg.90]    [Pg.748]    [Pg.139]    [Pg.263]    [Pg.182]    [Pg.118]    [Pg.99]    [Pg.258]    [Pg.150]    [Pg.3]    [Pg.37]    [Pg.173]    [Pg.145]    [Pg.5]    [Pg.480]    [Pg.146]    [Pg.404]    [Pg.13]    [Pg.417]    [Pg.273]    [Pg.148]    [Pg.417]    [Pg.24]   
See also in sourсe #XX -- [ Pg.25 ]




SEARCH



Mount

© 2024 chempedia.info