Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Acquisition of spectra

The spectra for the samples in the above-described set are typically recorded on the same equipment that will subsequently be used for the measurement of the unknown samples. Such spectra will include the variability in the instrument. The essential prerequisites for constracting the calibration model are that the spectrum should contain the information required to predict the property of interest and that the contribution [Pg.474]

The spectra for the samples in the above-described set are typically recorded on the same equipment that will subsequently be used for the measurement of the unknown samples. Such spectra will include the variability in the instrument. The essential prerequisites for constructing the calibration model are that the spectrum should contain the information required to predict the property of interest and that the contribution of such a property to the spectrum should be much greater than that of the noise it may contain. Some authors recommend recording the spectra for the calibration samples in a more careful manner than those for the other samples in order to minimize their noise (e.g. by strictly controlling the environmental conditions, averaging an increased number of scans, recording spectra on different days) on the assumption that a better model will result in better predictions - even if the latter spectra are noisier. [Pg.377]


Somsen et al. [796] have reported the use of SERR spectroscopy for the in situ selective determination and semi-quantitative analysis of structurally similar dyes separated by TLC. The limits of identification of the TLC-SERRS method (ca. 5ng applied) were sufficient for acquisition of spectra of impurities present in the certified dye standards. SERRS may also be used for in situ identification of highly fluorescent molecules on HPTLC plates. [Pg.538]

The parameters K1/ K2/ and K3 are defined by the refractive indices of the crystal and sample and by the incidence angle [32]. If the sample has uniaxial symmetry, only two polarized spectra are necessary to characterize the orientation. If the optical axis is along the plane of the sample, such as for stretched polymer films, only the two s-polarized spectra are needed to determine kz and kx. These are then used to calculate a dichroic ratio or a P2) value with Equation (25) (replacing absorbance with absorption index). In contrast, a uniaxial sample with its optical axis perpendicular to the crystal surface requires the acquisition of spectra with both p- and s-polarizations, but the Z- and X-axes are now equivalent. This approach was used, through dichroic ratio measurements, to monitor the orientation of polymer chains at various depths during the drying of latex [33]. This type of symmetry is often encountered in non-polymeric samples, for instance, in ultrathin films of lipids or self-assembled monolayers. [Pg.310]

FT Spectrometers FT spectrometers (Figure 3) differ from scanning spectrometers by the fact that the recorded signal is an interferogram [14] (see Chapter 6.2). They can be coupled to a microscope or macrochamber with an FPA detector. FT chemical imaging systems (CISs) are available for Raman, NIR, and IR spectroscopy. However, they can only be considered as research instruments. For example, most IR imaging systems are FT spectrometers coupled to microscopes. This type of spectrometer allows the acquisition of spectra in reflection, attenuated total reflection (ATR), or transmission mode. [Pg.414]

An NMR spectrometer under the control of software which allows the automatic acquisition of spectra with solvent suppression. Communication with the chromatography software is also necessary. [Pg.44]

This pseudo-double beam spectrometer requires that the source be stable over the measurement of the three required arrays. It also consumes a large amount of computational time forbidding the very rapid acquisition of spectra. [Pg.140]

S solid-state MAS spectra at 19.6 T have been acquired for some bulk compounds representative of sulphate speciation in cement paste ettringite, [Ca6Al2(S04)3(0H)6 26H20],107,108 and three different samples of gypsum, [CaS04 -2H20].108 At 33S natural abundance, the acquisition of spectra with a... [Pg.42]

Fig. 3 The measurement of FTIR images does not require much sample preparation. Acquisition of spectra is very fast due to the huge amount of elements in the detector array. By far most time is spent for the evaluation of the molecular data. The figure gives a semi-quantitative impression of the size of the time slices... Fig. 3 The measurement of FTIR images does not require much sample preparation. Acquisition of spectra is very fast due to the huge amount of elements in the detector array. By far most time is spent for the evaluation of the molecular data. The figure gives a semi-quantitative impression of the size of the time slices...
Simultaneous application of spectroscopy in voltammetry — Numerous spectroelectrochemical methods employ spectrometers fast enough to acquire complete spectra within seconds or fractions of a second. Acquisition of spectra during a slow electrode potential scan as employed in -> cyclic voltammetry or during single scan voltammetry can thus be related to a single electrode potential (within a range of a few millivolts or even more narrow). Thus interfacial phenomena as detected with the applied spectroelectrochemical technique can be related to the electrode potential directly even in case of unstable intermediates. [Pg.631]

