Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Instrumentation differentiator

Electrophoretic methods of separation of LD Tsoenzymes have become routine in clinical laboratories. Efforts are now being made to standardize the methodologies used for LD isoenzymes, particularly by Rosalki (38). The preferred methods are based on electrophoresis on a solid medium, so that the several bands may be scanned instrumentally. Differential isoenzyme inhibition with urea or other inhibitors is based on the fact that the heart LD isoenzyme is more resistant to inhibition than other isoenzymes. However, the analyst then has the problem of allocating the observed degree of inhibition between the different isoenzymes of a given sample, a problem that has not been resolved satisfactorily thus far. Hence, differential inhibition is not as reliable for isoenzyme separation as is electrophoresis. [Pg.193]

The increasing sophistication of automated cell counters has resulted in virtual replacement of many manual determinations, such as the white blood cell (WBC) differential (Hunt, 2004). Because the instrument is evaluating thousands of cells individually, as opposed to the traditional 100 used to perform a manual differential, the instrument differential is almost always more accurate than the manual differential. However, notable exceptions... [Pg.17]

Radon Detecting buried U Instruments differentiating radon and thoron... [Pg.71]

Fourier Transformation Infrared (FT-IR) spectra were recorded using a Peridn-Elmer RX-1 spectrometer with KBr pellet from 4,000 to 400 cm . The H NMR and NMR speetra were acquired at 300 MHz on a Bruker-300 spectrometer with 1% tetra-methylsilane (TMS) as an internal standard. The DSC analysis was carried out with a Qj series TA instruments differential scanning calorimeter using 5-7 mg of the sample crimped in alumininm pans at a heating rate of 10°C/min and nnder nitrogen atmosphere with a flow rate of 40 ml/mia The MW reactions were carried out in a Milestone Ine., laboratory MW system with a frequency of 2,450 MHz and controllable power system (max 1,000 W). A 50 ml (diameter 5 cm) Teflon reaction vessel was used. The temperature and time of the reaction were controlled by pre-programmed Easywave software system. [Pg.45]

Particle size distribution of the examined materials was determined both on a volume (Method A) and on a number/volume basis (Method B) both in inert (isopropyl alcohol) and in swelling media (isotonic saline). From the differential number distribution supplied by the instrument, differential and cumulative volume distributions were calculated by means of a suitable program run on an IBM AT personal computer. [Pg.386]

Tg values for NOA 63 (and composites thereof) cured to varying extents were measured using a TA Instruments Differential Scanning Calorimeter (DSC) model 2920 using heating/cooling rate of 10 C/min under nitrogen atmosphere. A PerkinElmer Dynamic Mechanical Analyzer (DMA) 7e was run in tensile mode at an... [Pg.2382]

Allen, H. C. Brauers, T. Finlayson-Pitts, B. J. Illustrating Deviations in the Beer-Lambert Law in an Instrumental Analysis Laboratory Measuring Atmospheric Pollutants by Differential Optical Absorption Spectrometry, /. Chem. [Pg.447]

As m increases, At becomes progressively smaller (compare the difference between the square roots of 1 and 2 (= 0.4) with the difference between 100 and 101 (= 0.05). Thus, the difference in arrival times of ions arriving at the detector become increasingly smaller and more difficult to differentiate as mass increases. This inherent problem is a severe restriction even without the second difficulty, which is that not all ions of any one given m/z value reach the same velocity after acceleration nor are they all formed at exactly the same point in the ion source. Therefore, even for any one m/z value, ions at each m/z reach the detector over an interval of time instead of all at one time. Clearly, where separation of flight times is very short, as with TOF instruments, the spread for individual ion m/z values means there will be overlap in arrival times between ions of closely similar m/z values. This effect (Figure 26.2) decreases available (theoretical) resolution, but it can be ameliorated by modifying the instrument to include a reflectron. [Pg.191]

In a normal quadrupole instrument, the field-free regions are very short. Ions formed in region t will be transmitted by the quadrupole as normal ions. In region 2 there is no differentiation between metastable and normal ions. [Pg.232]

In a sector instrument, which acts as a combined mass/velocity filter, this difference in forward velocity is used to effect a separation of normal and metastable mj" ions (see Chapter 24, Ion Optics of Magnetic/Electric-Sector Mass Spectrometers ). However, as discussed above, the velocity difference is of no consequence to the quadmpole instrument, which acts only as a mass filter, so the normal and metastable mj ions formed in the first field-free region (Figure 33.1) are not differentiated. [Pg.233]

