Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Calibration of the method

A series of standards is prepared by adding decreasing volumes of working solution to 100 ml calibrated flasks by means of a pipette and dilution to volume with oxygen-free water via a siphon (see Table S-1). Calculate the exact sulphide concentration using the formulae [Pg.96]


The determination of the relationship between detector response and the sample concentration is termed the calibration of the method. There are two types of methods in use for the quantitative analysis of a sample, i.e., the external standard and the internal standard method. An external standard method is a direct comparison of the detector response of a pure compound (standard) to a sample.2 The calibration of the method is performed by preparing standards of varying concentration and analyzing them by a developed method. Method 1 (below) was developed for toluene, and standards of varying concentration were prepared and analyzed. The results obtained are summarized in Table 2 see Figure 3. [Pg.156]

For a qualitative analysis it is sufficient to be able to apply a test which has a known sensitivity limit so that negative and positive results may be seen in the right perspective. Where a quantitative analysis is made, however, the relation between measurement and analyte must obey a strict and measurable proportionality only then can the amount of analyte in the sample be derived from the measurement. To maintain this proportionality it is generally essential that all reactions used in the preparation of a sample for measurement are controlled and reproducible and that the conditions of measurement remain constant for all similar measurements. A premium is also placed upon careful calibration of the methods used in a quantitative analysis. These aspects of chemical analysis are a major pre-occupation of the analyst. [Pg.612]

Where the calibration of the method is very stable a one-point calibration to verify a previous multipoint calibration maybe sufficient... [Pg.192]

The target valne of a Difference Chart shonld be aronnd zero. Otherwise there is a drift in the analyses dnring the series and again the laboratory has to decide if this is acceptable. If not, the calibration of the method has to be made more freqnentiy. [Pg.280]

These methods allow a superior spectroscopic experiment to be performed because they allow the calibration of the method to depend in an optimal way on the experimentation done. These calibration methods ensure that the best measurements are done. They are especially useful for noisy data with substantial interferences, such as those found in many types of ocean science measurements. [Pg.81]

A solution viscosity measurement is a hydrodynamic-thermodynamic technique, and the extent to which a polymer molecule increases the viscosity of a solvent depends on the nature of its interactions with that solvent (as well as on its own molecular weight). These interactions are characterized by the quantity a that appears in equation (3). The calibration of the method, using samples, of the same polymer having known molecular weights, in essence determines its value. The disadvantage of this calibration requirement is offset by the simplicity of the experimental measurements. [Pg.10]

Limitations. The K/K method of calculating CRE ages works poorly for ages less than 100 Myr because of the need for large corrections for native potassium. The absolute calibration of the method is indirect and rests on educated guesses and assumptions. Nickel content may influence the results but is not normally considered explicitly. [Pg.356]

The case presented here is simple because all six samples can be analysed as one set of samples in one analytical run. For a single run the estimation of uncertainty can be reduced to the calibration of the method with an in-house reference material and the uncertainty of the analysis itself. In this example inhomogeneity of the sample, the evaluation uncertainty and the uncertainty of the reference material (i.e. inhomogeneity) as given in Table 1 can be neglected because the results of only one analysis are compared. The weighing uncertainties can be neglected because they are small compared to the sampler uncertainty. [Pg.77]

In the exercises, you re asked to do similar calibrations of the method using quasi-periodic data as well as time series f rom the Lorenz and Rossler attractors. [Pg.441]

Additional factors to consider in the calibration of the method are the curve fit of choice and possible weighting factors associated with the chosen model. The statistical... [Pg.170]

The determination of chemical species in an environmental or biological sample can be difficult and time consuming. The accuracy and precision generally achieved for this type of work is not as good as that for total elemental analysis, even for laboratories with a proven track record in speciation analysis. Only a limited number of high purity standards are available, and so calibration of the method is an important area for improvement in methodology. Standard additions and external calibration strategies have been applied and recently, the use of IDMS has been described. [Pg.622]

In the present time, many issues in regard to the appropriateness of PT methods are obviated by the use of density functional methods, although this in no way reduces the need for calibration of the methods being used. Indeed, wave-function-based methods such as HF and MP2 are excellent choices along with DFT for conducting sensitivity analyses of calculated properties, as they are typically quick calculations and reasonably different in approach from density functional theory. [Pg.649]

In this method the voltage developed between the measuring electrode and the reference electrode is measured. Under certain conditions this measured value is proportional to the logarithm of the fluoride concentration. Attention must be paid to the water temperature, the ionic strength of the dissolved substances and the pH value. These parameters should be standardized in the calibration of the method in such a way that interference in the course of the actual measurement can be largely ruled out. [Pg.215]

Evaluate the performance of the method to ensure that it is performing according to the standard expected based on method validation data and historical data. This could include a check on the overall calibration of the method by comparison of the within-run calibrators with freshly prepared calibrators, and similarly for the QC samples. [Pg.581]

The greatest advantage is in the simple and very stable calibration of the method. [Pg.163]


See other pages where Calibration of the method is mentioned: [Pg.77]    [Pg.56]    [Pg.340]    [Pg.289]    [Pg.296]    [Pg.651]    [Pg.42]    [Pg.77]    [Pg.378]    [Pg.48]    [Pg.292]    [Pg.175]    [Pg.183]    [Pg.39]    [Pg.342]    [Pg.129]    [Pg.660]    [Pg.184]    [Pg.96]    [Pg.98]    [Pg.274]    [Pg.140]   


SEARCH



Calibration of methods

Method calibration

© 2024 chempedia.info