Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Collimation of the beam

The property characterizing the collimation of the beam formed is the spatial distribution 1(9, emitted particles. Due to the axial symmetry of the tube, the angle dependence is only on which is the angle of the emitted particles against the tube axis. The connection between 1(9) and the total rate of flow N is given by... [Pg.411]

The x-ray tube assembly is a simple and maintenance-free device. However, the overall efficiency of an x-ray tube is very low - approximately 1% or less. Most of the energy supplied to the tube is converted into heat, and therefore, the anode must be continuously cooled with chilled water to avoid target meltdown. The input power to the sealed x-ray tube ( 0.5 to 3 kW) is therefore, limited by the tube s ability to dissipate heat, but the resultant energy of the usable x-ray beam is much lower than 1% of the input power because only a small fraction of the generated photons exits through each window. Additional losses occur during the monochromatization and collimation of the beam (see section 2.3). [Pg.105]

Collimation of the beam hardly works beyond a reduction of 50% in illuminated area, The collimator must be placed some distance in front of the sample to avoid that forward or backward detectors pick up positrons stemming from muons stopped in the collimator, The divergence of the muon beam and especially scattering of muons by the collimator and also in the start counter, the cryostat windows etc. break up the collimated beam again in the case of surface muons. [Pg.85]

When this requirement is fulfilled, and when the resolution related to collimation of the beam is improved, the method is indeed powerful eventually the singularities associated with the existence of a slowly decreasing perturbation potentials becomes observable. [Pg.266]

X-ray mirrors are used for rejecting harmonics, focusing, power filtering, displacing the beam, and improving collimation of the beam incident on the first crystal in order to improve energy resolution. [Pg.1279]

The Doppler width at optical frequencies can be considerably reduced by using a well-collimated beam of atoms, as described by Odintsov (1961) and Stanley (1966). The radiation emitted or absorbed is observed in a direction at right angles to the beam and the effective velocity in the line of sight can be reduced indefinitely by increasing the collimation of the beam (Problem 8.8). These sources are, however, considerably more difficult to construct than conventional discharge tubes or absorption cells and have very low intensities and absorption coefficients. In addition... [Pg.249]

The experiment was carried out by a continuously working Nd YAG-laser fabricated by NEC. The laser has a maximum output of 1200 W and is controlled by handling facility with a linear axle. A stage index fiber optical waveguide with a diameter of d=1000 pm was used for the control of the beam. The focusing optics consist of a focusing lens (f=l 16 mm) and a collimation lens (f=70 mm). [Pg.543]

A stream of a liquid solution can be broken up into a spray of fine drops from which, under the action of aligned nozzles (skimmers) and vacuum regions, the solvent is removed to leave a beam of solute molecules, ready for ionization. The collimation of the initial spray into a linearly directed assembly of droplets, which become clusters and then single molecules, gives rise to the term particle beam interface. [Pg.393]

The value of the beam divergence angle given by equation 3 corresponds to that emerging from the laser. The beam may be collimated further by a telescope. The improvement of the coUimation is the inverse of the magnification of the telescope ... [Pg.2]

A beam from an actual sample will require a more elaborate slit S3rstem for collimation if the sample is broad. The Soller slit (Figure 4-7), a stack of thin parallel plates, is such a system. The reasoning that supports this construction is as follows. Were the sample a point or a line source, a slit between sample and crystal or a slit between crystal and detector would be enough for satisfactory collimation. With a two-dimensional sample, both slits would be needed to get this done. But this arrangement is wasteful of emitted intensity because the detector sees the sample as a line source. To use all the sample area effectively, a system of parallel slits is needed. To eliminate the divergent rays in such a system, the slits must be extended in the direction of the beam, and this leads to the parallel-plate construction in the Seller slit system. [Pg.111]

Fig. 4-7. Diffraction of a divergent beam from a broad sample by a large crystal. Collimation of this beam requires the Soller slit system shown. This system is equivalent to simple slits at A and B with separators provided to make certain that only parallel rays leave the exit slit. Fig. 4-7. Diffraction of a divergent beam from a broad sample by a large crystal. Collimation of this beam requires the Soller slit system shown. This system is equivalent to simple slits at A and B with separators provided to make certain that only parallel rays leave the exit slit.
The cross-section of the primary X-ray beam is extended and not an ideal point. This fact results in a blurring of the recorded scattering pattern. By keeping the cross-section tiny, modern equipment is close to the point-focus collimation approximation - because, in general, the features of the scattering patterns are relatively broad. Care must be taken, if narrow peaks like equatorial streaks (cf. p. 166) are observed and discussed. The solution is either to desmear the scattering pattern or to correct the determined structure parameters for the integral breadth of the beam profile (Sect. 9.7). [Pg.56]

Information on particle size may be obtained from the sedimentation of particles in dilute suspensions. The use of pipette techniques can be rather tedious and care is required to ensure that measurements are sufficiently precise. Instruments such as X-ray or photo-sedimentometers serve to automate this method in a non-intrusive manner. The attenuation of a narrow collimated beam of radiation passing horizontally through a sample of suspension is related to the mass of solid material in the path of the beam. This attenuation can be monitored at a fixed height in the suspension, or can be monitored as the beam is raised at a known rate. This latter procedure serves to reduce the time required to obtain sufficient data from which the particle size distribution may be calculated. This technique is limited to the analysis of particles whose settling behaviour follows Stokes law, as discussed in Section 3.3.4, and to conditions where any diffusive motion of particles is negligible. [Pg.9]

