Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Dc current decay

For a heterogeneous system, slow polarizations should always occur once an electric field is applied. Those slow polarizations include the Debye polarization, the interfacial polarization, and the electrode polarization if there is electrolyte in the system. Those slow polarizations will result in an unstable dc current dc current decays with time and finally become stable, as shown in Figure 28. Once the applied electric field is turned off, the dc current goes down to zero quickly and continuously drops to a negative current and then gradually decays to zero. The charge and discharge curves... [Pg.413]

The dc current decay function has been determined experimentally. At low temperature the dc current decay curve of a solid material can given as... [Pg.414]

Calculate the space charge amount from dc current decay curve... [Pg.415]

The major advantage of Eq. (208) is that parameters k and n are not involved in the calculation, though n is needed for calculating the dielectric constant. It should be noted that this transient dc current method is only valid for the frequency 10 to 0.1 Hz [I], The reason is obvious The dc current decay function is only valid in a certain time period. [Pg.420]

In such case, the calculation of DC is simple. The criterion for infinite sample dimensions yields that the largest possible D = 10 cm /s. Thus, the electrode of 60 pm thickness behaves as semi-infinite during the 10 s current decay. Therefore, if the experiment s duration is limited to few minutes (10 s) then the allowed sample thickness would be 20 pm. Such consideration is necessary when film or single-crystal electrodes are employed. [Pg.50]

Figure 28. dc current vs. time. After an electric field is applied, dc current immediately jumps lo a high value and then exponentially decay to a steady value once the applied electric field is removed, dc current drops to a negative value, and then decays to zero. [Pg.414]

Figure 7 Typical decay of dc current for insulating polymers. Figure 7 Typical decay of dc current for insulating polymers.
O, a large current is detected, which decays steadily with time. The change in potential from will initiate the very rapid reduction of all the oxidized species at the electrode surface and consequently of all the electroactive species diffrising to the surface. It is effectively an instruction to the electrode to instantaneously change the concentration of O at its surface from the bulk value to zero. The chemical change will lead to concentration gradients, which will decrease with time, ultimately to zero, as the diffrision-layer thickness increases. At time t = 0, on the other hand, dc-Jdx) r. will tend to infinity. The linearity of a plot of i versus r... [Pg.1929]

The important point is that capacitors will, therefore, allow the flow of AC in preference to DC. Because there is less time for current to decay in a high-frequency AC circuit before the polarity reverses, the mean current flow is greater. The acronym CLiFF may help to remind you that capacitors act as low-frequency filters in that they tend to oppose the flow of low frequency or DC. [Pg.43]

Graphs show how capacitors alter current flow within a circuit. The points to demonstrate are that DC decays rapidly to zero and that the mean current flow is less in a low-frequency AC circuit than in a high-frequency one. [Pg.43]

The effective tunneling barrier may be quantified by measuring the distance dependence of the tunnel conductance or tunneling current [Eq. (1) or (2), respectively]. Experimentally, the decay constant, k, may be derived from dc or ac measurements. The more accurate ac modulation... [Pg.224]

It is a particular characteristic in the solution of differential equations involving resistances and inductances that a DC component accompanies the symmetrical AC component. The magnitude of the DC component can equal that of the peak AC component since both are determined by X J. The decay of the DC component can be reasonably slow and is determined by which is a function of X J and the armature winding resistance Ra With machines that have significant values of X J and particularly low values of R, the value of Ta can become relatively high. When Ta is high in relation to Tj and rj it is possible that the initial AC decay is faster than the DC decay. When this happens the AC instantaneous current does not reach zero until several cycles have passed. This puts an extra strain on the circuit breaker and can cause problems at the point when it starts to open to clear the fault current. [Pg.150]

The increase of clock time at a properly chosen pulse amplitude and scan rate does not effect the peak current or the peak potential seriously. The curves are smoothed as T decreases (v and AE fixed) and can consequently be evaluated more precisely. In spite of this a time T = 1.0 is recommended, because the complete decay of the reverse current spike due to the potential step back to the actual dc potential disturbes less the current sample before the application of the next potential pulse, as can be seen in Table 1. The thickness of the spacer does not have any greater effect on the measured dpv current. [Pg.416]

As seen in previous sections, the response to a potential step is a pulse of current, which decreases with time as the electroactive species near the electrode surface is consumed and consists of a faradaic, /f, and a capacitive contribution, Iq. The advantage of most pulse techniques results from the measurement of the current flow near the end of the pulse when the faradaic current has decayed, often to a diffusion-limited value but when the capacitive current is insignificant. Pulse widths, tp, are adjusted to satisfy this condition and the additional condition that time has not been allowed for natural convection effects to influence the response. There is a greatly improved signal-to-noise ratio (sensitivity) compared to steady state techniques and in many cases, greater selectivity. Detection limits are of the order of 10 M. Furthermore, for analytical purposes, most current-voltage profiles from the pulse techniques are faster to interpret than those of dc voltammograms, because they are peak-shaped rather than the typical step curve of conventional voltammet-ric methods. [Pg.111]


See other pages where Dc current decay is mentioned: [Pg.284]    [Pg.415]    [Pg.284]    [Pg.415]    [Pg.172]    [Pg.164]    [Pg.128]    [Pg.129]    [Pg.159]    [Pg.135]    [Pg.315]    [Pg.369]    [Pg.321]    [Pg.328]    [Pg.331]    [Pg.322]    [Pg.278]    [Pg.171]    [Pg.632]    [Pg.344]    [Pg.17]    [Pg.231]    [Pg.66]    [Pg.46]    [Pg.37]    [Pg.363]    [Pg.368]    [Pg.370]    [Pg.173]    [Pg.492]    [Pg.255]    [Pg.379]    [Pg.151]    [Pg.76]    [Pg.76]    [Pg.178]    [Pg.89]   
See also in sourсe #XX -- [ Pg.413 , Pg.414 ]




SEARCH



Current DC

Current decay

© 2024 chempedia.info