Magnetic Sector Field Analyzer

Apart from the quadrupole and TOP analyzers described in Sect. 3.2.2, the most important types of mass analyzer used in common dynamic SIMS instruments employ a magnetic-sector field. [Pg.109]

Fig. 3.19. Basic set-up of a direct imaging magnetic sector instrument. The stigmatic secondary ion optics consists of an electrostatic analyzer (ESA) and a magnet sector field. Fig. 3.19. Basic set-up of a direct imaging magnetic sector instrument. The stigmatic secondary ion optics consists of an electrostatic analyzer (ESA) and a magnet sector field.
The direct imaging magnetic sector mass analyzer (Fig. 3.19) has the unique property that all parts (lenses, electrostatic analyzer and magnetic sector field) of the secondary ion optics are stigmatic (comparable with light microscopes). This means that all points of the surface are simultaneously projected into the analyzer. [Pg.111]

In both electron post-ionization techniques mass analysis is performed by means of a quadrupole mass analyzer (Sect. 3.1.2.2), and pulse counting by means of a dynode multiplier. In contrast with a magnetic sector field, a quadrupole enables swift switching between mass settings, thus enabling continuous data acquisition for many elements even at high sputter rates within thin layers. [Pg.126]

The most commonly used mass separators are quadrupoles, ions traps and time-of-flight analyzers, for which the principle of mass separation is discussed below. Additionally, other types such as magnetic sector field or Fourier-trans-form cyclotron-resonance instruments are available. [Pg.54]

Magnetic sector field (B) Combination of magnetic (B) and electric sector fields (E) Quadrupole mass analyzer (Q) Time-of-flight mass analyzer (ToF)... [Pg.4]

The ions formed are extracted by the extraction lens, EL, perpendicular to the electron beam. Due to the low energy spread of ions formed in an electron ionization source (< 1 eV) one single-focusing magnetic sector field mass spectrometer or quadrupole analyzer is sufficient for... [Pg.66]

Dynamic mass separation systems use the fact that ions with different masses (accelerated with the same voltage) possess several velocities and consequently their flight times are different. There are about 50 dynamic separation systems known2 using several types of ion movements (linear straight ahead, linear periodic or circular periodic as a function of the electric or magnetic sector field applied). The simplest dynamic mass separation system is a linear time-of-flight (ToF) mass analyzer, and a widely applied mass separation system is the quadrupole analyzer. [Pg.87]

ToF mass spectrometers as dynamic instruments gained popularity with the introduction of matrix assisted laser desorption/ionization (MALDI) and electrospray ionization (ESI) as effective pulsed ion sources for the soft ionization of large biomolecules (up to 10s dalton) due to their high ion transmission.38 ToF mass spectrometers, quadrupole analyzers and/or magnetic sector fields can be combined in tandem mass spectrometers (MS/MS) for the analysis of organic compounds. [Pg.133]

Fig. 120. Glow discharge mass spectrometry. A Sector field based instrument (a) glow discharge source (b) magnetic sector field (c) electrostatic energy analyzer (d) Daly detector (e) ... Fig. 120. Glow discharge mass spectrometry. A Sector field based instrument (a) glow discharge source (b) magnetic sector field (c) electrostatic energy analyzer (d) Daly detector (e) ...
The ICP/MS is an elemental and isotopic analysis method that was first developed in the early 1980s. The ICP had been used only as a source for emission spectroscopy until it was adapted for producing ions for a mass analyzer (Douglas and French, 1981 Houk et al., 1980 Houk et al., 1981 Houk and Thompson, 1982). Since 1983, several manufacturers have sold ICP/MS instruments that incorporate various mass analyzer systems, such as quadrupole mass filter, magnetic sector field, time-of-flight, Paul ion trap, and ion detection systems such as the electron... [Pg.386]

In contrast, the methylated higher ohgosaccharides were overestimated by the other instruments. In the study already mentioned, Naven and Harvey found that there was no discrimination of lower molecular weight analytes when a MALDI ion source and a magnetic-sector field mass analyzer were combined. ... [Pg.166]


See other pages where Magnetic Sector Field Analyzer is mentioned: [Pg.156]    [Pg.156]    [Pg.156]    [Pg.156]    [Pg.52]    [Pg.53]    [Pg.378]    [Pg.5]    [Pg.7]    [Pg.11]    [Pg.28]    [Pg.78]    [Pg.83]    [Pg.135]    [Pg.138]    [Pg.160]    [Pg.168]    [Pg.53]    [Pg.476]    [Pg.7]    [Pg.11]    [Pg.78]    [Pg.83]    [Pg.135]    [Pg.138]    [Pg.160]    [Pg.168]    [Pg.295]    [Pg.1545]    [Pg.1692]    [Pg.6]    [Pg.384]    [Pg.387]    [Pg.264]   
See also in sourсe #XX -- [ Pg.113 ]




SEARCH



Analyzer magnet

Analyzer magnetic field

Magnetic analyzer

Magnetic sector

Magnetic sector analyzer

Magnetic sector field

Sector

Sector analyzers

Sector field

Sector field analyzers

Sectorization

© 2024 chempedia.info