Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Visible-ultraviolet spectroscopy polarity

Microscopy (qv) plays a key role in examining trace evidence owing to the small size of the evidence and a desire to use nondestmctive testing (qv) techniques whenever possible. Polarizing light microscopy (43,44) is a method of choice for crystalline materials. Microscopy and microchemical analysis techniques (45,46) work well on small samples, are relatively nondestmctive, and are fast. Evidence such as sod, minerals, synthetic fibers, explosive debris, foodstuff, cosmetics (qv), and the like, lend themselves to this technique as do comparison microscopy, refractive index, and density comparisons with known specimens. Other microscopic procedures involving infrared, visible, and ultraviolet spectroscopy (qv) also are used to examine many types of trace evidence. [Pg.487]

The symmetry of the LB films was determined by polarized ultraviolet-visible (UV-Vis) absorption spectroscopy, optical rotation, and second-harmonic generation. All studies showed that the constructed LB films are anisotropic in the plane of the film and that the symmetry of the film is C2 with the twofold rotation axis perpendicular to the film plane. For example, when the SH intensity is plotted as a function of the azimuthal rotation angle (rotation around an axis perpendicular to the plane of the film), the twofold symmetry becomes evident (Figure 9.23). Isotropic films generate an SH signal independent of the azimuthal rotation angle. On the other hand, the LB... [Pg.559]

There have been relatively little ultraviolet-visible (UV-Vis) spectroscopic data for 1,4-oxazines, but selected data are presented in Table 8. UV spectroscopy is important for photochromic compounds, such as spirooxazines. The UV spectra of 33 spirooxazines in five different solvents are collected in a review <2002RCR893>, and the more recently reported examples of photochromic oxazines 65, 66, 101, and 102 are shown here. It can be seen from Table 8 that both adding methoxy substituents to the oxazine and changing to a more polar solvent give a UV maximum at a higher wavelength. This solvent effect can also be seen in the case of 102, which also has important fluorescence properties, discussed in Section 8.06.12.2. [Pg.471]

The tautomerism of 4 (Figure 1) was also studied by UV-Vis (ultraviolet-visible) spectroscopy in polar aprotic solvents the effect of added water, darkness, and indirect sunlight were also evaluated. The experimental spectroscopic results are discussed in Section 13.14.3.1.1 (i). Theoretical calculations using ZINDO/S were performed to state the allowed absorption transitions <2005SAA875>. The five tautomeric structures of 4 as well as the calculated energies for each are depicted in Figure 2. [Pg.401]

These devices are based on the anisotropic absorption of light. Usually molecular crystals exhibit this property and tourmaline is the classical example for this. For practical purposes, however, micro crystals are oriented in polymer sheets. Polymers containing chromophors become after stretching dichroic polarizers. The devices produced in this manner are called polawids. They have found a broad application in many technologies. Their application in spectroscopy is limited to the near ultraviolet and to the visible and near infrared range of the spectrum. In vibrational spectroscopy polaroids are employed as analyzers only for Raman spectroscopy. [Pg.91]

Volume 50 of Advances in Catalysis, published in 2006, was the hrst of a set of three focused on physical characterization of solid catalysts in the functioning state. This volume is the second in the set. The hrst four chapters are devoted to vibrational spectroscopies, including Fourier transform infrared (Lamberti et al.), ultraviolet Raman (Stair), inelastic neutron scattering (Albers and Parker), and infrared-visible sum frequency generation and polarization-modulation infrared rehection absorption (Rupprechter). Additional chapters deal with electron paramagnetic resonance (EPR) (Bruckner) and Mossbauer spectroscopies (Millet) and oscillating microbalance catalytic reactors (Chen et al.). [Pg.392]

CD spectroscopy measures the difference in absorbance between left- and right-circularly polarized incident radiation, in the near ultraviolet to visible and infrared regions of the spectrum. [Pg.6437]

