Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Depolarized detection

There are different embodiments of SMSLS. They can contain flow cells, batch cells, or both. They can use lasers with beam splitters or individual lasers per cell. Cells can be in series, for applications where there is no tmbidity, in parallel when tmbidity is an issue, or some combination of the two. Other featmes are currently being added to SMSLS. Depolarized detection, for example, can monitor processes in which there are transitions from isotropic to anisotropic morphologies or vice versa, snch as nano- and micro-fibrillar modes of protein aggregation. [Pg.296]

Detection of Bromine Vapor. Bromine vapor in air can be monitored by using an oxidant monitor instmment that sounds an alarm when a certain level is reached. An oxidant monitor operates on an amperometric principle. The bromine oxidizes potassium iodide in solution, producing an electrical output by depolarizing one sensor electrode. Detector tubes, usefiil for determining the level of respiratory protection required, contain (9-toluidine that produces a yellow-orange stain when reacted with bromine. These tubes and sample pumps are available through safety supply companies (54). The usefiil concentration range is 0.2—30 ppm. [Pg.288]

Figure 4.6 shows an apparatus for the fluorescence depolarization measurement. The linearly polarized excitation pulse from a mode-locked Ti-Sapphire laser illuminated a polymer brush sample through a microscope objective. The fluorescence from a specimen was collected by the same objective and input to a polarizing beam splitter to detect 7 and I by photomultipliers (PMTs). The photon signal from the PMT was fed to a time-correlated single photon counting electronics to obtain the time profiles of 7 and I simultaneously. The experimental data of the fluorescence anisotropy was fitted to a double exponential function. [Pg.62]

The ECG has several noteworthy characteristics. First, the firing of the SA node, which initiates the heart beat, precedes atrial depolarization and therefore should be apparent immediately prior to the P wave. However, due to its small size, it does not generate enough electrical activity to spread to the surface of the body and be detected by the electrodes. Therefore, there is no recording of the depolarization of the SA node. [Pg.175]

Cardiac Action Potential In Vitro Purkinje Fibers. Intracellular recording of action potentials from cardiac Purkinje fibers isolated from dog or sheep ventricle. Measurement of maximum rate of depolarization and action potential duration to detect sodium and potassium channel interactions, respectively, according to recommendations in EM A CPMP Points to Consider document, CPMP 986/96 (1998). [Pg.746]

Homo-FRET is a useful tool to study the interactions in living cells that can be detected by the decrease in anisotropy [106, 107]. Since commonly the donor and acceptor dipoles are not perfectly aligned in space, the energy transfer results in depolarization of acceptor emission. Imaging in polarized light can be provided both in confocal and time-resolved microscopies. However, a decrease of steady-state anisotropy can be observed not only due to homo-FRET, but also due to rotation of the fluorescence emitter. The only possibility of discriminating them in an unknown system is to use the variation of excitation wavelength and apply the... [Pg.125]

Presently, the only commercially available dyes that are applied because of then-ability to form fluorescent aggregates are trimethine cyanines JC-1 and JC-9 (Fig. 11) [25], the first one being studied much more extensively than the second one. The dye JC-1 is known to form red-fluorescent (emission maximum at 590 nm) J-aggregates in mitochondria possessing strong intramitochondrial negative potential, while upon depolarization of the mitochondrial membrane, the dye monomer green emission (maximum at 527 nm) is observed [25]. JC-9 demonstrates similar properties [25]. Such properties permit the application of these dyes for, e.g., detection of apoptotic electrical depolarization of mitochondria [25]. [Pg.154]

End-point Detection The end-point of the Karl Fischer titration may be determined quite easily by adopting the electrometric technique employing the dead-stop end-point method. When a small quantum of e.m.f. is applied across two platinum electrodes immersed in the reaction mixture, a current shall tend to flow till free iodine exists, to remove hydrogen and ultimately depolarize the cathode. A situation will soon arise when practically all the traces of iodine have reacted completely thereby setting the current to almost zero or veiy close to zero or attain the end-point. [Pg.224]

Otherwise, depolarization would also be a result of energy transfer between probe molecules. Because the transition moments of two interacting probes are unlikely to be parallel, this effect is indeed formally equivalent to a rotation. Moreover, artefacts may arise from scattering light that is not totally rejected in the detection system. [Pg.245]

Our experiments are typically carried out at DNA concentrations of 20-50 /ig/ml with 1 ethidium per 300 bp, so that depolarization by excitation transfer is negligible.(18) The sample is excited with 575-nm light, and the fluorescence is detected at 630, 640, or 645 nm. Less than one fluorescent photon is detected for every 100 laser shots. The instrument response function e(t) is determined using 575-nm incident light scattered from a suspension of polystyrene latex spheres. [Pg.170]

Similar results were obtained by De Shazer using a different detection technique, where laser oscillations in the sample were forced to develop from the narrow-band radiation, injected from a second small aperture laser into the sample laser cavity. The interionic transfer allowed the feeding of this narrow-band radiation by ions having frequencies outside this interval. The effeciency of energy extraction within the narrow bandwidth and the degree of depolarization of the laser oscillations parametrize the cross relaxation effects. [Pg.77]

C. Atropine will not directly paralyze the respiratory muscles. However, it can prevent the detection of early signs of an overdose of neostigmine, which can quickly progress to a depolarizing block of skeletal muscle and paralysis of the respiratory muscles. Dry mouth, ocular disturbances, and tachycardia are common side effects of atropine given alone, but these effects are less likely to occur with competition between atropine and the increase in the synaptic ACh produced by inhibition of AChE by neostigmine. [Pg.139]


See other pages where Depolarized detection is mentioned: [Pg.246]    [Pg.246]    [Pg.683]    [Pg.80]    [Pg.85]    [Pg.7]    [Pg.115]    [Pg.91]    [Pg.277]    [Pg.136]    [Pg.162]    [Pg.114]    [Pg.20]    [Pg.87]    [Pg.106]    [Pg.560]    [Pg.631]    [Pg.835]    [Pg.863]    [Pg.930]    [Pg.105]    [Pg.198]    [Pg.222]    [Pg.33]    [Pg.271]    [Pg.33]    [Pg.58]    [Pg.536]    [Pg.121]    [Pg.17]    [Pg.192]    [Pg.120]    [Pg.51]    [Pg.404]    [Pg.346]    [Pg.62]    [Pg.264]   
See also in sourсe #XX -- [ Pg.296 ]




SEARCH



Depolarization

Depolarizer (

Depolarizers

© 2024 chempedia.info