Pump, diffusion

Operation. In a diffusion pump, the pump fluid is heated so that a vapour pressure of 1-10 mbar is established in the boiler. The vapour rises in the jet assembly where it is expanded through nozzles and enters the space between the nozzle and the cooled wall of the pump at high supersonic velocity. Pumping action is based on the transfer of momentum in collisions between the high speed (several times the speed of sound) pump fluid vapour molecules and particles that have entered the vapour jet. [Pg.78]

A possible problem with diffusion pumps is that of backstreaming whereby pump fluid molecules emerge, particularly from the top jet, in the direction of the chamber, where they can condense. Oil contamination can be significantly reduced by the use of baffles (cold-cap, shell) and traps between the pump and the system. [Pg.78]

At the high pressure end of the S vs pMet curve, (here labelled, overload region also known as the forepressure break-down region), S declines from iSmax until a critical pressure is reached in the backing line (critical backing pressure or forevacuum tolerance) which, if exceeded, causes the pumping action of the diffusion pump to cease. [Pg.79]

The diagram below shows a manually operated, diffusion-pumped high vacuum pump set. [Pg.79]

Note It is usual to have a combined gauge accepting signals from the gauge heads 9 and 10 and automatically switching from 9 to 10 and vice versa. [Pg.80]


A diffusion mechanism is also used in dialysis as a means of separating colloids from crystalloids. The rate of diffusion of molecules in gels is practically the same as in water, indicating the continuous nature of the aqueous phase. The diffusion of gases into a stream of vapour is of considerable importance in diffusion pumps. [Pg.137]

The so-called hydro-vac pump, shown in Fig. 11, 22, 2 (the upper half of the mercury reservoir and the column above it are insulated by a layer of asbestos), is an inexpensive, all-glass, mercury diffusion pump, which can be used in series either with an oil pmnp or with a water Alter pmnp (compare Fig. 11,21, 1) capable of producing a vacuum of at least 2 mm. It is accordingly of particular value in the organic laboratory for vacuum distillations, fractionations, sublimations and pyrolyses as well as for molecular distillations (see Section 11,26). The hydro-vac... [Pg.111]

The metal is widely used in laboratory work for making thermometers, barometers, diffusion pumps, and many other instruments. It is used in making mercury-vapor lamps and advertising... [Pg.81]

Perfluoroalkyl ether greases thickened with polytetrafluoroethylene (MIL-G-38220 and MIL-G-27617) are used from —40 to 200°C in missiles, aircraft, and appHcations where fuel, oil, and Hquid oxygen resistance is needed (55). Polyphenyl ether greases find special use from 10 to 315°C in high vacuum diffusion pumps and for radiation resistance. [Pg.248]

A represents mechanical pump or steam ejector B, booster pump D, cryo, turbomolecular, sorption, ion, or trapped diffusion pumps. [Pg.366]

Diffusion Pump System. After the pump line and trap have been shut off, a large valve is opened slowly enough that the mass flow of gas from the chamber through the valve into the od-diffusion pump system does not dismpt the top jet of the diffusion pump (DP) (Fig. 4). When the Hquid nitrogen is replenished after the trap has been operated for some time, release of previously trapped gas must be avoided. The so-caded ionization-gauge response pips at the start of the Hquid-nitrogen replenishment are an indication of trap ineffectiveness. [Pg.369]

Fig. 4. (a) Multistage diffusion pump (b) insert A, jet spray, where closed circles indicate gas molecules and open circles indicate vapor-jet molecules P. ... [Pg.370]

An optimized relationship is obtained between the beU jar, 60° swing-leaf valve, LN trap, baffle for the oil, and the plane of action for the diffusion pump (DP) top jet. The valve open area equals 0.38 of the cross-sectional area of the inside diameter of the furnace. The volumetric speed factor for water vapor is thus 0.38 x 0.9 crr 0.34, where 0.9 is the Clausing factor. [Pg.377]

Diffusion pumps operate at veiy low pressures. The ultimate vacuum attainable depends somewhat upon the vapor pressure of the pump liquid at the temperature of the condensing surfaces. By providing a cold trap between the diffusion pump and the region being evacuated, pressures as low as 10 mmHg absolute are achieved in... [Pg.936]

