Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Amplifiers reference wire

Figure 8.14 Differential preamplifier well adapted for serving as biopotential preamplifier. Input equipped with three-wire connections for two PU electrodes and one Ind electrode. The indifferent electrode is coupled to the amplifier reference wire. The purpose is to equipotentialize... Figure 8.14 Differential preamplifier well adapted for serving as biopotential preamplifier. Input equipped with three-wire connections for two PU electrodes and one Ind electrode. The indifferent electrode is coupled to the amplifier reference wire. The purpose is to equipotentialize...
Are there lock-in amplifier models that can measure nonlinearity on a continuous basis What is an Amplifier Reference Wire ... [Pg.328]

The amplifier s reference wire is the electronic zero of the electronic circuitry. The reference wire of the amplifier may be floating or referred to ground. The wire is connected to the indifferent electrode wire of the person to be measured it is the system for potential equalization between the measured person and the instrumentation. [Pg.232]

These amplifiers are active devices in which the inputs need a certain bias current and must operate within the limits of the power supply voltage. The power supply wires and the reference wire (0 V), as illustrated in Figure 8.12(a), are frequently omitted on most circuit diagrams. Notice that the inputs are galvanically separated neither from the output, nor from the power supply. Transcouplers or optocouplers are needed if galvanic separation is necessary (Min et ah, 2006). [Pg.285]

The CMV must usually be within hmits set by the power supply of the amplifier. If the supply is 12 V, the CMV input range is perhaps 9 V. For this reason, the patient/test person usually must have a third electrode connected to the reference wire of the input amplifier. Without this third wire, the input amphfier s CMV range may easily be exceeded. Both DC and AC must be considered in this respect. In BF and CF equipment (c.f. Section 10.16.6), the third electrode is a floating reference electrode, in B equipment the amplifier common reference wire is grounded and therefore grounds the patient. [Pg.288]

In more demanding applications, the electrode lead wires should be shielded, preferably by purchasing the type that has a fine coaxial braided copper wire around the central signal-carrying lead. The shield should be connected to the amplifier reference but not connected to the electrode signal wire itself since this would abrogate its shielding effect. [Pg.424]

FIG. 3 (a) Block schematic of the typical instrumentation for SECM with an amperometric UME tip. The tip position may be controlled with various micropositioners, as outlined in the text. The tip potential is applied, with respect to a reference electrode, using a potential programmer, and the current is measured with a simple amplifier device. The tip position may be viewed using a video microscope, (b) Schematic of the submarine UME configuration, which facilitates interfacial electrochemical measurements when the phase containing the UME is more dense than the second phase. In this case, the glass capillary is attached to suitable micropositioners and electrical contact is made via the insulated copper wire shown. [Pg.294]

Obtain the sensor amplifier module board from your instructor, and examine it closely as follows. Notice the outlined horizontal box in the top center of the board—it is labeled instrumentation amplifier. Notice the symbol of an amplifier in this box (refer to Figure 6.10). Also, notice the inputs to the amplifier (labeled P14 and P15). These are sockets for the wires that will bring the input signals to the amplifier. It is the difference between these two signals that is amplified. The output of the amplifier (the amplified signal) is connected to sockets P19 and P20. Wires inserted into these sockets allow us to observe the amplified signal with a voltmeter. [Pg.171]

The connection with the potentiostat should shielded shielded reference electrodes are commercially available. However, in the case of excessive capacitance to ground ii tability of the control potential may occur, in such cases the shielding may be maintained at the reference potential via an operational amplifier [5]. Another possibility is to provide a platinum wire to be used as the reference in the control loop, next to the reference electrode proper. [Pg.21]

The result is that small line-frequency currents flow in loops between different instruments interconnected by multiple signal and ground references. Line-frequency voltage drops in the interconnecting reference path wiring can appear as a signal across the amplifier inputs. [Pg.426]

Figure 2 The principle of measurement by means of an ion-selective double-barreled microelectrode inside a cell. The cell is in a bath the solution of which is grounded via a reference electrode. Each barrel is connected via a chlorided silver wire (shown coiled) to amplifiers (triangles). The reference barrel of the double-barreled electrode directly records the intracellular electrical potential, the membrane potential (Em). The ion-selective barrel, indicated by the plug of ion exchanger in the tip, records the sum of the membrane potential and a potential , related to the chemical potential of the ion in question (of activity a) (see eqn [1]). / is obtained by electronic subtraction. The influence on from other ions (indicated by index j and the valencies Zy) can be obtained from calibration. Figure 2 The principle of measurement by means of an ion-selective double-barreled microelectrode inside a cell. The cell is in a bath the solution of which is grounded via a reference electrode. Each barrel is connected via a chlorided silver wire (shown coiled) to amplifiers (triangles). The reference barrel of the double-barreled electrode directly records the intracellular electrical potential, the membrane potential (Em). The ion-selective barrel, indicated by the plug of ion exchanger in the tip, records the sum of the membrane potential and a potential , related to the chemical potential of the ion in question (of activity a) (see eqn [1]). / is obtained by electronic subtraction. The influence on from other ions (indicated by index j and the valencies Zy) can be obtained from calibration.

See other pages where Amplifiers reference wire is mentioned: [Pg.287]    [Pg.287]    [Pg.231]    [Pg.283]    [Pg.286]    [Pg.292]    [Pg.426]    [Pg.96]    [Pg.295]    [Pg.1317]    [Pg.53]    [Pg.207]    [Pg.195]    [Pg.247]    [Pg.295]    [Pg.57]    [Pg.225]    [Pg.410]    [Pg.19]    [Pg.197]    [Pg.188]    [Pg.57]    [Pg.295]    [Pg.91]    [Pg.232]    [Pg.338]    [Pg.20]    [Pg.577]    [Pg.272]    [Pg.398]    [Pg.410]    [Pg.219]    [Pg.829]    [Pg.426]    [Pg.3043]    [Pg.96]    [Pg.567]    [Pg.627]    [Pg.303]   
See also in sourсe #XX -- [ Pg.283 ]




SEARCH



Amplifiers

© 2024 chempedia.info