Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Concentration determination analysis

Benchtop X-ray energy dispersive analyzer BRA-17-02 based on a gas-filled electroluminescent detector with an x-ray tube excitation and range of the elements to be determined from K (Z=19) to U (Z=92) an electroluminescent detector ensures two times better resolution compared with traditional proportional counters and possesses 20 times greater x-ray efficiency compared with semiconductor detectors. The device is used usually for grits concentration determination when analysing of aviation oils (certified analysis procedures are available) and in mining industry. [Pg.76]

Where a, b, and c = van Deemter coefficients, dp = particle size of column, L = column length, Dm = diffusion coefficients of analytes, t = column dead time (depends on flow rate F), tg= gradient time (determines analysis time via tA = tg + t0), Ac = difference in concentrations of the organic modifier at the end and the beginning of the gradient (a continuous linear gradient is assumed), and B = slope of the linear relationship between the logarithm of the retention factor and the solvent composition. [Pg.97]

Fig. 2 Determination of Bt values (amount of functional immobilized ligand in the column) for the immobilized Erythrina cristagalli agglutinin. / -Nitrophenyl, (pNP)-lactose, diluted to various concentrations (8 to 50 pM), was used for concentration-dependence analysis. (A) The solid and dotted lines demonstrate elution profiles of pNP-lactose and control sugar (pNP-mannose), respectively. (B) Woolf-Hofstee-type plot was made by using V-V0 values. Adapted from 47 with permission. Fig. 2 Determination of Bt values (amount of functional immobilized ligand in the column) for the immobilized Erythrina cristagalli agglutinin. / -Nitrophenyl, (pNP)-lactose, diluted to various concentrations (8 to 50 pM), was used for concentration-dependence analysis. (A) The solid and dotted lines demonstrate elution profiles of pNP-lactose and control sugar (pNP-mannose), respectively. (B) Woolf-Hofstee-type plot was made by using V-V0 values. Adapted from 47 with permission.
The calcium antagonist nicardipine (17) and its pyridine metabolite M-5 were determined in plasma after LLE and concentration. End analysis was by capillary GC-NPD with temperature gradient LOD 0.5 pg/L for both compounds93. See also reaction 27 in Section IV.H for electrochemical processes undergone by similar compounds. [Pg.1062]

Correlation functions are powerful tools in statistical physics, and in the above example they permit one to examine the behavior of a fluctuating system from a reference time back to previous times. Such fluctuations can occur in the concentration of two (or more) interconverting chemical species in dynamic equilibrium, and the technique of concentration correlation analysis permits one to determine the forward and reverse rate constants for their interconversion. See Concentration Correlation Analysis... [Pg.170]

The rate equation to be used for kinetic analysis of enzyme depletion is that for simple noncompetitive inhibition. If the Henderson equation or similar types are not employed, keep in mind that the inhibitor concentration [I] is the free inhibitor concentration. Determination of Ki may not be feasible if the rate assay is insensitive and requires an enzyme concentration much greater than K[. Alternatively, Ki may be obtained by measuring the on-off rate constants of the E l complex, provided the rate constants for any conformation change steps involved are also known. [Pg.242]

Once temperature equilibrium was established, 600 g of alumina seeds were added to the crystallizer to make up a slurry of solids density 200 g/1. Samples for size distribution analysis and solution concentration determinations were t2dcen periodically for some 24 hours. [Pg.331]

Metal Analysis The concentration of Pd in the effluent was determined by ICP or AA and the concentration of Ru was extrapolated from a UV-Vis calibration curve. A Feeman Tabs, Inc. Direct Reading Eschelle ICP-AES was used for Pd concentration determination in the Deloxan MP studies while a Varian Spectra AA 55B flame atomic absorption spectrophotometer was used to determine the Pd concentration in the Deloxan THP II pilot studies. Once the organic solvent was... [Pg.498]