Fouling Industrial streams may contain condensable or reactive components which may coat, solvate, fill the free volume, or react with the membrane. Gases compressed by an oil-lubricated compressor may contain oil, or may be at the water dew point. Materials that will coat or harm the membrane must be removed before the gas is treated. Most membranes require removal of compressor oil. The extremely permeable poly(trimethylsilylpropyne) may not become a practical membrane because it loses its permeability rapidly. Part of the problem is pore collapse, but it seems extremely sensitive to contamination even by diffusion pump oil and gaskets [Robeson, op. cit., (1994)]. [Pg.2050]

If the pump is a filter pump off a high-pressure water supply, its performance will be limited by the temperature of the water because the vapour pressure of water at 10°, 15°, 20° and 25° is 9.2, 12.8, 17.5 and 23.8 mm Hg respectively. The pressure can be measured with an ordinary manometer. For vacuums in the range lO" mm Hg to 10 mm Hg, rotary mechanical pumps (oil pumps) are used and the pressure can be measured with a Vacustat McLeod type gauge. If still higher vacuums are required, for example for high vacuum sublimations, a mercury diffusion pump is suitable. Such a pump can provide a vacuum up to 10" mm Hg. For better efficiencies, the pump can be backed up by a mechanical pump. In all cases, the mercury pump is connected to the distillation apparatus through several traps to remove mercury vapours. These traps may operate by chemical action, for example the use of sodium hydroxide pellets to react with acids, or by condensation, in which case empty tubes cooled in solid carbon dioxide-ethanol or liquid nitrogen (contained in wide-mouthed Dewar flasks) are used. [Pg.12]

Sodium borate (borax) [1330-43-4] M 201.2, m 741 , d 2.37. Most of the water of hydration was removed from the decahydrate by evacuation at 25° for three days, followed by heating to 100° and evacuation with a high-speed diffusion pump. The dried sample was then heated gradually to fusion (above 966°), allowed to cool gradually to 200°, then tranferred to a desiccator containing P2O5 [Grenier and Westrum J Am Chem Soc 78 6226 1956]. [Pg.466]

Two vacuum systems are used to provide both the high vacuum needed for the mass spectrometer and the differential pumping required for the interface region. Rotary pumps are used for the interface region. The high vacuum is obtained using diffusion pumps, cryogenic pumps, or turbo pumps. [Pg.626]

In addition to the fluoroplastics and fluororubbers already described other fluoropolymers have been marketed. Polymers of hexafluoropropylene oxide are marketed by Du Pont (Krytox). These materials have a low molecular weight (2000-7000) and are either oils or greases. The oils are uses as lubricants, heat transfer fluids and non-flammable oils for diffusion pumps. The greases are also used as lubricants. They have good heat and oil resistance but it is said that explosions may result from contact with the surfaces aluminium or magnesium cuttings. [Pg.384]

With the rotary and diffusion pumps in tandem, aided by a liquid-nitrogen trap, a vacuum of 10 Torr became readily attainable between the wars by degrees, as oils and vacuum greases improved, this was inched up towards 10 Torr (a hundred-billionth of atmospheric pressure), but there it stuck. These low pressures were beyond the range of the McLeod gauge and even beyond the Pirani gauge based on heat conduction from a hot filament (limit Torr), and it was necessary to... [Pg.405]

Centrifugal pumps with diffusion vanes are called diffusion pumps or, more recently, vertical turbine pumps. Those pumps with volute casings are called volute pumps. [Pg.473]

For very low pressures, a diffusion pump is used with a rotary pump as the first stage. The principle of operation is that the gas diffuses into a stream of oil or mercury and is driven out of the pump by molecular bombardment. [Pg.367]

The apparatuses used for the studies of both ammonia synthesis emd hydrodesulfurization were almost identical, consisting of a UHV chamber pumped by both ion and oil diffusion pumps to base pressures of 1 x10 " Torr. Each chamber was equipped with Low Energy Electron Diffraction optics used to determine the orientation of the surfaces and to ascertain that the surfaces were indeed well-ordered. The LEED optics doubled as retarding field analyzers used for Auger Electron Spectroscopy. In addition, each chamber was equipped with a UTI 100C quadrupole mass spectrometer used for analysis of background gases and for Thermal Desorption Spectroscopy studies. [Pg.155]

During the last year we have built an FTMS instrument specifically designed for laser-induced thermal desorption from single-crystal surfaces. Figure 5 is a perspective drawing of the Instrument. The chamber is pumped by a 150 1/s ion pump and has a base pressure of 2.0 X 10- torr. Gases are Introduced through sapphire-sealed leak valves from a diffusion pumped gas manifold. [Pg.243]

