Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Differential thermocouple 65 temperature

The construction of DTA apparatus is simple and consists of a furnace, differential thermocouple, temperature thermocouple, specimen holders, temperature programmer and recorder. The schematic of a typical DTA apparatus is shown in Figure 3.5. [Pg.183]

In differential thermal aned rsls, l.e.- DTA, we use one thermocouple Taucked" against the voltage output of another of the same composition to produce a "net" EMF. What this means Is that either the positive (or negative) legs of both thermocouples are electrically connected so that the net EMF at any given temperature of the two Is zero. Only If one thermocouple temperature differs from that of the other does one obtain an EMF response. [Pg.361]

The first thing to note is that the furnace surrounds the sample-holder containing the differential thermocouples. A separate control thermocouple controls the furnace temperature and should be placed as close as possible to the position of the sample holder. Some commercial manufacturers use the Reference leg of the differential thermocouple to control the temperature. However, if you were to build a DTA using the components as shown in 7.1.14,... [Pg.364]

Figure 3.1 is a schematic of the differential thermal analyzer (DTA) design. The device measures the difference in temperature between a sample and reference which are exposed to the same heating schedule via symmetric placement with respect to the furnace. The reference material is any substance, with about the same thermal mass as the sample, which undergoes no transformations in the temperature range of interest. The temperature difference between sample and reference is measured by a differential thermocouple in which one junction is in contact with the underside of the sample crucible, and the other is in contact with the underside of the reference crucible.1 The sample temperature is measured via the voltage across the appropriate screw terminals (Vt,) and similarly for the reference temperature (Vrr) generally only one or the other is recorded (see section 3.5.1). Sample and reference... [Pg.35]

The type of thermocouple used for the differential measurement is critical since the measured temperature differences can be quite small. If measurements are to be made only to 1200°C, for example, a type K thermocouple could be used, which has approximately five times the output of platinum thermocouples such as types 5 or R. Type K thermocouples, however, tend to oxidize rapidly over repeated cycles, and hence need to be replaced more often than platinum-based thermocouples. The most critical component of a good DTA is the differential thermocouple signal amplifier, which must amplify minute voltages while eliminating random noise. [Pg.35]

Generally the y-axis is left as an amplification of differential thermocouple voltage, showing exothermic and endothermic trends, where no effort is made to convert the abscissa values into temperatures. [Pg.36]

For reactions of minute thermal effect, e.g. second order transitions, it is advantageous to use as much sample mass as feasible in the heat-flux DSC sample pan. It is advisable to use an adequate thermal mass of reference powder to match that of the sample. This has the advantage of not only minimizing baseline float, but also smooths out what may appear to be signal noise When the reference lacks thermal mass, its temperature will vary responsively to random thermal fluctuations in its surroundings. On a sensitive scale, the changing reference temperature will be manifested as noise on the amplified differential thermocouple signal. [Pg.75]

Figure 3.27 Calculation of heat capacity of an unknown using a Netzsch DSC200 heat-flux DSC [7]. The distinct shift in heat capacity at 690°C corresponds to the glass transition temperature (see section 7.6). A 191 mg sapphire standard was used as calibrant for a 130 mg (laser special) glass sample. All heating ramps were at 20°C/min (faster heating rates permit greater temperature lags). The right hand scale, in the original units of the differential thermocouple, is inverted in exothermic and endothermic directions as compared to the usual convention in this book. Figure 3.27 Calculation of heat capacity of an unknown using a Netzsch DSC200 heat-flux DSC [7]. The distinct shift in heat capacity at 690°C corresponds to the glass transition temperature (see section 7.6). A 191 mg sapphire standard was used as calibrant for a 130 mg (laser special) glass sample. All heating ramps were at 20°C/min (faster heating rates permit greater temperature lags). The right hand scale, in the original units of the differential thermocouple, is inverted in exothermic and endothermic directions as compared to the usual convention in this book.
Two thermocouples separated by distance L are imbedded in the test specimen, one directly above the other, whereby the temperature drop T2 — T between them is measured. A differential thermocouple measures the temperature rise ATw of the exit water of the calorimeter as compared to its entrance temperature. The mass flow rate of water F into the calorimeter is monitored, so that over a specific time interval At, the total heat absorbed by the calorimeter may be calculated, knowing the specific heat cp of water. Dividing by the time interval will give the rate of heat flow into the calorimeter under steady state conditions ... [Pg.232]

