Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Signal ratio method

Another method for finding class-distinguishing features, but not necessarily between known classes, is the signal ratio method SraUo method). The details of the method will not be presented here, but it is designed to find the chromatographic (and mass spectral) locations of greatest difference, measured as a ratio, in signal intensity across all samples and mass channels in a set of... [Pg.120]

A sixth spectrophotometric method for the quantitative determination of Pb + levels in blood uses CQ+ as an internal standard. A standard containing 1.75 ppb Pb + and 2.25 ppb CQ+ yields a ratio of Sa/Sis of 2.37. A sample of blood is spiked with the same concentration of Cu +, giving a signal ratio of 1.80. Determine the concentration of Pb + in the sample of blood. [Pg.116]

When the fluorescence spectra of the probe shifts on protonation, two emission wavelengths with opposite proton-sensitive response are chosen to give a pH-dependent emission intensity ratio. In this ratio method a number of ion-independent factors that affect the signal intensity like photobleaching, variations in probe concentration, and illumination instability are eliminated. [Pg.128]

In general it can be shown that the approach following the signal-to-noise ratio method is mathematically very similar to the standard formulation of the chi-square test using observed and expected frequencies, and in practice they will invariably give very similar results. Altman (1991) (Section 10.7.4) provides more detail on this connection. [Pg.67]

Standards used for this ratio method are generally quinine sulphate, rhodamine B or 2-aminopyridine solutions. In order to eliminate fluctuations due to the radiation source and other instrument parameters, comparative measurements are used. Instrument sensitivity is commonly expressed in terms of the signal to noise ratio of the Raman band of water (cf. 12.4). The nature of the solvent, the temperature,... [Pg.228]

The intense absorption of water over most of the infrared spectrum restricts the regions where aqueous solutions of carbohydrates can be usefully studied. Absorbance subtraction makes it possible to eliminate water absorbance and magnify the remaining spectral features to the limit of the signal-to-noise ratio. Many other data-processing techniques, such as the ratio method,4 the least-squares refinement,5 and factor analysis,6 should be of benefit in the study of carbohydrate mixtures. [Pg.7]

Eig. 5. Several endpoint detection methods were compared for the detection of immuno-polymerase chain reaction (IPCR) amplificate from a direct IPCR (Fig. 3A) of mouse-IgG. Although all IPCR/DNA-detection combinations were able to improve the detection limit of a comparable enzyme-linked immunosorbent assays (ELISA) of approximately 10 amol IgG in a 30-fL sample volume, several differences were observed in actual detection limit, and the linearity of the concentration/signal ratio dependent on the DNA quantification was applied. Best results were obtained for PCR-ELISA (see also Fig. 6) in combination with fluorescence- or chemiluminescence-generating substrates (b, c). With photometric substrates (d) or gel electrophoresis and subsequent spot densitometry (a), a 10-fold decrease in sensitivity was observed. In addition to the more sigmoid curve in gel electrophoresis, an enhanced overall error of 20% compared to 13% in PCR-ELISA was observed for two independent assays. The simple addition of a double-strand sensitive intercalation marker to the PCR-amplificate and measurement in a fluorescence spectrometer further decreased sensitivity (e) and appears therefore to be unsuited for IPCR amplificate quantification. (Figure modified according to references 37 and 65.)... [Pg.260]

Internal standards could be used in external calibration, matrix-matched external calibration, and standard addition calibration [2], However, the use of internal standards in LC-MS quantitative methods should not be confused with internal calibration in which an internal standard is employed as a calibrant and the concentration of a unknown sample is calculated from the concentration of this internal standard and its analyte/IS signal ratio, i.e., the concentration of the unknown sample is calculated without the need for a calibration curve [3], The use of internal standards in most LC-MS quantitative methods belongs to signal-ratio calibration or internal standardization [2,4], In fact, the majority of bioanalytical LC-MS methods use matrix-matched signal-ratio external calibration. [Pg.2]

All of these error sources, other than the minimal intrinsic error, can be reduced by using the internal standard method. The absolute measurement of the signal is replaced by the measurement of the signal ratio for the molecule that is measured and for the internal standard. The same compound can play the role of the internal standard for the quantification of various compounds within the mixture. [Pg.266]

Fig. 3.6 Internal standard method RJRst)o- RJRsd-ti signal ratios measured for analyte and internal standard in a set of standard solutions, Cj-Cn analyte concentrations in the standard solutions, (RJRsdx signal ratio measured for analyte and internal standard in the sample, calculated analyte concentration in the sample... Fig. 3.6 Internal standard method RJRst)o- RJRsd-ti signal ratios measured for analyte and internal standard in a set of standard solutions, Cj-Cn analyte concentrations in the standard solutions, (RJRsdx signal ratio measured for analyte and internal standard in the sample, calculated analyte concentration in the sample...
The fundamental advantage of the internal standard method is the possibility of increasing the precision of the analytical result, as the signal ratio can be measured with greater precision than individual signals. An essential condition is that the... [Pg.33]

The use of online data together with steady-state models, as in Real Time Optimization applications, requires the identification of steady-state regimes in a process and the detection of the presence of gross errors. In this paper a method is proposed which makes use of polynomial interpolation on time windows. The method is simple because the parameters in which it is based are easy to tune as they are rather intuitive. In order to assess the performance of the method, a comparison based on Monte-Carlo simulations was performed, comparing the proposed method to three methods extracted from literature, for different noise to signal ratios and autocorrelations. [Pg.459]

In addition to the glass transition, which is also called the a transition, p relaxation is common to most polymers and copolymers. We attempted to obtain T in a similar fashion as with Tg using DSC, but our efforts were in vain due to a high noise/signal ratio. Thermomechanical analysis (TMA) offered an alternative in this case, although the method was not suitable for measuring the Tg of the sample in a state other than bone-dry. Illustrated in Figure 7 is... [Pg.19]


See other pages where Signal ratio method is mentioned: [Pg.446]    [Pg.429]    [Pg.376]    [Pg.268]    [Pg.227]    [Pg.26]    [Pg.53]    [Pg.66]    [Pg.155]    [Pg.2]    [Pg.90]    [Pg.419]    [Pg.89]    [Pg.76]    [Pg.212]    [Pg.283]    [Pg.391]    [Pg.418]    [Pg.1077]    [Pg.75]    [Pg.123]    [Pg.166]    [Pg.67]    [Pg.44]    [Pg.142]    [Pg.819]    [Pg.334]    [Pg.240]    [Pg.105]    [Pg.299]    [Pg.21]    [Pg.15]    [Pg.118]    [Pg.418]    [Pg.459]    [Pg.1280]   


SEARCH



Ratio methods

© 2024 chempedia.info