Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Thermocouple differential

The first thing to note is that the furnace surrounds the sample-holder containing the differential thermocouples. A separate control thermocouple controls the furnace temperature and should be placed as close as possible to the position of the sample holder. Some commercial manufacturers use the Reference leg of the differential thermocouple to control the temperature. However, if you were to build a DTA using the components as shown in 7.1.14,... [Pg.364]

Several techniques are available for thermal conductivity measurements, in the steady state technique a steady state thermal gradient is established with a known heat source and efficient heat sink. Since heat losses accompany this non-equilibrium measurement the thermal gradient is kept small and thus carefully calibrated thermometers and heat source must be used. A differential thermocouple technique and ac methods have been used. Wire connections to the sample can represent a perturbation to the measurement. Techniques with pulsed heat sources (including laser pulses) have been used in these cases the dynamic response interpretation is more complicated. [Pg.656]

The construction of DTA apparatus is simple and consists of a furnace, differential thermocouple, temperature thermocouple, specimen holders, temperature programmer and recorder. The schematic of a typical DTA apparatus is shown in Figure 3.5. [Pg.183]

Figure 3.1 is a schematic of the differential thermal analyzer (DTA) design. The device measures the difference in temperature between a sample and reference which are exposed to the same heating schedule via symmetric placement with respect to the furnace. The reference material is any substance, with about the same thermal mass as the sample, which undergoes no transformations in the temperature range of interest. The temperature difference between sample and reference is measured by a differential thermocouple in which one junction is in contact with the underside of the sample crucible, and the other is in contact with the underside of the reference crucible.1 The sample temperature is measured via the voltage across the appropriate screw terminals (Vt,) and similarly for the reference temperature (Vrr) generally only one or the other is recorded (see section 3.5.1). Sample and reference... [Pg.35]

The type of thermocouple used for the differential measurement is critical since the measured temperature differences can be quite small. If measurements are to be made only to 1200°C, for example, a type K thermocouple could be used, which has approximately five times the output of platinum thermocouples such as types 5 or R. Type K thermocouples, however, tend to oxidize rapidly over repeated cycles, and hence need to be replaced more often than platinum-based thermocouples. The most critical component of a good DTA is the differential thermocouple signal amplifier, which must amplify minute voltages while eliminating random noise. [Pg.35]

Generally the y-axis is left as an amplification of differential thermocouple voltage, showing exothermic and endothermic trends, where no effort is made to convert the abscissa values into temperatures. [Pg.36]

These devices have a disk (e.g. constantan alloy) on which the sample and reference pans rest on symmetrically placed platforms. Thermocouple wire (e g. chromel alloy) Is welded to the underside of each platform. The chromel-constantan junctions make up the differential thermocouple junctions with the constantan disk acting as one leg of the thermocouple pair. [Pg.40]

The DTA design most susceptible to this form of baseline float is the post-type (Figure 3.23a), where care must be taken to ensure that sample and reference are centered in the tube,15 and are of nearly identical total heat capacity. The older nickel (or other refractory metal) block design, shown in Figure 3.23b, where differential thermocouple beads reside directly within... [Pg.72]

For reactions of minute thermal effect, e.g. second order transitions, it is advantageous to use as much sample mass as feasible in the heat-flux DSC sample pan. It is advisable to use an adequate thermal mass of reference powder to match that of the sample. This has the advantage of not only minimizing baseline float, but also smooths out what may appear to be signal noise When the reference lacks thermal mass, its temperature will vary responsively to random thermal fluctuations in its surroundings. On a sensitive scale, the changing reference temperature will be manifested as noise on the amplified differential thermocouple signal. [Pg.75]

Figure 3.27 Calculation of heat capacity of an unknown using a Netzsch DSC200 heat-flux DSC [7]. The distinct shift in heat capacity at 690°C corresponds to the glass transition temperature (see section 7.6). A 191 mg sapphire standard was used as calibrant for a 130 mg (laser special) glass sample. All heating ramps were at 20°C/min (faster heating rates permit greater temperature lags). The right hand scale, in the original units of the differential thermocouple, is inverted in exothermic and endothermic directions as compared to the usual convention in this book. Figure 3.27 Calculation of heat capacity of an unknown using a Netzsch DSC200 heat-flux DSC [7]. The distinct shift in heat capacity at 690°C corresponds to the glass transition temperature (see section 7.6). A 191 mg sapphire standard was used as calibrant for a 130 mg (laser special) glass sample. All heating ramps were at 20°C/min (faster heating rates permit greater temperature lags). The right hand scale, in the original units of the differential thermocouple, is inverted in exothermic and endothermic directions as compared to the usual convention in this book.
Two thermocouples separated by distance L are imbedded in the test specimen, one directly above the other, whereby the temperature drop T2 — T between them is measured. A differential thermocouple measures the temperature rise ATw of the exit water of the calorimeter as compared to its entrance temperature. The mass flow rate of water F into the calorimeter is monitored, so that over a specific time interval At, the total heat absorbed by the calorimeter may be calculated, knowing the specific heat cp of water. Dividing by the time interval will give the rate of heat flow into the calorimeter under steady state conditions ... [Pg.232]

Here, the heat evolved on adsorption increases the temperature of the sample and its container (usually a copper cylinder). The heat is prevented from flowing to the peripheral shield (the surroundings ) by an appropriate control of the shield temperature. Thus, the shield is usually maintained at the same temperature as the sample container by the use of a differential thermocouple and a heat coil - as indicated in Figure 3.14. The temperature rise is measured by means of a resistance thermometer attached to the sample container. [Pg.63]

