Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Instrumentation inlet system

There are other characteristics of quadrupoles that make them cheaper for attainment of certain objectives. For example, quadrupoles can easily scan a mass spectrum extremely quickly and are useful for following fast reactions. Moreover, the quadrupole does not operate at the high voltages used for magnetic sector instruments, so coupling to atmospheric-pressure inlet systems becomes that much easier because electrical arcing is much less of a problem. [Pg.185]

The choice of a mass spectrometer to fulfill any particular task must take into account the nature of the substances to be examined, the degree of separation required for mixtures, the types of ion source and inlet systems, and the types of mass analyzer. Once these individual requirements have been defined, it is much easier to discriminate among the numerous commercial instruments that are available. Once suitable mass spectrometers have been identified, it is then often a case of balancing capital and running costs, reUability, ea.se of routine use, after-sales service, and manufacturer reputation. [Pg.285]

Because an increase in resolution causes a decrease in sensitivity, it is best to operate at the lowest resolution commensurate with good results. Some instrument data systems will allow calibration with an external reference material such as perfluorokerosene and then use of a secondary reference material for the internal mass reference. Tetraiodothiophene, vaporized using the solids probe inlet, is recommended as the secondary reference. The accurate masses are 79.9721, 127.9045, 162.9045, 206.8765, 253.8090, 293.7950, 333.7810, 460.6855, and 587.5900. For a higher mass standard, use hexaiodobenzene. Because the mass defect for these internal reference ions are so large, a resolution of 2000 is ample to separate these ions from almost any sample ions encountered in GC/MS. [Pg.375]

Before describing the single parts, let us examine the general features of a mass spectrometer (Figure 2.1). In every MS experiment the first step is the introduction of the sample into the mass spectrometer. It follows that the first part of every instrument is the inlet system that allows the introduction of the sample, generally molecules, into the mass spectrometer. There are different ways to introduce the sample, depending on its purity and properties. [Pg.41]

Of the explosives listed in Table 4, only those such as NG with vapour pressures greater than 10 Pa at 25°C are good candidates for the direct detection of vapour by current instrumental techniques. However, vapour pressure rises markedly with temperature. In addition, consideration of the thermal stability data in Table 4 offers the possibility of heating samples containing traces of involatile explosives such as RDX or PETN to increase their vapour pressure and render them detectable. This is the basis of the common technique of combining a heated inlet system with a vapour-type detector, for example, the method of desorption from a swab on a heated stage often used with IMS or TEA systems. This approach has greatly broadened the scope of what were previously viewed as vapour-type detectors and is now standard practice such instruments are now known as particle detectors. [Pg.25]

Because it is particles in the smaller size range, <2.5 /x.m (PM25), that are of greatest interest with respect to health effects, inlet systems are normally used that exclude larger particles. These size exclusion inlets are usually based on filters, cyclone collectors, or impactors, the principle behind which is discussed shortly. Inlet cutoff diameters from 2.2 to 15 /im are achieved in commercial instruments using these techniques (Chow, 1995). [Pg.608]

The third factor that is important in determining the detection limit is the conversion efficiency of the kinetics. A conversion efficiency of 1.0 requires that the airstream have a velocity substantially less than 200 m/s because uniform mixing of NO is very difficult. At the same time, collisions of the sample airstream with wall surfaces in slower inlet systems may cause a chemical loss of CIO and BrO, because they are both reactive with wall surfaces. The solution to this problem was suggested by Soderman (83). Soderman s novel design consists of two nested ducts in which the air speed is decreased from 200 m/s to 60 m/s in a 14-cm-diameter outer duct that protrudes 60 cm in front of the left wing pod and is reduced to 20 m/s inside a smaller 5-cm-square duct in which the measurements are made. The entrance to the smaller measurement duct is 60 cm downstream of the entrance to the outer duct, and the NO injector tubes, the two CIO detection axes, and the one BrO axis are 25 cm, 37.5 cm, 55 cm, and 72.5 cm downstream of the entrance of the measurement duct. Ninety percent of the air that enters the outer duct bypasses the measurement duct through additional duct work, and only the center 10% of the airstream is captured and sampled by the measurement duct. These two flows are recombined downstream of the instrument and are vented out the side of the wing pod that houses the instrument. [Pg.181]

Some compounds such as long chain esters and polyethers decompose in the inlet system, and the spectrum obtained is not that of the initial substance. Calibration coefficients are required for quantitative analyses. The sample introduced to the instrument cannot usually be recovered. [Pg.59]

Apparatus. A diagram of the custom built gas chromatographic instrument is shown in Figure 1. Because of the highly reactive and corrosive nature of C1F3, the construction materials tor the instrument were stainless steel, Monel, and nickel. The apparatus is best considered as four parts sample inlet system, column, detector, and detector bypass. [Pg.230]


See other pages where Instrumentation inlet system is mentioned: [Pg.98]    [Pg.104]    [Pg.164]    [Pg.353]    [Pg.205]    [Pg.471]    [Pg.52]    [Pg.194]    [Pg.352]    [Pg.327]    [Pg.16]    [Pg.391]    [Pg.163]    [Pg.287]    [Pg.219]    [Pg.115]    [Pg.230]    [Pg.169]    [Pg.216]    [Pg.305]    [Pg.396]    [Pg.80]    [Pg.114]    [Pg.401]    [Pg.41]    [Pg.222]    [Pg.103]    [Pg.270]    [Pg.68]    [Pg.95]    [Pg.177]    [Pg.212]    [Pg.302]    [Pg.6]    [Pg.19]    [Pg.163]    [Pg.216]    [Pg.305]   
See also in sourсe #XX -- [ Pg.3 , Pg.42 ]




SEARCH



Inlet

Inlet system

© 2024 chempedia.info