Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Capacitance measurements, signal

Position sensors convert the position into an electrically measurable signal such as resistance, voltage, current, inductance, pulses, or capacitance. The simplest and most widely used position sensor is the potentiometer. The potentiometer has three terminals, one for each end of the resistive element and one for the brush. As the brush moves, the resistance between the center tap and end terminal changes (Figure 3.130). [Pg.468]

Dielectric spectroscopy or culture capacitance measurement is used as an on-line, non-invasive method for biomass estimation (see the chapter by Sonnleitner in this issue - the section on electrical properties) and responds mainly to living cells [43,44]. Observed difficulties in using the signal as a pure biomass concentration sensor, i.e. deviations from the simple correlation with cell density, were attributed to dependencies on the physiological state [43], and could be used to discriminate different populations in yeast cultures [45]. Connections with morphological features could be found for budding yeast... [Pg.193]

Film thickness could be varied from 500 nm to 800 nm by changing the polymer concentration in the solution and the spinning speed. The film thickness was measured either by means of the X-ray grazing angle reflectivity method, or by an Alpha-step profilometer, in the case of thicker films. Independent low-signal capacitance measurements yield consistent results for film thickness. [Pg.189]

The use of enzyme labels in ELIS A-type immunosensors and simple amperometric detection schemes resulted in simple and cost-effective alternatives to fluorescence immunosensors. In particular, the use of alkaline phosphatase as enzyme label allowed for the fabrication of advanced immunosensors with signal amphfi-cation by means of redox cycling, which has been a success story of its own. This detection scheme has been used in immunosensors and other biosensors and has stimulated significant developments in electrode fabrication. Instrumental electroanalysis, namely capacitance measurements and EIS allow for label-free detection of immunoreactions. [Pg.55]

To excite the polymer film, an input voltage waveform causes current to flow through the cells in a cartridge. An AC signal was chosen to (i) facilitate the capacitance measurement (ii) avoid electrode polarization and (iii) lower susceptibility to noise. [Pg.473]

Two-Position Control. The simplest case is two-position (on-off) control. Here, any deviation of the measured value from a set point drives the final control-operator to either a full-on or full-off position. This forces the measured value back and forth across the set point, and the measurement signal cycles about this point. The amplitude and frequency of this cycle depend on the response characteristics of the process. As the process dead-time becomes small, the frequency of the cycle becomes high likewise, as the process capacitance becomes high, the amplitude of the cycle becomes small. This mode of control is used only for processes in which this cycling effect can be tolerated it is most successful with those having large capacitance. [Pg.775]

For monitoring electrophysiological signals the measuring electrodes can be either placed directly onto the human skin or based on capacitive measurements, in which direct contact between the human skin and the electrodes is not needed. [Pg.211]

The measurement object, the working electrode, and the reference electrode of the Kelvin probe form, due to the small gap between them, a capacitor. Between them a potential is developed, the amplitude of which gives a measure of corrosion activity. A periodic variation in separation by means of an actuator built into the sensor changes the capacitance of the setup. The resulting signal is converted to a measurement signal by means of a lock-in amplifier [88]. The Volta-potential difference is directly determined by the corrosion potential [89]. [Pg.1081]

In Figure 2.6, the movement of solvent is detected by the movement of a bubble in a very narrow tube attached to the solvent compartment of the osmometer. When the photocell detects movement of the bubble, an external head of solvent is established by the raising of the solvent reservoir. In Figure 2.7, a capacitance device is used to detect solvent movement across the membrane. The measured signal is used to generate an appropriate external solvent head equivalent to the osmotic pressure. Modern instrumentation, once set up properly, can reduce the measurement time to a matter of minutes thus reducing the errors which are found if the membranes are anything less than truly semi-permeable. [Pg.61]

The nanostructured electrodes can store a large number of electrons, which impUes that the photocnrrent is driven into a capacitive element. This introduces an additional time constant, the RC time, in the photocurrent response. To deal with it, Eqs (51) and (52) mnst be multiplied by the transfer function of the measuring system, presented in Figure 4.3.31. The transfer function of the measuring system and the measured signals in the frequency domain are given by Eqs (32) and (33), respectively. [Pg.340]

Photocurrent-voltage data showed that all the supra-band-edge redox species listed above are reduced upon illumination of the p-Si electrode no appreciable reduction was achieved in the dark. Typical results for p-Si with methanol and acetonitrile are shown in Figure 11. Capacitance measurements as a function of electrode potential, ac signal frequency, and light intensity are shown in Figures 12 and 13. [Pg.285]

Fig. 3 Equivalent circuit for the coupling of a ferroelectret sensor to a measuring device. The device is represented by its input capacitance Ca and by its input resistance Ra. The measurement signal is the voltage across the capacitors and the resistance... Fig. 3 Equivalent circuit for the coupling of a ferroelectret sensor to a measuring device. The device is represented by its input capacitance Ca and by its input resistance Ra. The measurement signal is the voltage across the capacitors and the resistance...

See other pages where Capacitance measurements, signal is mentioned: [Pg.508]    [Pg.291]    [Pg.216]    [Pg.146]    [Pg.257]    [Pg.330]    [Pg.32]    [Pg.181]    [Pg.469]    [Pg.375]    [Pg.1255]    [Pg.46]    [Pg.280]    [Pg.443]    [Pg.186]    [Pg.193]    [Pg.205]    [Pg.193]    [Pg.121]    [Pg.636]    [Pg.149]    [Pg.5348]    [Pg.643]    [Pg.356]    [Pg.790]    [Pg.106]    [Pg.864]    [Pg.665]    [Pg.126]    [Pg.862]    [Pg.33]    [Pg.159]    [Pg.92]    [Pg.46]    [Pg.84]    [Pg.407]    [Pg.661]   


SEARCH



Capacitance measurements

Capacitance measurements, signal transduction

Signal measurements

Signal, measuring

© 2024 chempedia.info