Fig. 5. Colloid concentrations determined under in situ conditions at the Aspo hard rock laboratory, Sweden. Analysis of colloids has been performed on line by (a) using a mobile laser-induced breakdown detection (LIBD) arrangement, and (b) DOC analysis of collected samples in the laboratory (Hauser et al. 2003). Bars in the upper diagram represent colloid concentration ranges detected during the campaign. Colloid concentrations and DOC are plotted against salinity expressed as the groundwater CP concentration (lower x-axis) and ionic strength (upper x-axis). Fig. 5. Colloid concentrations determined under in situ conditions at the Aspo hard rock laboratory, Sweden. Analysis of colloids has been performed on line by (a) using a mobile laser-induced breakdown detection (LIBD) arrangement, and (b) DOC analysis of collected samples in the laboratory (Hauser et al. 2003). Bars in the upper diagram represent colloid concentration ranges detected during the campaign. Colloid concentrations and DOC are plotted against salinity expressed as the groundwater CP concentration (lower x-axis) and ionic strength (upper x-axis).
In air monitoring program accurate concentration determinations require careful attention to the sample collection and subsequent analysis of the collected sample. The analysis is the most critical step, especially if very low levels (ppb) of contaminants are being determined. However, in many cases, sampling may be the least accurate step since this job is performed in the field with portable equipment under conditions far less favorable than those that can be created and controlled in the lab ). ... [Pg.491]

A Linde NaY zeolite without binder was ion exchanged in an ammonia-cal solution of PdCl2 which provides exchangeable (Pd(NH3)4)2+ cations. The solution was stirred at room temperature for 24 hr and then filtered. The zeolite was washed with ammonia solution to eliminate Cl ions. The desired exchange level was readily obtained by allowing the zeolite to equilibrate in a solution where a suitable amount of palladium has been introduced. Chemical analysis for palladium and sodium showed the composition of the calcined sample to be Pdtf.sNaiQ.sHn.sAlfieSiiaeCW (10 wt % of Pd proton concentration determined by difference). [Pg.74]

Conditioning of the manganese oxide suspension with each cation was conducted in a thermostatted cell (25° 0.05°C.) described previously (13). Analyses of residual lithium, potassium, sodium, calcium, and barium were obtained by standard flame photometry techniques on a Beckman DU-2 spectrophotometer with flame attachment. Analyses of copper, nickel, and cobalt were conducted on a Sargent Model XR recording polarograph. Samples for analysis were removed upon equilibration of the system, the solid centrifuged off and analytical concentrations determined from calibration curves. In contrast to Morgan and Stumm (10) who report fairly rapid equilibration, final attainment of equilibrium at constant pH, for example, upon addition of metal ions was often very slow, in some cases of the order of several hours. [Pg.83]

If compounds with very low odor thresholds and very small concentrations contribute to a material s odor their detection can be very challenging, especially when only applying routine emission measurements like GC—MS. Such compounds will easily be overlooked, for their detection GC—O can often be the only choice, but so far this method is seldom used in material analysis. Instead concentrations determined by emission measurements are compared with published odor thresholds to decide whether a compound might contribute to the odor or not. One problem is that published odor thresholds can differ quite a lot, even by several orders of magnitude (van Gemert, 2003). The value depends on the method and the panel but also on the purity of the compound used for threshold determination (if small impurities of a substance with a low odor threshold were present in a sample the odor threshold determined would have been too low ). Many factors influence odor threshold determination, therefore many published values are questionable and they are hard to rely on. Some authors (Knudsen et al., 1999 Wolkoff, 1999 Wolkoff et al., 2006) assume that many of the odor thresholds reported in the literature are actually much lower, because if they compare concentrations of compounds emitted and measured with odor thresholds published,... [Pg.181]

The method of analysis must be very much faster than the reaction itself, so that virtually no reaction will occur during the period of concentration determination. [Pg.19]

Correlation approach Objective is to determine if there is a consistent relationship between the concentration of the suspected toxicant(s) and effluent toxicity. A wide range of toxicity responses with several samples must be obtained in order to provide an adequate range of effect concentrations for analysis. Two key problems associated with the correlation approach are 1) lack of additivity requires careful data analysis, and 2) analysis is difficult when matrix effects are present. [Pg.194]