In many industrial laboratories, Raman spectrscopy is routinely used, together with infrared spectroscopy, for acquisition of vibrational spectra. Raman spectrometer systems for routine analytical applications are commercially available. An important expansion of the potential of the technique has arisen from the use of the microprobe, which permits acquisition of spectra from domains as small as one micron. [Pg.162]

While still insufficient to remove the very strong dipolar interactions between protons in organic solids, this spinning rate allows the acquisition of spectra of perfluorinated polymers, such as Teflon. [Pg.7]

Broad resonance lines, accompanied by Tj relaxation times much longer than the T2 relaxation times, are indications that the fluorine nucleus is in a restricted-mobility environment [14, 20, 21, 71]. The specific MR characteristics of a compound in its in vivo state must be evaluated and the acquisition and analysis adjusted to account for the values of these characteristics prior to performing human 19F spectroscopy studies. Effective T2 values (T2 ) that are shorter than 2 ms have spectra that are difficult to measure using whole-body MR systems. In general, acquisition of spectra that contain short T2 -value species require the use of short (<0.5 ms) excitation RF pulses and careful attention... [Pg.499]

In solid-state NMR, assignment of resonances to specific nuclei relies upon correlation spectroscopy. This involves the acquisition of spectra in which the transfer of magnetization between nearby nuclei can be traced. The patterns of transfer may then be used to assign the individual resonances to the correct positions in the protein. Three mechanisms for magnetization transfer are available through-space, through-bond or relaxation mediated. [Pg.131]

The data system also provides the opportunity of ejecting a mass of choice from the cell just prior to spectrum acquisition, thus permitting the acquisition of spectra from which specific peak related attenuation or interference has been removed. This type of peak removal may remove whole interference spectra, if the parent mass of a compound is chosen for ejection initially. [Pg.282]

For measurement of steady-state fluorescence, the rapid acquisition of spectra is a convenience but not essential. However, the rapid acquisition of spectral information is essential for measurement of transient species or where the luminescence signal continually changes with time. This will be demonstrated first for kinetic-based luminescence measurements and then chemiluminescence measurements. [Pg.163]

Fluorescence Kinetic-Based Measurements. Our studies of the reaction rate determination of thiamine (vitamin Bl) will be used to demonstrate the unique capabilities of rapid acquisition of spectra in kinetic measurements. The kinetic method is based on the oxidation of thiamine by Hg + in basic solutions to highly fluorescent thiochrome (16) The initial rate, taken as the change in fluorescence signal at 444 nm that occurs in a fixed time after mixing the sample and reagents, is directly proportional to the thiamine concentration. [Pg.163]

Checking water quality for solution with acquisition of spectra against air... [Pg.268]

Checking solvent quality with acquisition of spectra against air and water (Abs < 0.1 at 200 nm if not another wavelength window)... [Pg.268]

Pholoacouslic spectroscopy (PAS) provides a way to obtain ultraviolet, visible, and IR absorption. spectra of solids, semisolids, or turbid liquids. Acquisition of spectra for these materials by ordinary methods is usually difficult at best and often impossible because of light scattering and reflection. [Pg.472]


See other pages where Acquisition of spectra is mentioned: [Pg.211]    [Pg.53]    [Pg.146]    [Pg.119]    [Pg.172]    [Pg.474]    [Pg.73]    [Pg.211]    [Pg.377]    [Pg.245]    [Pg.154]    [Pg.261]    [Pg.64]    [Pg.391]    [Pg.96]    [Pg.56]    [Pg.159]    [Pg.318]    [Pg.1121]    [Pg.111]    [Pg.171]    [Pg.144]    [Pg.172]    [Pg.137]    [Pg.48]    [Pg.144]    [Pg.499]    [Pg.296]    [Pg.393]   


SEARCH



Determination of NMR Spectra Acquisition Parameters

Spectrum acquisition

© 2024 chempedia.info