Here, the heat evolved on adsorption increases the temperature of the sample and its container (usually a copper cylinder). The heat is prevented from flowing to the peripheral shield (the surroundings ) by an appropriate control of the shield temperature. Thus, the shield is usually maintained at the same temperature as the sample container by the use of a differential thermocouple and a heat coil - as indicated in Figure 3.14. The temperature rise is measured by means of a resistance thermometer attached to the sample container. [Pg.63]

The PID-SCR temperature control technique is adopted for the adiabatic control. Besides, other devices are also adopted to attain as complete an adiabatic control as possible. To cite an instance, a pre-amplifier is incorporated before the PID controller to amplify the A i.e., the temperature difference between the temperature of 2 cm of a chemical of the TD type, including every gas-penneable oxidatively-heating substance, and the T),. A zero suppression circuit is composed of this amplifier to cancel the slight stray-, or pseudo-, Ihermoelcctromotive force of the differential thermocouple. Such a pscudo-ihcmioclcctromolivc force of a differential thermocouple may still appear even if the temperature of 2 cm of the chemical and the r , . are physically the same, and even if the two thermocouples to make up the differential thermocouple are... [Pg.65]

A procedure to cancel electrically the slight pseudo-thermoelectromotive force of the differential thermocouple, which is composed of the thermocouple to measure the temperature of the reference material, or the temperature of the chemical tested, and that to measure the T by operating the zero-suppression circuit. The concrete procedure is to set the indicator of the analog D.C. microvoltmeter to the graduation line of zero at the center of the scale span of the meter. [Pg.89]

Aluminium cylinders in vented furnace The surface heat-transfer coefficients, h, for the aluminium containers in the stirred air of the working space in the vented furnace were estimated from measurements of the heat-transfer coefficients for solid aluminium cylinders of similar dimensions and surface finish. The cylinders were heated electrically at a known power input by small heaters in central cavities (0.6 cm diameter) and the steady-state temperature difference between the cylinder surface and the air in the working space was measured by means of a differential thermocouple. Measurements were made on two sizes of cylinder with a length to diameter ratio of 1.7, and heat-transfer coefficients for other sizes were estimated by fitting the following equation for heat and mass transfer from small spheres, due to Ranz and Marshall [1952], to the observations ... [Pg.155]

The still pot is then heated so that the liquid boils gently, and a steady reflux is returned from the head of the column. The jacket temperature is adjusted to correspond to the vapor temperature at the head if a thermometer is used, or it is adjusted to adiabatic operation if a differential thermocouple is employed. The boil-up rate (throughput) is adjusted to a value which is appropriate for the column being used (Table 1-12) by regulating the amount of heat supplied to the still pot. The column is allowed to achieve equilibrium before any material is withdrawn. This is usually determined by the constancy of the vapor temperature or of the refractive index of the material at the column head and usually requires from one-half to several hours. The time necessary for establishing equilibrium is usually greatest for the columns with the highest number of theoretical plates. [Pg.52]

Considerable supercoolings are realized in small liquid drops. Water drops from 500 to 20 pm in diameter in oil were located on the junction of a differential thermocouple. Every drop was melted down and crystallized several tens of times. Measurements at the same temperature were made on 5-10 drops similar in size. The distribution of crystallization events of isolated drops was studied in repeated experiments under isothermal conditions and continuous supercooling. ... [Pg.260]

Figure 6.36. DSC cell by David (110). 1. thermocouple for x axis or system temperature readout 2. limit switch thermocouple 3. programming or furnace thermocouple 4. dynamic gas port entry 5. dynamic gas port exit 6, sample side of differential thermocouple 7, reference side of differential thermocouple 8, ceramic thermal insulator. 9, ceramic support rods 10. sample pans. Figure 6.36. DSC cell by David (110). 1. thermocouple for x axis or system temperature readout 2. limit switch thermocouple 3. programming or furnace thermocouple 4. dynamic gas port entry 5. dynamic gas port exit 6, sample side of differential thermocouple 7, reference side of differential thermocouple 8, ceramic thermal insulator. 9, ceramic support rods 10. sample pans.
Identification of the apparatus, including the geometry and materials of the thermocouples and the locations of the differential and temperature-measuring thermocouples. [Pg.800]

The differential thermocouple or AT thermocouple is the thermocouple system used to measure temperature difference. Should another thermosensing device be used, its name should replace thermocouple. ... [Pg.806]

The relationship between oxygen concentration and temperature suggests that a stack can be kept nonflammable by temperature control. The steam rate to the stack would be controlled by a differential thermocouple with one junction in the isothermal zone and the other in the declining temperature zone. The latter junction could be placed at the elevation below which non-flammable conditions are to be maintained. [Pg.186]