Better conventional collimation will not do, except for the largest synchrotron radiation installations to obtain snb-arc-second collimation in the laboratory would require a collimator some 100m long with a sealed-tube source, and at this distance the intensity would be impracticably low. The problem is solved by the use of a beam conditioner, which is a further diffracting system before the specimen The measnred rocking cnrve is then the correlation of the plane wave rocking cnrves of the beam conditioner and the specimen crystals, from which most of the diffracting characteristics of the specimen crystal may be deduced. [Pg.9]

We therefore need to limit the divergence and wavelength spread of the beam incident upon the specimen. Beam conditioners are used to collimate and to... [Pg.15]

X-rays are removed from a collimated beam either by scattering out of the beam or by being absorbed. The process is quantized and probabilistic in nature. An X-ray beam can be considered as a flow of discrete photon particles each having a specific... [Pg.92]

All the above techniques use incident monochromatic radiation, usually focus in one or two dimensions. However for cases a) and d) the reduction of radiation damage and more particularly in kinetic crystallography the use of polychromatic data collection is yielding promising results. This technique makes combined use of the intensity and collimation of the SR beam with a large wavelength spread for Laue data colla tion from protein single crystals. [Pg.35]

Otherwise commercially interesting proteins, which are not readily crystallizable. The high intensity and collimation of the SR is currently allowing data to be collected and several structures to be studied where samples as small as 100 x 100 x 15 pm and 200 x 50 x 50 pm (Ealick, Helliwell and Cook, unpublished) with unit cell volumes in the range of 10 A or so.. Sample volumes as small as (20 pm) have yielded very strong diffraction patterns with the white SR beam, (Fig. 6) (Hedman et al. 1985). In this application area efforts to reduce the X-ray background at the detector include replacement of traditional glass capillaries in the sample with thin film capillaries. [Pg.45]

The energy of photons with their optimal beam penetration, the reliability of the beam delivery and collimation systems, and the mechanical stability of the new generation of accelerators have achieved a nearly optimum level of performance. Actually, with the modern linear electron accelerators, it is now possible to irradiate at the prescribed dose (nearly) any target volume of any shape with reduced irradiation of the surrounding organs at risk (OAR). [Pg.747]

Figure B3.6.5 The inner filter effect. A cuvette (10 x 10-mm) is represented in plan view, with the collimated incident beam from the monochromator having intensity /0. As a result of absorption by the protein solution, the intensity of the beam through the cuvette will decrease steadily, emerging with intensity /. The values are illustrated for a solution having an absorbance at the excitation wavelength of 0.1. The optics of the fluorescence detector are focused so that only fluorescence originating from the volume depicted by the heavily shaded square is seen by the photomultiplier. Thus the observed normalized fluorescence intensity will be less than that expected from the protein at infinite dilution. The fluorescence passes through the protein solution on its way to the detector and will be further decreased in intensity if the solution absorbs at the wavelengths of the emitted radiation. Figure B3.6.5 The inner filter effect. A cuvette (10 x 10-mm) is represented in plan view, with the collimated incident beam from the monochromator having intensity /0. As a result of absorption by the protein solution, the intensity of the beam through the cuvette will decrease steadily, emerging with intensity /. The values are illustrated for a solution having an absorbance at the excitation wavelength of 0.1. The optics of the fluorescence detector are focused so that only fluorescence originating from the volume depicted by the heavily shaded square is seen by the photomultiplier. Thus the observed normalized fluorescence intensity will be less than that expected from the protein at infinite dilution. The fluorescence passes through the protein solution on its way to the detector and will be further decreased in intensity if the solution absorbs at the wavelengths of the emitted radiation.
A stretched fiber containing many poly-L-alanine molecules is suspended vertically and exposed to a collimated monochromatic beam of CuKa x-rays, as shown in figure 1(a). Only a small percentage of the x-ray beam is diffracted most of the beam travels through the specimen with no change in direction. A photographic film is held in back of the specimen. A hole in the center of the film allows the incident undiffracted beam to pass through. [Pg.96]


See other pages where Collimation of the beam is mentioned: [Pg.127]    [Pg.125]    [Pg.433]    [Pg.123]    [Pg.195]    [Pg.231]    [Pg.269]    [Pg.104]    [Pg.426]    [Pg.282]    [Pg.81]    [Pg.350]    [Pg.673]    [Pg.127]    [Pg.125]    [Pg.433]    [Pg.123]    [Pg.195]    [Pg.231]    [Pg.269]    [Pg.104]    [Pg.426]    [Pg.282]    [Pg.81]    [Pg.350]    [Pg.673]    [Pg.593]    [Pg.2]    [Pg.13]    [Pg.427]    [Pg.210]    [Pg.185]    [Pg.100]    [Pg.331]    [Pg.263]    [Pg.207]    [Pg.245]    [Pg.118]    [Pg.43]    [Pg.649]    [Pg.108]   
See also in sourсe #XX -- [ Pg.266 ]




SEARCH



Beam collimator

Collimated

Collimated beam

© 2024 chempedia.info