Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Hotness levels measurements

Hot spots are either macroscopic hot spots (measurable) or microscopic hot spots (on a molecular level, immeasurable and similar to those in sonochemistry). [Pg.364]

Figure 13 shows a typical pressurizer level system. Pressurizer temperature is held fairly constant during normal operation. The AP detector for level is calibrated with the pressurizer hot, and the effects of density changes do not occur. The pressurizer will not always be hot. It may be cooled down for non-operating maintenance conditions, in which case a second AP detector, calibrated for level measurement at low temperatures, replaces the normal AP detector. The density has not really been compensated for it has actually been aligned out of the instrument by calibration. [Pg.77]

The intercept of the straight line generated by the two fixed points (that is, the value of t at which V would vanish on that straight line if He could be maintained as an ideal gas down to extremely low temperatures ) is found to be 7b = — 100Vb/( ioo — o) = —273.15°C. This suggests a natural lower limit to temperature, namely, the point where V vanishes. It also suggests a shift of scale whereby the quantity T = lOOL/(Tioo — Vb) is the fundamental entity of interest. Adoption of this method leads an absolute scale for quantifying hotness levels we construct a thermodynamic temperature scale T(K) = r(°C) -I- 273.15, where K stands for kelvins as the temperature unit. This still maintains the desired proportionality between absolute temperature and measured volumes of He at fixed, low pressures. [Pg.7]

The hot/cold intraface level in the upper density lock is determined by the total volume of the primary loop water mass, when the position of the interface level in the lower density lock is kept constant. The temperature measurements for the interface level in the upper lock are basically used for reactor pool volume control purposes. (The reactor primary loop volume control utilizes level measurements in the pressurizer.)... [Pg.237]

In many temperature determinations, one maintains the gas thermometer at a fixed low pressure. A useful quantification scheme is the so-called Celsius scale that assigns the values t = 0 °C (this was the original intent, but nowadays the standard value is t = 0.01 °C) and t = 100 °C to the He gas thermometer maintained at equilibrium respectively with water containing ice and with water equilibrated with steam maintained at 1 bar." Let the volume of heUum gas in a flexible container at a fixed low pressure be the indicator of hotness levels, such that the measured volumes V, Vq, and Vioo correspond to temperatures x, 0, and 100 °C respectively then x is to be specified by... [Pg.7]

Now if the first term in brackets exceeds the second, dEA will necessarily increase i.e., heat will flow into A without external intervention, which requires, according to Clausius, that A will be colder than B. Thus, (din g/dE) is seen to be a measure of the hotness level of a system. To quote Waldram A larger value of (din g/dE) sucks energy into itself. Normally, g increases very rapidly with E. In fight of the above, it is then sensible to postulate the association... [Pg.434]

Installation of a permanent monitoring of the cavity and the RCS level from the control room. For each cold shutdown, when the RCS is to be opened, the utility has installed a precise and reliable level measurement, with an ultrasonic probe which has a contact with the primary hot leg. (This is meant to replaces mobile ultrasonic level measurement). This measurement shall provide a low level alarm in the control room. [Pg.61]

Positive ions are obtained from a sample by placing it in contact with the filament, which can be done by directing a gas or vapor over the hot filament but usually the sample is placed directly onto a cold filament, which is then inserted into the instrument and heated. The positive ions are accelerated from the filament by a negative electrode and then passed into a mass analyzer, where their m/z values are measured (Figure 7.1). The use of a suppressor grid in the ion source assembly reduces background ion effects to a very low level. Many types of mass analyzer could be used, but since very high resolutions are normally not needed and the masses involved are quite low, the mass analyzer can be a simple quadrupole. [Pg.45]