Surfactin in the supernatant was confirmed and the concentration determined by high-performance liquid chromatography analysis as described by Noah et al. (4). Surface tension was measured using video image analysis of inverted pendant drops as previously described (5). All data points are an average of five measurements taken of the cell-free supernatant. [Pg.829]

Determination of the single operator figures of merit. Accuracy, precision, detection limits, linear dynamic range, and sensitivity are determined. Analysis is performed at different concentrations using standards. [Pg.16]

Following procedure A the concentration of total Hyo was determined and following procedure B only the remaining i -Hyo portion was measured. 5-Hyo was calculated by the difference of both concentrations. Enantioselective analysis was performed according to John et al. [49]... [Pg.323]

True Concentration Concentration Determined by Analysis Simultaneous Concentration Determined by Analysis of Neutral Solution Determined by Analysis of Acid Solution ... [Pg.111]

Waters are susceptible to change by differing extents as a result of physical, chemical or biological reactions that take place between the time of sampling and analysis. If suitable precautions are not taken before and during transport as well as the time spent in the laboratory, then the nature and rate of these reactions are often such that concentrations determined will be different from those existing at the time of sampling. The causes of variation may include... [Pg.45]

NMR-characterization. 27 Corbin et al. (35) were able to show by a systematic study that Z/A1 MAS NMR gives the true Si/Al ratio with a mean error of 10 %, if two conditions are met a) The amount of paramagnetic species is less than 0.05 % and b) the sample does not contain "NMR-invisible" aluminium. Chemical analyses of the samples under study showed that condition one is fulfilled. If samples contain "NMR-invisible" aluminium a difference between the concentration determined by chemical analysis and the framework aluminium concentration determined by NMR should be observed. From the absence of such a difference we conclude that "NMR-invisible" aluminium species do not exist in our samples. Also a line at the position of about 0 ppm due to octahedrally coordinated non-framework aluminium and a broad line at about 30 ppm due to tetrahedrally coordinated nonframework aluminium (36) could not be observed. The values for the concentration of framework aluminium atoms derived from the intensities of the line at about 60 ppm (see below) are in good agreement with those corresponding to the amount of alumina used in the synthesis mixtures. In conclusion, through the Al MAS NMR measurements it was possible to show that all aluminium atoms are incorporated in tetrahedrally oxygen coordinated framework positions. [Pg.281]

The term toxic unit (TU) plays an important role in mixture concentration-response analysis. It is defined as the actual concentration of a chemical in the mixture divided by its effect concentration (e.g., c/EC50 Sprague 1970). The toxic unit is equivalent to the hazard quotient (HQ), which is used for calculating the hazard index (HI Hertzberg and Teuschler 2002). The term hazard quotient is generally used more in the context of risk assessment (see Chapter 5 on risk assessment), and the term toxic unit is used more in the context of concentration-response analysis, and therefore the latter term is used here. Toxic units are important for 2 reasons. First, toxic units are the core of the concept of concentration addition concentration addition occurs if the toxic units of the chemicals in a mixture that causes 50% effect sum up to 1. Second, toxic units can help to determine which concentrations of the chemicals to test when a mixture experiment needs to be designed. [Pg.122]


See other pages where Concentration determination analysis is mentioned: [Pg.136]    [Pg.373]    [Pg.605]    [Pg.621]    [Pg.99]    [Pg.171]    [Pg.146]    [Pg.506]    [Pg.435]    [Pg.729]    [Pg.158]    [Pg.160]    [Pg.219]    [Pg.197]    [Pg.139]    [Pg.200]    [Pg.192]    [Pg.335]    [Pg.108]    [Pg.376]    [Pg.13]    [Pg.514]    [Pg.88]    [Pg.277]    [Pg.318]    [Pg.276]    [Pg.300]    [Pg.353]   
See also in sourсe #XX -- [ Pg.44 ]




SEARCH



Concentration determine

Concentration, determination

Determining concentration

© 2024 chempedia.info