Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Detector accuracy

Apart from the qualification dossiers provided by vendors there seems, at present, to be very little information published on the performance of an operational qualification for capillary electrophoresis (CE) instruments other than a chapter in Analytical Method Validation and Instrument Performance. The chapter, written by Nichole E. Baryla of Eli Lilly Canada, Inc, discusses the various functions (injection, separation, and detection) within the instrument and provides guidance on the type of tests, including suggested acceptance criteria, that may be performed to ensure the correct working of the instrument. These include injection reproducibility and linearity, temperature and voltage stability, detector accuracy, linearity, and noise. [Pg.171]

Accuracy is a measure of the closeness of a measurement to the true value. Precision is a measure of how reproducible the measurements are. For many detectors, the accuracy of a measurement is maintained by user calibration. For some detectors, however, such as photodiode array detectors, accuracy relies on internal calibration. The linear dynamic range of the detector is the maximum linear response, divided by the detector noise. The detector response is said to be linear if the difference in response for two concentrations of a given compound is proportional to the difference in concentration between the two samples. Most detectors become nonlinear as the sample concentration increases. [Pg.91]

Detector Accuracy Test—This test method assumes that the FID response approximates the theoretical unit carbon response. To verify this assumption, analyze the performance mixture and calculate the response factors, relative to hexadecane, (RRF for each of the components in the performance mix, using Ae following equations ... [Pg.808]

The following procedure has been found to be useflil for verifying detector linearity. However, the size of the n-C diluent peak tends to exceed the linear range of the FID. Should this occur, the detector accuracy test (8.2.3) provides one indication of linear performance. [Pg.808]

The INTROS Flaw Detector is able to inspect ropes moving through the magnetic head at speed 0...2 m/s. Limit of sensitivity to wire brake is 1% of the rope meatallic cross-section area, the LMA measure accuracy is not less than 2%. [Pg.337]

Other limitation for the spatial resolution can be found in the detector. A limited number of pixels in the camera array can be a reason for pure resolution in the case of a big field of view. For example, if field of view should be 10 by 10 nun with camera division 512x512 pixels the pixel size will be approximately 20 microns. To improve the relation of the field of view and the spatial resolution a mega-pixel sensor can be used. One more limitation for the spatial resolution is in mechanical movement (rotation) of the object, camera and source. In the case of a mechanical movement all displacements and rotations should be done with accuracy better than the spatial resolution in any tested place of the object. In the case of big-size assemblies and PCB s it is difficult to avoid vibrations, axle play and object non-planarity during testing. [Pg.570]

In case of some samples besides the cross sectional CT-slice also a projectional image is of interest. In these cases the test mode Digital Radiography (DR) is applied. In the DR-mode the object is not turned, but scanned horizontally and vertically. Again the very high dynamic of the detector and the mechanical accuracy of the complete system are of large benefit to the image quality. [Pg.586]

Measurements have been made in a static laboratory set-up. A simulation model for generating supplementary data has been developed and verified. A statistical data treatment method has been applied to estimate tracer concentration from detector measurements. Accuracy in parameter estimation in the range of 5-10% has been obtained. [Pg.1057]

The first detector for optical spectroscopy was the human eye, which, of course, is limited both by its accuracy and its limited sensitivity to electromagnetic radiation. Modern detectors use a sensitive transducer to convert a signal consisting of photons into an easily measured electrical signal. Ideally the detector s signal, S, should be a linear function of the electromagnetic radiation s power, P,... [Pg.379]

TOF mass spectrometers are very robust and usable with a wide variety of ion sources and inlet systems. Having only simple electrostatic and no magnetic fields, their construction, maintenance, and calibration are usually straightforward. There is no upper theoretical mass limitation all ions can be made to proceed from source to detector. In practice, there is a mass limitation in that it becomes increasingly difficult to discriminate between times of arrival at the detector as the m/z value becomes large. This effect, coupled with the spread in arrival times for any one m/z value, means that discrimination between unit masses becomes difficult at about m/z 3000. At m/z 50,000, overlap of 50 mass units is more typical i.e., mass accuracy is no better than about 50-100 mass... [Pg.191]

Any one bin can be electronically distinguished from the next one, and therefore the bins can be used like the tick of a standard clock. Each bin serves as one tick, which lasts for only 0.3 nsec. By counting the ticks and knowing into which bin the ion pulse has gone, the time taken for the ion to arrive at the detector can be measured to an accuracy of 0.3 nsec, which is the basis for measuring very short ion arrival times after the ions have traveled along the TOE analyzer tube. Each ion arrival pulse (event) is extracted from its time bin and stored in an associated computer memory location. [Pg.222]

