Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

External standard compound

This is a much simpler problem (a) the analysis is performed in a short period of time (a few hours, depending on the number of samples) by analysing one sample after other so no sensor drift is expected and (b) no stable external standard compounds are necessary. In connection with this is the following, more difficult problem is it possible to detect the presence of a trace of a perfumery note in a fragrance This problem requires the manufacturer to provide a large number of samples containing the note in order to create a good database to perform ANN (a first exploratory approach to this problem was reported by Branca et at, 2003). [Pg.285]

At X-ray fluorescence analysis (XRF) of samples of the limited weight is perspective to prepare for specimens as polymeric films on a basis of methylcellulose [1]. By the example of definition of heavy metals in film specimens have studied dependence of intensity of X-ray radiation from their chemical compound, surface density (P ) and the size (D) particles of the powder introduced to polymer. Have theoretically established, that the basic source of an error of results XRF is dependence of intensity (F) analytical lines of determined elements from a specimen. Thus the best account of variations P provides a method of the internal standard at change P from 2 up to 6 mg/sm the coefficient of variation describing an error of definition Mo, Zn, Cu, Co, Fe and Mn in a method of the direct external standard, reaches 40 %, and at use of a method of the internal standard (an element of comparison Ga) value does not exceed 2,2 %. Experiment within the limits of a casual error (V changes from 2,9 up to 7,4 %) has confirmed theoretical conclusions. [Pg.104]

To analyze pesticides from the sample, several GC techniques were used GC with FID and EC detectors and GC/MS with external standards. Pesticides are mostly analyzed using split/splitless technique where higher amount of injected solution exits the gas chromatograph without decomposing therefore by quantification of the several pesticides in the filter, we found out how harmful is exposing analysts to pesticide compounds during the GC analysis. [Pg.192]

For details of the clean-up of the pyrolysate of the DIN oven see reference 12. Identification and quantification of PBDD/F was performed by GC/MS (refs. 8-12). This was done for all brominated PBDD and PBDF from mono- through octabromo compounds using external standards which were either prepared (refs. 11-13) or purchased. There exists a total of 210 brominated compounds of PBDD/F. Since not all isomers are available a complete isomer-specific determination could not be performed. [Pg.368]

Adequate precision and accuracy are only likely to be achieved if some standardization procedure is employed and the nature of this, internal or external standards or the method of standard additions, needs to be chosen carefully. If internal standardization procedures are adopted then appropriate compound(s) must be chosen and their effect on the chromatographic and mass spectrometry methods assessed. The ideal internal standard is an isotopically labelled analogue of the analyte but, although there are a number of commercial companies who produce a range of such molecules, these are not always readily available. An analytical laboratory is then faced with the choice of carrying out the synthesis of the internal standard themselves or choosing a less appropriate alternative with implications on the accuracy and precision of the method to be developed. [Pg.270]

Quantitation of anthocyanins has become simple and fast since many anthocy-anin standards became commercially available as external standards in the past decade. When the standards are not available, individual anthocyanins or total monomeric anthocyanins can be determined by the use of a generic external standard such as commercial cyanidin-3-glucoside or other compound structurally similar to the analytes of interest. Individual and total peak areas are measured at 520 nm or their and quantified using external standards by which values are typically slightly different from those via the pH differential method. ... [Pg.486]

The determination of the relationship between detector response and the sample concentration is termed the calibration of the method. There are two types of methods in use for the quantitative analysis of a sample, i.e., the external standard and the internal standard method. An external standard method is a direct comparison of the detector response of a pure compound (standard) to a sample.2 The calibration of the method is performed by preparing standards of varying concentration and analyzing them by a developed method. Method 1 (below) was developed for toluene, and standards of varying concentration were prepared and analyzed. The results obtained are summarized in Table 2 see Figure 3. [Pg.156]

An internal standard method gives more reliable results when elaborate sample preparation is required, as in extraction of a drug substance from biological fluids, or extraction of pesticides and herbicides from soil and plant matter. The addition of internal standard (IS) to the sample and standard acts as a marker to give accurate values of the recovery of the desired compound(s). Since the determination of wt% involves the ratio of the detector responses in the two chromatograms, the injection volume is not critical as in an external standard method. [Pg.159]

To determine the weight percent of compounds 2 and 3 of the LANA reaction presented in scheme 3, Method 6 was developed as an external standard technique. Figure 17 shows a typical chromatogram. [Pg.186]

To determine the weight percent of each compound in the reaction sample, a standard and sample of known weight concentration were prepared and analyzed. The weight percent of each component purity was determined as described in the External Standard Method. [Pg.187]

All NMR spectra were run by Spectral Data Services. No external standard such as TMS was added to the sample as the compounds had (CH3)3Si groups. X-ray diffraction patterns were obtained on Rigaku-Geiger flex x-ray diffraction machine. Melting points and... [Pg.530]

Compound 6 contains seven iron-based units [ 12], of which the six peripheral ones are chemically and topologically equivalent, whereas that constituting the core (Fe(Cp)(C6Me6)+) has a different chemical nature. Accordingly, two redox processes are observed, i.e., oxidation of the peripheral ferrocene moieties and reduction of the core, whose cyclic voltammetric waves have current intensities in the 6 1 ratio. Clearly, the one-electron process of the core is a convenient internal standard to calibrate the number of electron exchanged in the multi-electron process. In the absence of an internal standard, the number of exchanged electrons has to be obtained by coulometry measurements, or by comparison with the intensity of the wave of an external standard after correction for the different diffusion coefficients [15]. [Pg.208]

