Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Detectors diffusion chamber

This diffusion chamber was modified to provide a uniform flow from two channels at the entrance, one for the filtered room air and the other for the gas from the radon chamber. This modified mobility analyzer is schematically shown in Figure 2. The pressure heads are adjusted so that the gas velocities, v, are the same in both channels. An adjustable vertical electric field, E, is provided through the analyzer so that charged particles are drawn toward the detector located at x cm from the entrance. With the known distance, d, between the radon-laden gas channel and the detector implanted plate, the mobility can then be determined from... [Pg.363]

Schematic representation of the experimental setup is shown in Fig 1.1. The electrochemical system is coupled on-line to a Quadrupole Mass Spectrometer (Balzers QMS 311 or QMG 112). Volatile substances diffusing through the PTFE membrane enter into a first chamber where a pressure between 10 1 and 10 2 mbar is maintained by means of a turbomolecular pump. In this chamber most of the gases entering in the MS (mainly solvent molecules) are eliminated, a minor part enters in a second chamber where the analyzer is placed. A second turbo molecular pump evacuates this chamber promptly and the pressure can be controlled by changing the aperture between both chambers. Depending on the type of detector used (see below) pressures in the range 10 4-10 5 mbar, (for Faraday Collector, FC), or 10 7-10 9 mbar (for Secondary Electrton Multiplier, SEM) may be established. Schematic representation of the experimental setup is shown in Fig 1.1. The electrochemical system is coupled on-line to a Quadrupole Mass Spectrometer (Balzers QMS 311 or QMG 112). Volatile substances diffusing through the PTFE membrane enter into a first chamber where a pressure between 10 1 and 10 2 mbar is maintained by means of a turbomolecular pump. In this chamber most of the gases entering in the MS (mainly solvent molecules) are eliminated, a minor part enters in a second chamber where the analyzer is placed. A second turbo molecular pump evacuates this chamber promptly and the pressure can be controlled by changing the aperture between both chambers. Depending on the type of detector used (see below) pressures in the range 10 4-10 5 mbar, (for Faraday Collector, FC), or 10 7-10 9 mbar (for Secondary Electrton Multiplier, SEM) may be established.
The concentration of 222Rn in air was determined with a radon measurement detector. The detector allows realizing continuous radon monitoring. It consists of an electronic unit and a scintillation cell. The electronic unit contains power supply, amplifier, discriminator, timer, counter, and indicator. The scintillation cell contains the zinc sulfide scintillator, photomultiplier, preamplifier, high voltage power supply and chamber with a volume of 200 mL over the scintillator. This chamber is filled with the gas to be analyzed. The air is either pumped or diffuses into the scintillation cell. The scintillation count is processed by electronics, and radon concentrations for predetermined intervals are stored in the memory of the device. [Pg.221]

Diffuse reflectance FTIR spectra of the ground Mo03/Al203 catalysts were recorded on an FTIR instrument (Nicolet, Model 740, MCT detector). The microreactor in the flow system was replaced by an FTIR cell. The cell used a Harrick diffuse reflectance accessory (DRA-2CO) fitted with a controlled environmental chamber (HVC-DRP). Spectra (500 scans, 4 cm 1 resolution) were presented in Kubelka-Munk units and recorded at RT. [Pg.455]

The key part of the system is the multiport valve, which interconnects the different parts and solutions used by the system. The common port is connected to a reversible pump with the retention coil placed in between. The pump is connected to the carrier solution reservoir. The common port can access any of the other ports, which lead to sample, standard solutions or reagents, mixing chamber and sensor array, by electrical rotation of the valve. Since the system is bidirectional, volumes can not only be propelled directly to the detector, but also be injected into the retention coil, therefore merging accurate aliquots of different solutions. In order to assure proper mixing of the solutions not only via diffusion... [Pg.745]

The chamber in Example 4.11 has a volume of 5 m3. The volumes of the diffusion pump and the connecting line to the chamber are 100L and 660 L, respectively. The volume of the connection between the Roots and diffusion pumps is 10 L. Calculate the response time for 95% of the final signal on the leak detector when it is connected in positions 1 and 2. [Pg.129]

Two types of track-etch monitor occur, open and closed types. In the open type, the SSNTD is not contained in a volume and is exposed to the air as a bare foil. This detector will register the alpha radiation from the Rn and RnD in the air, and the track density on the foil represents the sum of these activities. However, the Rn signal will be much larger than the signal from the RnD, except at very high levels of RnD (high F factor), and the track density has to be interpreted in terms of this ratio, which is typically unknown. In close monitors the SSNTD is enclosed in a closed container into which Rn diffuses through a filter. This prevents the entry of RnD and dust particles into the chamber, and the foil is then sensitive only to the alpha radiation from Rn and RnD formed in the container. There is a repeatable equilibrium between the isotopes in the container, and calibration provides the relationship between the Rn concentration and the track density on the foil. A typical track-etch radon monitor of the closed type is shown in Fig. 9.27. [Pg.446]