The prepared flask is supported so that the tube B can be sealed to a vacuum system. If a suitable furnace is available the flask is surrounded by this, with the constriction A still within reach of the hand torch flame otherwise the heating must be done by flame. A liquid air trap is included between the flask and diffusion pump. With a vacuuih of below 10 mm of mercury the temperature of the flask is raised to near the softening point—the heating may be rapid... [Pg.67]

Mercury diffusion pumps are standard pieces of laboratory equipment and, for most cases, quite adequate ones can be constructed from glass. In general, Pyrex glass (or similar glass) is necessary because hot mercury vapour circulates and differential strain is... [Pg.77]

Very Simple Glass Mercury Diffusion Pump... [Pg.77]

HREELS experiments [66] were performed in a UHV chamber. The chamber was pre-evacuated by polyphenylether-oil diffusion pump the base pressure reached 2 x 10 Torr. The HREELS spectrometer consisted of a double-pass electrostatic cylindrical-deflector-type monochromator and the same type of analyzer. The energy resolution of the spectrometer is 4-6 meV (32-48 cm ). A sample was transferred from the ICP growth chamber to the HREELS chamber in the atmosphere. It was clipped by a small tantalum plate, which was suspended by tantalum wires. The sample was radia-tively heated in vacuum by a tungsten filament placed at the rear. The sample temperature was measured by an infrared (A = 2.0 yum) optical pyrometer. All HREELS measurements were taken at room temperature. The electron incident and detection angles were each 72° to the surface normal. The primary electron energy was 15 eV. [Pg.6]


See other pages where Pump, diffusion is mentioned: [Pg.137]    [Pg.122]    [Pg.270]    [Pg.381]    [Pg.370]    [Pg.374]    [Pg.375]    [Pg.375]    [Pg.375]    [Pg.377]    [Pg.936]    [Pg.158]    [Pg.405]    [Pg.147]    [Pg.352]    [Pg.28]    [Pg.35]    [Pg.28]    [Pg.77]    [Pg.78]    [Pg.88]    [Pg.90]    [Pg.172]    [Pg.172]    [Pg.174]    [Pg.112]   
See also in sourсe #XX -- [ Pg.1663 ]

See also in sourсe #XX -- [ Pg.56 , Pg.57 , Pg.76 , Pg.78 ]

See also in sourсe #XX -- [ Pg.365 , Pg.366 , Pg.367 , Pg.368 , Pg.369 , Pg.370 , Pg.371 , Pg.372 , Pg.373 , Pg.374 , Pg.375 , Pg.376 , Pg.377 , Pg.378 , Pg.379 , Pg.380 , Pg.381 , Pg.382 , Pg.383 ]

See also in sourсe #XX -- [ Pg.590 , Pg.591 ]

See also in sourсe #XX -- [ Pg.78 , Pg.529 ]

See also in sourсe #XX -- [ Pg.367 ]

See also in sourсe #XX -- [ Pg.330 , Pg.331 ]




SEARCH



Centrifugal pumps diffuser type

Diffusion barriers, pumping

Diffusion barriers, pumping cells

Diffusion pumps cleaning

Diffusion pumps draining

Diffusion pumps glass

Diffusion pumps limitations

Diffusion pumps maintenance

Diffusion pumps metal pump

Diffusion pumps pumping speed

Diffusion pumps safety

Diffusion pumps systems

High vacuum pumps diffusion pump

How to Use a Diffusion Pump

Mercury diffusion pump

Mixing pump diffusion system

Oil diffusion pump

Operating errors with diffusion and vapor-jet pumps

Pump, cryogenic diffusion

Pump, diffusion Quadrupole

Pump, diffusion Quantum

Pump, diffusion helium)

Pump, diffusion magnetic)

Pump, diffusion principal

Pump, diffusion radial

Pumping cells with different diffusion

Pumping diffusion

Pumping diffusion

Silicone oil diffusion pump

Ultimate pressure diffusion pumps

Vacuum distillation diffusion pumps

Vacuum oil diffusion pump

Vacuum pump diffusion

Vacuum pump, compression diffusion

Vacuum systems diffusion pump

Vapour diffusion pumps

© 2024 chempedia.info