Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Absolute flow rate, calibration

Absolute flow rate, calibration, 132 Acrylic polymers, membrane viscometer characterization, 163,165/... [Pg.299]

Experimental variables such as temperature, flow rate, sample concentration and mobile phase composition can cause changes in the elution volume of a polymer [439,457,460-464]. Chromatographic measurements made with modem equipment are limited more by the errors in the absolute methods used to characterize the molecular weight of the calibration standards than any errors Inherent in the measurements themselves, since the determination of molecular weights by SEC is not an absolute method and is dependent on calibration [462]. The Influence of temperature on retention in SEC is not very great, since no strong sorptive interactions are involved in the retention mechanism. Temperature differences between the column and solvent delivery... [Pg.228]

These investigations showed that the T04 olfactometer can deliver repeatable dilutions, but that an absolute calibration was needed before use. The main problems arise from the flow rates measured in the lower ranges of the rotameter scales, which are less accurate than the upper ranges. If the upper ranges only are used, then consistent results should be obtainable. The olfactometer should be used in the same orientation as when calibrated. [Pg.140]

The NO concentration measurements were made using a chemiluminescence analyzer calibrated with 89 ppm standard mixture of NO in N2. A choked flow orifice controls the sample flow rate through the analyzer and therefore the probe is not choked during sampling for NO measurements. The pressure drop across the analyzer is approximately 80 kPa and the exit of the analyzer is operated at 10 kPa absolute pressure. [Pg.444]

As is the case for LIF, calibration to obtain absolute concentrations is a challenge. In the instrument shown in Fig. 11.45, a calibration source based on the photolysis of water at 185 nm is installed in the inlet. From the absorption cross section of HzO gas at 185 nm, its concentration, the light intensity, and the sample flow rate, the concentration of OH generated by the photolysis can be calculated. However, not only is there significant uncertainty in the absorption cross section for HzO at 185 nm (e.g., see Lazendorf et al., 1997 Hofzumahaus et al., 1997, 1998 and Tanner et al., 1997), but the measured calibration factor was highly variable from day to day, by as much as a factor of two (Tanner et al., 1997). [Pg.602]

The calculation of the relative characteristic peak areas on the chromatograms of the volatile pyrolysis products, using an external standard irrespective of the pyrolysis procedure, permits one to take into account the sensitivity of the detector, with easy computation of the ratio between the peak areas of the component of interest and the standard which, under normal conditions (sample size, carrier gas flow-rate, pyrolysis temperatures, etc.) are proportional to the absolute amounts of the pyrolysis products. This method of calculation is essentially a modification of the absolute calibration method in gas chromatography, which had never been used before in Py—GC.To facilitate comparison of the results obtained at different times or on different instruments, the results of individual measurements should preferably be presented in terms of specific yields (or relative characteristic peak areas), i.e., the yield of the volatile pyrolysis products must be calculated per 1 mg (or g or ng) of the pyrolysed sample with respect to 1 mg (or g or Mg) of the external standard. Such a calculation makes sense in the range of sample sizes which affect only insignificantly the specific yield of light pyrolysis products. [Pg.126]

During testing in the glass furnace environment, the detector assembly was cooled using a vortex tube cooler, with an adjustable air flow rate, in order to minimize the temperature excursions. Additionally, the temperature of the detector assembly was measured and documented for later use in data reduction. A statistical analysis of all calibration data including the temperature dependence of the detectors reveals an estimated uncertainty in measured temperature of +2% of the total measured absolute temperature. [Pg.678]

All of the calibration gas generators for laboratory use work dynamically, ie, according to the dosing principle. The desired gases are dosed with definite flow rates and then mixed inside the apparatus. With the technical equipment available in laboratories, standard volumetric flow rates (V,) or molar flow rates (F) can be proportioned much more precisely than concentrations or absolute quantities (gas suppliers, however, often manufacture gas mixtures by a weighing-in process). Calibration gas generators are offered on the basis of three different methods of gas dosage (to be described in detail later). [Pg.152]

Differential scanning calorimetry is not an absolute measuring technique, calibrations are thus of prime importance. Calibrations are necessary for the measurement of temperature, T (in K) amplitude, expressed as temperature difference, AT (in K) or as heat-flow rate, dQ/dt (in J s or W) peak area AH (in J) and time, t (in s or min). Figure 4.62 shows the analysis of a typical first-order transition, a melting transition. [Pg.338]