A common mistake for beginners in mass spectrometry is to confuse average atomic mass and isotopic mass. For example, the average atomic mass for chlorine is close to 35.45, but this average is of the numbers and masses of Cl and Cl isotopes. This average must be used for instruments that cannot differentiate isotopes (for example, gravimetric balances). Mass spectrometers do differentiate isotopes by mass, so it is important in mass spectrometry that isotopic masses be used... [Pg.348]

Almost any type of analyzer could be used to separate isotopes, so their ratios of abundances can be measured. In practice, the type of analyzer employed will depend on the resolution needed to differentiate among a range of isotopes. When the isotopes are locked into multielement ions, it becomes difficult to separate all of the possible isotopes. For example, an ion of composition CgHijOj will actually consist of many compositions if all of the isotopes ( C, C, H, H, 0, O, and 0) are considered. To resolve all of these isotopic compositions before measurement of their abundances is difficult. For low-molecular-mass ions (HjO, COj) or for atomic ions (Ca, Cl), the problems are not so severe. Therefore, most accurate isotope ratio measurements are made on low-molecular-mass species, and resolution of these even with simple analyzers is not difficult. The most widely used analyzers are based on magnets, quadrupoles, ion traps, and time-of-flight instruments. [Pg.365]

To differentiate tteir functions and modes of operation, the array collector of spatially dispersed m/z values is still called an array collector for historical reasons, but the other multipoint detector of a temporally dispersed range of m/z values is called a microchannel plate (typically used in time-of-flight instruments). [Pg.410]

Detection limit. The detection limit of an instrument should be differentiated from its sensitivity. The detection limit reflects the smallest flow of sample or the lowest partial pressure that gives a signal that can be distinguished from the background noise. One must specify the experimental conditions used and give the value of signal-to-noise ratio corresponding to the detection limit. [Pg.436]

Linearizing the output of the transmitter. Functions such as square root extraction of the differential pressure for a head-type flowmeter can be done within the instrument instead of within the control system. [Pg.768]

Differential pressures and subatmospherie pressures should be measured by manometers with a fluid that is ehemieally stable when in eontaet with the test gas. Mereury traps should be used where neeessary to prevent the manometer fluid from entering the proeess piping. Errors in these instruments should not exeeed 0.25%. [Pg.696]

Differential pressure is included in the pressure instrument class. Good differential readout gauges are still not all that common. Fortunately, in the transducer category, they are more readily available. Figure 8-26 covers some installation details for pressure-oriented instrument piping, supplementing the information presented earlier in Figure 8-8. [Pg.343]

As time goes on, the ultimate resolution of the SEM operated in these modes will probably level out near 1 nm. The major growth of SEMs now seems to be in the development of specialized instruments. An environmental SEM has been developed that uses differential pumping to permit the observation of specimens at higher pressures. Photographs of the formation of ice crystals have been taken and the instrument has particular application to samples that are not vacuum compatible, such as biological samples. [Pg.83]


See other pages where Instrumentation differentiator is mentioned: [Pg.94]    [Pg.106]    [Pg.286]    [Pg.3326]    [Pg.231]    [Pg.87]    [Pg.94]    [Pg.106]    [Pg.286]    [Pg.3326]    [Pg.231]    [Pg.87]    [Pg.1308]    [Pg.1313]    [Pg.1794]    [Pg.1807]    [Pg.640]    [Pg.55]    [Pg.219]    [Pg.226]    [Pg.281]    [Pg.282]    [Pg.693]    [Pg.201]    [Pg.708]    [Pg.2311]    [Pg.2335]    [Pg.2554]    [Pg.72]    [Pg.366]    [Pg.224]    [Pg.242]   
See also in sourсe #XX -- [ Pg.638 ]




SEARCH



Amplifiers instrumentation differential

Differential instruments

Differential instruments

Differentially pumped TEM instruments

Instrument design differential scanning calorimetry

Instrument design differential thermal analysis

Instrumentation Long-Differential Mobility

Instrumentation differential probe

Instrumentation for differential

Instrumentation, differential scanning

Instrumentation, differential scanning calorimetry

Instrumentation, differential thermal

Instrumentation, differential thermal analysis

© 2024 chempedia.info