The qualifiers continuous and discrete as applied to pervaporation refer to different aspects of the process. In fact, analytical pervaporation is a continuous technique because, while the sample is in the separation module, mass transfer between the phases is continuous until equilibrium is reached. Continuous also refers to the way the sample is inserted into the dynamic manifold for transfer to the pervaporator. When the samples to be treated are liquids or slurries, the overall manifold to be used is one such as that of Fig. 4.18 (dashed lines included). The sample can be continuously aspirated and mixed with the reagent(s) if required (continuous sample insertion). Discrete sample insertion is done by injecting a liquid sample, either via an injection valve in the manifold (and followed by transfer to the pervaporator) or by using a syringe furnished with a hypodermic needle [directly into the lower (donor) chamber of the separation module when no dynamic manifold is connected to the lower chamber]. In any case, the sample reaches the lower chamber and the volatile analyte (or its reaction product) evaporates, diffuses across the membrane and is accepted in the upper chamber by a dynamic or static fluid that drives it continuously or intermittently, respectively, to the detector — except when separation and detection are integrated. [Pg.134]

Besides the stable ions, metastable ions are also known to be formed in the ionization process. These ions have a short half-life. Some start dissociating in the ionization chamber, but a significant number will still be able to leave it, continuing to dissociate during their passage to the detector. They generate diffuse peaks in the mass spectra. [Pg.133]

Diffuse reflectance infrared (DRIFT) spectra of pyridine adsorbed on the zeolite samples were obtained with a Nicolet Protege 460 ESP spectrometer, equipped with a controlled-temperature and environment diffuse reflectance chamber (Spectra-Tech) with KBr windows and a liquid nitrogen-cooled HgCdTe detector. All spectra were collected in the range of 4000-1000 cm averaging 400 scans at an instrumental resolution of 1 cm. ... [Pg.464]

Fig. 18. Diagram of reactive scattering apparatus for the study of non-metal reactions A, scattering chamber B, source chambers C, liquid nitrogen cooled cold shield D, detector E, source bulkheads G, liquid nitrogen trap H, oil diffusion pumps N, free radical source P, nozzle source Q, skimmer E, ion source H, liquid He trap I, ion lenses P, photomultiplier Q, quadrupole rods R, light baffle S, slide valve T, radial electric field pumps (from C. F. Carter et al. 02 by permission of the Chemical... Fig. 18. Diagram of reactive scattering apparatus for the study of non-metal reactions A, scattering chamber B, source chambers C, liquid nitrogen cooled cold shield D, detector E, source bulkheads G, liquid nitrogen trap H, oil diffusion pumps N, free radical source P, nozzle source Q, skimmer E, ion source H, liquid He trap I, ion lenses P, photomultiplier Q, quadrupole rods R, light baffle S, slide valve T, radial electric field pumps (from C. F. Carter et al. 02 by permission of the Chemical...
The presence of artefacts in the analytical path, such as mixing chambers, tubing connections, de-bubblers and other chamber-like components, can also affect sample dispersion in flow injection analysis. The effects of a mixing chamber and the detector inner volume are discussed in 3.1.2.2 and 6.3.2, respectively. The presence of devices for liquid—liquid extraction and gas diffusion (or dialysis) alters dispersion, and is dealt with in Chapter 8. [Pg.174]


See other pages where Detectors diffusion chamber is mentioned: [Pg.176]    [Pg.242]    [Pg.199]    [Pg.131]    [Pg.239]    [Pg.299]    [Pg.148]    [Pg.650]    [Pg.992]    [Pg.363]    [Pg.42]    [Pg.197]    [Pg.22]    [Pg.250]    [Pg.10]    [Pg.199]    [Pg.79]    [Pg.19]    [Pg.500]    [Pg.246]    [Pg.292]    [Pg.233]    [Pg.91]    [Pg.92]    [Pg.105]    [Pg.355]    [Pg.607]    [Pg.249]    [Pg.52]    [Pg.70]    [Pg.113]    [Pg.244]    [Pg.283]    [Pg.610]    [Pg.218]   
See also in sourсe #XX -- [ Pg.172 ]




SEARCH



Diffusion chamber

Diffusion detectors

© 2024 chempedia.info