Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

The Sample Measurement

Excluding the incorrect use of the instrumentation or its malfunctioning, the most common errors in UV-vis spectroscopic measurements are due to the composition and/or to the treatment of the sample. [Pg.59]


The take-home lesson is that the vast majority of high-pressure studies are on solids or other rigid media and are not done under hydrostatic conditions. The stresses and stress-related properties may vary throughout the sample. Unless the probes are very local and focus on a small region of the sample, measurements are averages over a range of, often uncharacterized, conditions. [Pg.1956]

The maximum power of a conventional X-ray tube is 2.4 kW for broad focus (approx.. 2x 12 mm focal spot size). Modern rotating anodes consume 18 kW and deliver fine focus (approx.. 0.1 x 1 mm focal spot size). Most important for high intensity is not the power consumption, but the product of focal spot power density and focal spot size or, more accurately, the flux on the sample measured in photons/s (cf. Sect. 7.6). [Pg.60]

Depending on the type of instrument, the reference measurement (top diagram) might be made simultaneously with the sample measurement (bottom diagram) or a reference measurement might be saved on computer to generate the full spectrum. [Pg.82]

Figure 3. Mach-Zehnder interferometer One of the arms is covered by inert material (reference arm), the other arm exposes to the sample (measuring arm). Figure 3. Mach-Zehnder interferometer One of the arms is covered by inert material (reference arm), the other arm exposes to the sample (measuring arm).
The error from pH measurement systems depends on the slope of the calibration line. For a pH sensor with a slope of 59mV/pH, an error of lmV in measuring the electrode potential will cause a 0.017 pH change in the concentration. The lower the slope, the higher the errors are on the sample measurements [39],... [Pg.291]

Heat flow as a function of time during rewarming of the samples measured by DSC. [Pg.221]

At each wavelength, subtract the dark-field measurement from the photometer response measured in step 6 to get the sample measurement. [Pg.142]

At each wavelength, express the sample measurement as a function of the blank measurement. [Pg.142]

It is possible to assess the proportion of stray light by measuring the amount of radiation transmitted by samples that are optically opaque at the wavelength to be assessed but that transmit radiation of other wavelengths. The instrument is set to zero and 100% transmittance in the normal way and the opaque substance introduced into the sample compartment. The amount of light transmitted by the sample, measured in percentage transmittance, is... [Pg.51]

How about measurement control Usually the measurement control process involves identical samples, which are introduced into the sample measurement sequence in order to monitor the stability of the system. [Pg.299]

The samples measured in Ref. [34) are probably strongly contaminated by K2Fe04. Some measurements of K3Fe04 [146) indicate that the vz band hes at approximately 760 cm-i. Accordingly, the corresponding vi should appear around 780—800 cm i. Hence, the n value given in Ref. (34) could be approximately correct, however, the same cannot be inferred for the vz value. Furthermore, VI >i 3 must be vahd as well for Fe04. ... [Pg.93]

The residuals are the portion of the sample measurement vector that is not explained using a given number of PCs. [Pg.55]

The residuals are that portion of the sample measurement vector not fit by the PCA model for a given class. [Pg.266]

For runs at MIT the splashing caused by the spray tubes resulted in difficulties in measuring the beam current, owing to electrical shorting. The spray tubes were replaced by two trickle tubes, placed about 1/16 to 1/8 inch from the broad sides of the sample. Measurement of the temperature in the center of polystyrene samples showed the same temperature rise for the two cooling systems, indicating that the same sample temperature profile was obtained in each case. The heat removed by the water is in the range of 0.1 to 0.3 cal. per second. [Pg.91]

Table 2.6. Critical 95% and 99% values of the Cochran statistic for k laboratories each repeating the sample measurement n times... Table 2.6. Critical 95% and 99% values of the Cochran statistic for k laboratories each repeating the sample measurement n times...
Wavelength accuracy is defined as the deviation of the wavelength reading at an absorption band or emission band from the known wavelength of the band. The wavelength deviation can cause significant errors in the qualitative and quantitative results of the UV-Vis measurement. It is quite obvious that if the spectrophotometer is not able to maintain an accurate wavelength scale, the UV absorption profile of the sample measured by the instrument will be inaccurate. The true Amax and A.min of the analyte cannot be characterized accurately. [Pg.155]

Additional Sample Measurement Requirements. The sample measurement should ideally be made under the same conditions as the electrode calibration. [Pg.239]


See other pages where The Sample Measurement is mentioned: [Pg.1146]    [Pg.467]    [Pg.613]    [Pg.78]    [Pg.492]    [Pg.84]    [Pg.462]    [Pg.309]    [Pg.26]    [Pg.228]    [Pg.7]    [Pg.49]    [Pg.172]    [Pg.837]    [Pg.155]    [Pg.199]    [Pg.35]    [Pg.267]    [Pg.277]    [Pg.221]    [Pg.101]    [Pg.410]    [Pg.43]    [Pg.156]    [Pg.162]    [Pg.837]    [Pg.20]   


SEARCH



Loading sample strontium solution on a filament for measurement in the thermal ionization mass spectrometer (TIMS)

Measuring sample

Sample measurements

The Sample

© 2024 chempedia.info