For differential thermal analysis, two tubes were positioned symmetrically in the nickel block. One tube contained the mixture to be examined by thermal analysis the other was used as a reference. Satisfactory baseline behavior in the record of the differential thermocouple was obtained by operating with the reference tube filled with air at 1 atm. pressure. The voltage indicating the difference between the sample and reference thermocouples was fed into a d.c. amplifier, capable of multiplying the difference signal by factors varying from 2.5 to 100. The amplified signal was displayed by a suitable strip chart recorder. [Pg.310]

The PID-SCR temperature control technique is adopted for the adiabatic control. Besides, other devices are also adopted to attain as complete an adiabatic control as possible. To cite an instance, a pre-amplifier is incorporated before the PID controller to amplify the A i.e., the temperature difference between the temperature of 2 cm of a chemical of the TD type, including every gas-penneable oxidatively-heating substance, and the T),. A zero suppression circuit is composed of this amplifier to cancel the slight stray-, or pseudo-, Ihermoelcctromotive force of the differential thermocouple. Such a pscudo-ihcmioclcctromolivc force of a differential thermocouple may still appear even if the temperature of 2 cm of the chemical and the r , . are physically the same, and even if the two thermocouples to make up the differential thermocouple are... [Pg.65]

A procedure to cancel electrically the slight pseudo-thermoelectromotive force of the differential thermocouple, which is composed of the thermocouple to measure the temperature of the reference material, or the temperature of the chemical tested, and that to measure the T by operating the zero-suppression circuit. The concrete procedure is to set the indicator of the analog D.C. microvoltmeter to the graduation line of zero at the center of the scale span of the meter. [Pg.89]

Aluminium cylinders in vented furnace The surface heat-transfer coefficients, h, for the aluminium containers in the stirred air of the working space in the vented furnace were estimated from measurements of the heat-transfer coefficients for solid aluminium cylinders of similar dimensions and surface finish. The cylinders were heated electrically at a known power input by small heaters in central cavities (0.6 cm diameter) and the steady-state temperature difference between the cylinder surface and the air in the working space was measured by means of a differential thermocouple. Measurements were made on two sizes of cylinder with a length to diameter ratio of 1.7, and heat-transfer coefficients for other sizes were estimated by fitting the following equation for heat and mass transfer from small spheres, due to Ranz and Marshall [1952], to the observations ... [Pg.155]

The still pot is then heated so that the liquid boils gently, and a steady reflux is returned from the head of the column. The jacket temperature is adjusted to correspond to the vapor temperature at the head if a thermometer is used, or it is adjusted to adiabatic operation if a differential thermocouple is employed. The boil-up rate (throughput) is adjusted to a value which is appropriate for the column being used (Table 1-12) by regulating the amount of heat supplied to the still pot. The column is allowed to achieve equilibrium before any material is withdrawn. This is usually determined by the constancy of the vapor temperature or of the refractive index of the material at the column head and usually requires from one-half to several hours. The time necessary for establishing equilibrium is usually greatest for the columns with the highest number of theoretical plates. [Pg.52]

Considerable supercoolings are realized in small liquid drops. Water drops from 500 to 20 pm in diameter in oil were located on the junction of a differential thermocouple. Every drop was melted down and crystallized several tens of times. Measurements at the same temperature were made on 5-10 drops similar in size. The distribution of crystallization events of isolated drops was studied in repeated experiments under isothermal conditions and continuous supercooling. ... [Pg.260]

The heat transfer problem for a DTA system containing ring thermocouples has been treated by David (56,57). In order to obtain a mathematical expression for Cp, the heat capacity of the sample (or reference), one must consider two factors (1) the effects of the system on the differential thermocouple and (2) the effects of the system plus sample on the differential thermocouple. The heat capacity of the sample holder containing a sample which is undergoing exothermic or endothermic changes can be expressed as... [Pg.220]

A new DSC cell, based on the DTA principle (as is the Du Pont DSC cell previously discussed) has been described by David (HO). The calorimeter cell, as shown in Figure 6.36, contains a differential thermocouple of a new thin-form design that is isolated from the cell wall and bottom to provide greater sensitivity. This thermocouple consists of a sheet of negative Platinel II type thermocouple alloy coupled to a positive Platinel II alloy. Flat shallow containers are employed for the sample and reference materials. Two addi-... [Pg.342]

Figure 6.36. DSC cell by David (110). 1. thermocouple for x axis or system temperature readout 2. limit switch thermocouple 3. programming or furnace thermocouple 4. dynamic gas port entry 5. dynamic gas port exit 6, sample side of differential thermocouple 7, reference side of differential thermocouple 8, ceramic thermal insulator. 9, ceramic support rods 10. sample pans. Figure 6.36. DSC cell by David (110). 1. thermocouple for x axis or system temperature readout 2. limit switch thermocouple 3. programming or furnace thermocouple 4. dynamic gas port entry 5. dynamic gas port exit 6, sample side of differential thermocouple 7, reference side of differential thermocouple 8, ceramic thermal insulator. 9, ceramic support rods 10. sample pans.

See other pages where Thermocouple differential is mentioned: [Pg.765]    [Pg.375]    [Pg.198]    [Pg.106]    [Pg.62]    [Pg.62]    [Pg.40]    [Pg.51]    [Pg.85]    [Pg.241]    [Pg.64]    [Pg.602]    [Pg.62]    [Pg.62]    [Pg.589]    [Pg.403]    [Pg.70]    [Pg.152]    [Pg.51]    [Pg.8]    [Pg.937]    [Pg.937]    [Pg.87]    [Pg.308]    [Pg.319]    [Pg.329]    [Pg.339]    [Pg.349]   
See also in sourсe #XX -- [ Pg.35 ]




SEARCH



Differential thermocouple 65 temperature

Thermocouple

Thermocoupling

© 2024 chempedia.info