A solvent for ultraviolet/visible spectroscopy must be transparent in the region of the spectrum where the solute absorbs and should dissolve a sufficient quantity of the sample to give a well-defined analyte spectrum. In addition, we must consider possible interactions of the solvent with the absorbing species. For example, polar solvents, such as water, alcohols, esters, and ketones, tend to obliterate vibration spectra and should thus be avoided to preserve spectral detail. Nonpolar solvents, such as cyclohexane, often provide spectra that more closely approach that of a gas (compare, for example, the three spectra in Figure 24-14). In addition, the polarity of the solvent often influences the position of absorption maxima. For qualitative analysis, it is therefore important to compare analyte spectra with spectra of known compounds measured in the same solvent. [Pg.788]

We have seen that the enantiomers of a molecule have the same properties (melting and boiling points, refractive index, etc.) and that spectroscopy that does not involve polarized light is incapable of distinguishing between them, so that their infrared, NMR, ultraviolet-visible and Raman spectra are identical. On the other hand, the optical rotation, optical dispersion and circular dichroism give results that are opposite in sign for the two enantiomers. We have also seen that in certain cases it is possible to determine the absolute configuration of a molecule in the crystal by X-ray diffraction. [Pg.48]

Rao and co-workers [82] used an inverted emulsion process for the synthesis of the emeraldine salt of PAM using a novel oxidising agent, benzoyl peroxide. The polymerisation was carried out in a non-polar solvent in the presence of four different protonic acids as dopants and an emulsifier (sodium lauryl sulfate). The polymer salts were characterised spectroscopically by ultraviolet-visible, Fourier-transform infrared, Fourier-transform Raman and electron paramagnetic resonance spectroscopy. Thermogravimetric analysis, was used to determine the stability of the salts and the activation energy for the degradation. The conductivity of the salts was found to be in the order of 10 S/cm. [Pg.112]

The monomers and polymers were characterized by infrared (IR) and ultraviolet-visible (UV-vis) spectroscopy, H and - C nuclear magnetic resonance (NMR), element analyses, differential scanning calorimetry (DSC), and polarizing optical microscopy. The molecular weights of the polymers were evaluated by gel permeation chromatography (GPC) using polystyrene standards, and electrical conductivities upon iodine doping for the cast film of the polymers were measured by the four-probe method. [Pg.1000]

Circular dichroism (CD) is a chiroptical spectroscopy that measures the differential absorption of left versus right circularly polarized light. Its higher sensitivity to molecular conformation and configuration has made CD spectroscopy a more powerful tool in the structural analysis of various chiral supramolecular systems than its parent achiral absorption spectroscopies such as ultraviolet (UV), visible (vis), and infrared (IR) spectra (Figure 1). CD measurements in the UV and vis regions are the most widely... [Pg.455]

In ordinary infrared measurements, the requirements for the optical properties of cell windows and other materials (flatness, transparency, etc.) are not as severe as needed for ultraviolet and visible spectroscopy. Sometimes it is even necessary to make the window surface slightly coarse or to make its surfaces slightly deviated from parallel to avoid interference effects. However, such measures should never be taken in VCD measurements in order to ensure that circularly polarized radiation may be generated without fail. In VCD measurements, cell windows must be flat and parallel. A mirror cannot be used for any purpose. Samples should be transparent. VCD will not be observed in samples such as colloids because of scattering. [Pg.329]


See other pages where Visible-ultraviolet spectroscopy polarity is mentioned: [Pg.180]    [Pg.596]    [Pg.174]    [Pg.379]    [Pg.6367]    [Pg.7]    [Pg.6366]    [Pg.114]    [Pg.309]    [Pg.8]    [Pg.77]    [Pg.121]    [Pg.120]    [Pg.76]    [Pg.1770]    [Pg.65]    [Pg.120]    [Pg.196]    [Pg.78]    [Pg.996]    [Pg.19]    [Pg.151]   
See also in sourсe #XX -- [ Pg.667 ]




SEARCH



Polarization spectroscopy

Spectroscopy ultraviolet visible

Ultraviolet spectroscopy

Ultraviolet-visible

© 2024 chempedia.info