For higher accuracy, a method involving ampHtude modulation of a continuous laser beam is used. Again, a detector receives light reflected from the object where the distance is to be measured. The phases of the modulation in the outgoing beam and in the reflected return are compared. For a total phase shift A( ) between the two signals, the range R is... [Pg.14]

Table 15 shows data for several radionucHdes the y-rays of which are often used to caHbrate the efficiency of y-ray detectors. For a number of these y-rays the very high accuracy arises because the y-ray occurs in essentially 100% of the decays of the nucHde, and only small corrections ate needed to deduce the y-emission probabiHty. In other cases the accuracy is high because a number of careful measurements have been made. The y-emission rate from a caHbration source also depends on the decay rate of the source, and for these nucHdes the uncertainty in the source activity is often the larger uncertainty. [Pg.456]

Industrial sterilization cycles tend to vary considerably, not only from manufacturer to manufacturer, but often from product type to product type, depending on the bioburden present on a given load. Chemical indicators have historically been used only to differentiate between sterilized and nonsterilized packages. More recent developments have resulted in the availability of chemical dosimeters of sufficient accuracy to permit their appHcation either as total monitors or as critical detectors of specific parameters. [Pg.407]

Several types of secondary pyrometer are available. In addition to those that measure by varying lamp current, some pyrometers maintain the lamp at constant current but interpose a wedge of graduated neutral density, whose position is a measure of temperature. Also, automatic pyrometers are available in which the eye is replaced by a detector and the measuring element is operated by a servo. In general, the accuracy of the automatic pyrometer is somewhat less than that achieved manually by a skilled operator. [Pg.404]

Due to the relative uniformity of ion formation by the RF spark (although its timing is erratic), the most widely used method of quantitation in SSMS is to assume equal sensitivity for all elements and to compare the signal for an individual element with that of the total number of ions recorded on the beam monitor. By empirically calibratii the number of ions necessary to produce a certain blackness on the plate detector, one can estimate the concentration. The signal detected must be corrected for isotopic abundance and the known mass response of the ion-sensitive plate. By this procedure to accuracies within a factor of 3 of the true value can be obtained without standards. [Pg.605]

Detectable concentration ranges are mbe-dependent and can be anywhere from one-hundredth to several thousand ppm. The limits of detection depend on the particular detector mbe. Accuracy ranges vary with each detector mbe. The pump may be handheld during operation (weighing from 8 to 11 ounces), or it may be an automatic type (weighing about 4 pounds) which collects a sample using a preset number of pump strokes. [Pg.249]

If the instrument is to be direct reading, the second (or cold ) junction must be kept at a constant reference temperature. If high temperatures are to be measured then the terminals of the detector can be used as the cold junction without an unacceptable loss of accuracy. [Pg.243]

In its simplest form, the Wheatstone bridge is used on D.C. for the measurement of an unknown resistance in terms of three known resistors. Its accuracy depends on that of the known units and the sensitivity of the detector. It is also used for sensing the changes which occur in the output from resistance strain-gauge detectors. The latter instruments can be made portable and can detect variations of less than 0.05 per cent. [Pg.245]

This is a method which is very attractive in principle and which has been applied to yield approximate barriers for a number of molecules. There are, however, difficulties in its use. In the first place, it is not easy to measure the intensities of microwave lines with accuracy. There are unsolved problems of saturation, reflections in the wave guide, and variation of detector efficiency with frequency which are presumably reponsible for the fact that measurements made with ordinary wave guide spectrometers are not very reproducible. In addition, both the spectral lines may be split into components by tunnelling from one potential minimum to another and this splitting, even though it is not resolved, can alter the apparent intensity. Furthermore, it is often difficult to find pairs of lines such that neither is obscured by Stark lobes from the other. [Pg.378]

From a practical point of view it is better to use the maximum volume of sample possible, assuming there is no mass overload. This will allow the detector to be operated at the lowest possible sensitivity and in doing so provide the greatest detector stability and, as a consequence, the highest accuracy. [Pg.97]


See other pages where Detector accuracy is mentioned: [Pg.605]    [Pg.115]    [Pg.687]    [Pg.242]    [Pg.605]    [Pg.115]    [Pg.687]    [Pg.242]    [Pg.81]    [Pg.210]    [Pg.484]    [Pg.585]    [Pg.389]    [Pg.636]    [Pg.651]    [Pg.14]    [Pg.66]    [Pg.465]    [Pg.316]    [Pg.766]    [Pg.1972]    [Pg.393]    [Pg.166]    [Pg.367]    [Pg.367]    [Pg.368]    [Pg.483]    [Pg.689]    [Pg.89]    [Pg.205]    [Pg.33]    [Pg.59]    [Pg.73]    [Pg.48]   
See also in sourсe #XX -- [ Pg.171 ]




SEARCH



Detectors wavelength accuracy

Diode array detector accuracy

© 2024 chempedia.info