The procedure of spin counting is then to use the EPR spectrum of another paramagnetic compound as an external standard (which we will label K to avoid confusion with the spin S) of known concentration (cK) to obtain the unknown (U) concentration (cv) of the paramagnetic compound of interest as... [Pg.97]

Chiral diazaborolidines catalysts in asymmetric reductions have been less described than the corresponding oxaza-borolidines. Although not isolated, the formation of compound 47 has been characterized by nB NMR spectroscopy with the detection of a signal at 24 ppm (from BF3.Et20 as an external standard) <2000TA4329>. [Pg.922]

If the precision offered by the external standard technique is not adequate, it is necessary to use an internal standard. In this procedure, a known amount of a reference substance, not originally present, is introduced into the sample. This will result in the appearance of an additional peak in the chromatogram of the modified sample and any variations in the injections volume will equally affect the standard and the test compounds. [Pg.110]

External standardization is obtained by constructing a calibration curve, i.e., from plotting measured intensities versus rising concentration of the target compound. Calibration curves are generally linear over a wide range of concentrations. When concentration approaches the detection limit (Chap. 5.2.3) the graph deviates from... [Pg.479]

In HPLC, a sample is separated into its components based on the interaction and partitioning of the different components of the sample between the liquid mobile phase and the stationary phase. In reversed phase HPLC, water is the primary solvent and a variety of organic solvents and modifiers are employed to change the selectivity of the separation. For ionizable components pH can play an important role in the separation. In addition, column temperature can effect the separation of some compounds. Quantitation of the interested components is achieved via comparison with an internal or external reference standard. Other standardization methods (normalization or 100% standardization) are of less importance in pharmaceutical quality control. External standards are analyzed on separate chromatograms from that of the sample while internal standards are added to the sample and thus appear on the same chromatogram. [Pg.274]

Much is known empirically on B-NMR data (39). Restricting boron compounds to noncyclic molecules with no or one nitrogen atom and two, three, or four carbon atoms bonded to boron and omitting sterically overcrowded alkyl groups R and R, we find four classes with the following typical chemical shift ranges (Et20 BF3 as the external standard) ... [Pg.137]

The products were identified by F NMR spectroscopy. The chemical shifts 8 (using CFCI3 as an external standard) and spin-spin coupling constants J of liquid fluorides are listed in Table I. Compound CF3SF... [Pg.144]

True profile analysis requires scanning over the whole mass range for the acquisition of all data on excreted compounds. Quantitation has been more challenging on a quadrupole instrument because total ion current peaks are seldom a single component and extracted-ion chromatograms (EICs) when recovered from scanned data are of poor quality due to the lower sensitivity of scanning GC-MS. Thus, we developed profile analysis based on SIM of selected analytes but tried to ensure the components of every steroid class of interest were included. For ion traps the fundamental form of data collection (in non-MS/MS mode must be full -scans). Thus, the quantitative data produced are EICs obtained from scanned data. The EICs are of the same ions used for SIM in quadrupole instruments and the calibration external standards are the same. [Pg.569]

Quantification of all compounds is by relating the intensity of specified ions to that of the internal standard stigmasterol [m/z 394 (M-90)]. Instrument calibration was achieved by running the external standard daily. [Pg.573]

In clinical chemistry, interpretation of the data can be quite simple or complex. In the case of MS/MS applications pertaining to a single analyte, all that is needed is the intensity value from the mass of a peak of interest and its internal standard. Viewing of a spectrum is not necessary. For profile methods such as full-scan acylcarni-tines, amino acids, or other compound families, the interpretation is more complex. With multiple related components, calculation of the concentration of many key metabolites is required. The system generally has multiple internal standards, external standards, or both. In addition to the concentration calculations, examination of a profile is often best achieved by viewing the spectra together with the quantitative information. [Pg.799]

In quantitative analyses, the use of the split/splitless injector can cause concentration errors because of the discrimination of compounds that are of different volatility. The composition of the fraction that enters the column can be very different from that which is eliminated. This mode of operation should be excluded when an external standard is used (cf. 4.9). However, this problem can be corrected, up to a point, by use of a proper glass sleeve inserted into the injector. [Pg.27]

Static mode the sample (liquid or solid matrix) is placed in a glass phial capped with a septum such that the sample occupies only part of the phial s volume. After thermodynamic equilibrium between the phases has been reached (1/2 to 1 h), a sample of the vapour at equilibrium is taken (Fig. 20.4). Under these conditions, the quantity of each volatile compound present in the headspace above the sample is proportional to its concentration in the matrix. The relationship between the amount of sample injected into the gas chromatograph and its concentration in the matrix can be obtained by calibration (using internal or external standards). [Pg.381]


See other pages where External standard compound is mentioned: [Pg.7]    [Pg.7]    [Pg.211]    [Pg.780]    [Pg.833]    [Pg.496]    [Pg.198]    [Pg.243]    [Pg.515]    [Pg.360]    [Pg.420]    [Pg.179]    [Pg.182]    [Pg.185]    [Pg.186]    [Pg.201]    [Pg.494]    [Pg.300]    [Pg.175]    [Pg.82]    [Pg.78]    [Pg.733]    [Pg.17]    [Pg.99]    [Pg.297]    [Pg.733]   
See also in sourсe #XX -- [ Pg.70 ]




SEARCH



External standardization

External standards

Standards compounds

© 2024 chempedia.info