The common problem with measuring reflux rate is that reflux meters are typieaUy set at startup and then never adjusted again. Therefore, the reflux flow rate is typically not reliable. The reflux ratio is checked and monitored as an important operating parameter, but the absolute value of the reflux rate is rarely monitored. However, to have a correct heat balance, the reflux flow meter must be checked and calibrated to achieve at least 5% closure of heat balance ([total heat input total heat output]/total heat input). Only with this accuracy of heat balance, tray efficiency can be accurately determined (Summers, 2009). [Pg.292]

The GPC analysis for PHEMA was made on a Tosoh CCP 8020-series high-speed liquid chromatograph (Tokyo, Japan) equipped with two Shodex gel columns LF804 (300 X 80 mm bead size = 6 pm pore size = 20-3000 A) (Tokyo). DMF was used as eluent with a flow rate of 0.8 mlV min (40 °C). The column system was calibrated with Tosoh standard polyethyleneglycols (PEGs). As an absolute number-average molecular weight of... [Pg.203]

The molecular weight distributions of the oligomers were determined by means of gel permeation chromatography. These were done by Dr. Julian F. Johnson who was then at the Chevron Research Company. A Waters Analytical G.P.C. Model 300 was used. A combination of one 100,000 R, one 15,000 A, one 100 R, and one 45 R column was used. The columns were calibrated with normal alkanes and monodisperse polystyrenes. Samples were dissolved in toluene to obtain 60 mg/100 ml concentration they were eluted with toluene at a flow rate of 1 ml per minute at room temperature. The results were machine-computed to obtain relative molecular weights. From these, absolute values of the molecular weights were obtained by means of scaling factors calculated from experimental viscosity or vapor pressure osmometry data. The calibration curve is shown... [Pg.105]

The calibration of in-line mass flow meters is an important issue. In one paper the authors argue that as direct measurement of instantaneous mass flow rate is measured it is absolutely essential that a standard mass flow rate sensor is used - rather than the weighing of accumulated product [11]. [Pg.801]

The calibration curve, strictly speaking, applies only to the particular polymer, solvent, temperarnre, flow rate, and column for which it was established. Change any one, and the calibration is no longer valid. Most SEC calibrations are obtained with polystyrene, because the necessary monodisperse standards are readily available at reasonable cost. What do you do when you want to analyze another polymer If absolute values are not needed, the molecular weights can be reported with respect to the polystryene (or other polymer)... [Pg.82]

Here, = tp-is the difference between the retention time of the solute, tp and of the marker, t . Air is usually used as a marker, when the Thermal Conductivity (TC) detector is used, to account for the dead volume in the chromatographic column. The retention time of the marker has to be subtracted from the solute retention time to reflect the absolute value of the solute retention time as tp. F is the flow rate of the carrier gas measured at room temperature 7V, w is the mass of the stationary phase, and J is a pressure correction factor which depends on /, and P , the inlet and outlet pressures respectively. Pi and P are measured using electronic transducers which are interfaced at the inlet and outlet of the column. These transducers are usually calibrated using a mercury manometer. To calculate the interaction parameter, xn, of the polymer-solute system, Vg from Eqn 1 is utilized as follows ... [Pg.324]


See other pages where Absolute flow rate, calibration is mentioned: [Pg.130]    [Pg.132]    [Pg.560]    [Pg.145]    [Pg.810]    [Pg.227]    [Pg.92]    [Pg.129]    [Pg.4]    [Pg.171]    [Pg.294]    [Pg.56]    [Pg.211]    [Pg.197]    [Pg.354]    [Pg.157]    [Pg.242]    [Pg.776]    [Pg.153]    [Pg.947]    [Pg.100]    [Pg.51]    [Pg.900]    [Pg.17]    [Pg.650]    [Pg.4]    [Pg.837]    [Pg.339]    [Pg.51]    [Pg.15]    [Pg.9]    [Pg.158]    [Pg.116]    [Pg.64]   
See also in sourсe #XX -- [ Pg.132 ]




SEARCH



Absolute calibration

Absolute rate

© 2024 chempedia.info