Required steam rates for safe purging are governed by heat losses, A rate of 7.5 pounds per minute per foot of diameter appears adequate for any weather. However, stack oxygen content is related to stack wall temperature. Thus, a more economical and safer operation would be one in which rate is automatically controlled by a differential thermocouple. [Pg.187]

The equipment used in DTA studies is shown schematically in Fig. 16.16. The sample is loaded into a crucible, which is then inserted into the sample well (marked S). A reference sample is made by placing a similar quantity of inert material (such as AI2O3) in a second crucible. This crucible is inserted in the reference well, marked R. The dimensions of the two crucibles and of the cell wells are as nearly identical as possible furthermore, the weights of the sample and the reference should be virtually equal. The sample and reference should be matched thermally and arranged symmetrically with the furnace so that they are both heated or cooled in an identical manner. The metal block surrounding the wells acts as a heat sink. The temperature of the heat sink is slowly increased using an internal heater. The sink in turn simultaneously heats the sample and reference material. A pair of matched thermocouples is used. One pair is in contact with the sample or the sample container (as shown) the other pair is in contact with the reference. The output of the differential thermocouple, or AT, is amplified and sent to the data... [Pg.1021]

The temperature distribution within the cryostat was measured with six 20-gauge copper-constantan thermocouples located throughout the cryostat, these thermocouples being referenced to the resistance thermometer. Thermocouple emf s were measured with an L N Type K-3 potentiometer in conjunction with a Model 9834 electronic dc null detector. Inasmuch as these differential thermocouples generate small emf s, their accuracy is limited by the potentiometer, this accuracy being 0.02°C in this investigation. During operation the temperature distribution in the bath liquid was within 0.1°C of the resistance thermometer. [Pg.199]

The differential scanning calorimeter evolved from an older instrument known as a differential thermal analyzer, or DTA. The DTA, which is based on the work of Le Chatelier in 1887, was developed in 1899 for identification of specific types of clays, which are difficult to differentiate by more traditional methods. The concept of the DTA is quite simple. A differential thermocouple, which consists of two otherwise identical thermocouples connected in opposing polarities, is placed in a furnace in a position which allows the bead of one thermocouple to be inserted into an inert reference material, while the bead of the other thermocouple is inserted into the sample. The difference in temperature between the reference and sample materials is obtained directly as a function of temperature as the entire assembly is heated at a controlled, usually linear, rate. In the absence of any thermal difference between the sample and reference material, the output of the differential thermocouple will be zero. When a thermal event occurs, c.g., heat released during crystallization, the change in specific heat at the glass... [Pg.237]

Example 7.1. The thermal conductivity of a 30 cm by 30 cm insulation insulator is measured in a guarded hot plate. The uncertainty of a differential thermocouple measuring the temperature is 0.3 K. The power applied to the sample is 5 kW 1 % and the temperature differential is 55 K across the 2.0 mm thick sample. What is the thermal conductivity of the insulation and what is the measurement uncertainty ... [Pg.234]

The DTA cell shown in Fig. 1 was made from the parts of a Du Pont standard DTA cell. Holes were drilled in the silver heating block to accommodate six 2mm o.d. glass capillary tubes which hold the reference and five samples. The Chromel-Alumel thermocouples in each of the five samples are connected in a circuit to the Chromel-Alumel thermocouple in the reference to provide five differential thermocouple voltages of the samples with respect to the reference. These five differential voltages, as well as the reference thermocouple voltage, are each amplified by one of the six low-level amplifiers in the data system to provide outputs representing AT-, AT, AT, AT, AT- and reference temperature T, which are multi-... [Pg.134]

The experimental cell was used, offering the possibility to take samples of anode gases for chromatographic and chemical analysis by the method described in [33], The anode was a graphite cylinder with a thermocouple positioned in a weU inside for measurement of temperature. Temperature differences were measured with differential thermocouple, with one jtmction being inside the anode and the other junction in the bulk electrolyte. Other experimental details can be found elsewhere [32]. [Pg.120]


See other pages where Differential thermocouple 65 temperature is mentioned: [Pg.198]    [Pg.106]    [Pg.51]    [Pg.85]    [Pg.241]    [Pg.602]    [Pg.403]    [Pg.70]    [Pg.152]    [Pg.51]    [Pg.87]    [Pg.308]    [Pg.319]    [Pg.329]    [Pg.339]    [Pg.349]    [Pg.679]    [Pg.100]    [Pg.375]    [Pg.212]    [Pg.1155]    [Pg.362]    [Pg.840]   
See also in sourсe #XX -- [ Pg.32 ]




SEARCH



Differential thermocouple

Temperature differential

Temperature thermocouples

Thermocouple

Thermocoupling

© 2024 chempedia.info