Because of its small size and portabiHty, the hot-wire anemometer is ideally suited to measure gas velocities either continuously or on a troubleshooting basis in systems where excess pressure drop cannot be tolerated. Furnaces, smokestacks, electrostatic precipitators, and air ducts are typical areas of appHcation. Its fast response to velocity or temperature fluctuations in the surrounding gas makes it particularly useful in studying the turbulence characteristics and rapidity of mixing in gas streams. The constant current mode of operation has a wide frequency response and relatively lower noise level, provided a sufficiently small wire can be used. Where a more mgged wire is required, the constant temperature mode is employed because of its insensitivity to sensor heat capacity. In Hquids, hot-film sensors are employed instead of wires. The sensor consists of a thin metallic film mounted on the surface of a thermally and electrically insulated probe. [Pg.110]

In any leaf test program there is always a question as to what vacuum level should be used. With very porous materials, a vacuum in the range of 0.1 to 0.3 bar (3 to 9 in Hg) should be used, and, except for thermal-drying apphcations using hot air, the vacuum level should be adjusted to give an air rate in the range of 450 to 900 mVm h (30 to 40 cfm/ft measured at the vacuum. [Pg.1696]

Fitzgerald et al. (1984) measured pressure fluctuations in an atmospheric fluidized bed combustor and a quarter-scale cold model. The full set of scaling parameters was matched between the beds. The autocorrelation function of the pressure fluctuations was similar for the two beds but not within the 95% confidence levels they had anticipated. The amplitude of the autocorrelation function for the hot combustor was significantly lower than that for the cold model. Also, the experimentally determined time-scaling factor differed from the theoretical value by 24%. They suggested that the differences could be due to electrostatic effects. Particle sphericity and size distribution were not discussed failure to match these could also have influenced the hydrodynamic similarity of the two beds. Bed pressure fluctuations were measured using a single pressure point which, as discussed previously, may not accurately represent the local hydrodynamics within the bed. Similar results were... [Pg.69]

After the activation period, the reactor temperature was decreased to 453 K, synthesis gas (H2 CO = 2 1) was introduced to the reactor, and the pressure was increased to 2.03 MPa (20.7 atm). The reactor temperature was increased to 493 K at a rate of 1 K/min, and the space velocity was maintained at 5 SL/h/gcat. The reaction products were continuously removed from the vapor space of the reactor and passed through two traps, a warm trap maintained at 373 K and a cold trap held at 273 K. The uncondensed vapor stream was reduced to atmospheric pressure through a letdown valve. The gas flow was measured using a wet test meter and analyzed by an online GC. The accumulated reactor liquid products were removed every 24 h by passing through a 2 pm sintered metal filter located below the liquid level in the CSTR. The conversions of CO and H2 were obtained by gas chromatography (GC) analysis (micro-GC equipped with thermal conductivity detectors) of the reactor exit gas mixture. The reaction products were collected in three traps maintained at different temperatures a hot trap (200°C), a warm trap (100°C), and a cold trap (0°C). The products were separated into different fractions (rewax, wax, oil, and aqueous) for quantification. However, the oil and wax fractions were mixed prior to GC analysis. [Pg.250]


See other pages where Hotness levels measurements is mentioned: [Pg.123]    [Pg.195]    [Pg.13]    [Pg.7]    [Pg.8]    [Pg.2054]    [Pg.73]    [Pg.502]    [Pg.6]    [Pg.7]    [Pg.8]    [Pg.478]    [Pg.389]    [Pg.182]    [Pg.285]    [Pg.129]    [Pg.66]    [Pg.72]    [Pg.273]    [Pg.120]    [Pg.627]    [Pg.905]    [Pg.154]    [Pg.913]    [Pg.122]    [Pg.633]    [Pg.502]    [Pg.746]    [Pg.478]    [Pg.245]    [Pg.171]    [Pg.280]    [Pg.274]    [Pg.334]    [Pg.434]    [Pg.84]    [Pg.266]   
See also in sourсe #XX -- [ Pg.7 ]




SEARCH



Hotness levels

Level measurements

© 2024 chempedia.info