Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Process mass analyzers

Metastable Peaks. If the mass spectrometer has a field-free region between the exit of the ion source and the entrance to the mass analyzer, metastable peaks m may appear as a weak, diffuse (often humped-shape) peak, usually at a nonintegral mass. The one-step decomposition process takes the general form ... [Pg.814]

A mass spectrometer consists of four basic parts a sample inlet system, an ion source, a means of separating ions according to the mass-to-charge ratios, ie, a mass analyzer, and an ion detection system. AdditionaUy, modem instmments are usuaUy suppUed with a data system for instmment control, data acquisition, and data processing. Only a limited number of combinations of these four parts are compatible and thus available commercially (Table 1). [Pg.539]

Model output validation is essential to any soil modeling effort, although this term has a broad meaning in the literature. For the purpose of this section we can define validation as the process which analyzes the validity of final model output, namely the validity of the predicted pollutant concentrations or mass in the soil... [Pg.62]

For the characterization of compounds extracted from plants, wool and dye baths, acquisition in the NI mode is used. The main signals in the mass spectra of each colourant are attributed to deprotonated molecular ions [M H]. More detailed studies can be performed by ESI MS" with a quadrupole ion trap mass analyzer, and such a set-up was used e.g. for the investigation of photo-oxidation processes of components of weld and onion skins.[29]... [Pg.375]

Multiple mass analyzers exist that can perform tandem mass spectrometry. Some use a tandem-in-space configuration, such as the triple quadrupole mass analyzers illustrated (Fig.3.9). Others use a tandem-in-time configuration and include instruments such as ion-traps (ITMS) and Fourier transform ion cyclotron resonance mass spectrometry (FTICRMS or FTMS). A triple quadrupole mass spectrometer can only perform the tandem process once for an isolated precursor ion (e.g., MS/MS), but trapping or tandem-in-time instruments can perform repetitive tandem mass spectrometry (MS ), thus adding n 1 degrees of structural characterization and elucidation. When an ion-trap is combined with HPLC and photodiode array detection, the net result is a profiling tool that is a powerful tool for both metabolite profiling and metabolite identification. [Pg.47]

Fig. 11.5. Diagram illustrating the components of an ESI source. A solution from a pump or the eluent from an HPLC is introduced through a narrow gage needle (approximately 150 pm i.d.). The voltage differential (4-5 kV) between the needle and the counter electrode causes the solution to form a fine spray of small charged droplets. At elevated flow rates (greater than a few pl/min up to 1 ml/min), the formation of droplets is assisted by a high velocity flow of N2 (pneumatically assisted ESI). Once formed, the droplets diminish in size due to evaporative processes and droplet fission resulting from coulombic repulsion (the so-called coulombic explosions ). The preformed ions in the droplets remain after complete evaporation of the solvent or are ejected from the droplet surface (ion evaporation) by the same forces of coulombic repulsion that cause droplet fission. The ions are transformed into the vacuum envelope of the instrument and to the mass analyzer(s) through the heated transfer tube, one or more skimmers and a series of lenses. Fig. 11.5. Diagram illustrating the components of an ESI source. A solution from a pump or the eluent from an HPLC is introduced through a narrow gage needle (approximately 150 pm i.d.). The voltage differential (4-5 kV) between the needle and the counter electrode causes the solution to form a fine spray of small charged droplets. At elevated flow rates (greater than a few pl/min up to 1 ml/min), the formation of droplets is assisted by a high velocity flow of N2 (pneumatically assisted ESI). Once formed, the droplets diminish in size due to evaporative processes and droplet fission resulting from coulombic repulsion (the so-called coulombic explosions ). The preformed ions in the droplets remain after complete evaporation of the solvent or are ejected from the droplet surface (ion evaporation) by the same forces of coulombic repulsion that cause droplet fission. The ions are transformed into the vacuum envelope of the instrument and to the mass analyzer(s) through the heated transfer tube, one or more skimmers and a series of lenses.
Fig. 11.16. Representation of three tandem mass spectrometry (MS/MS) scan modes illustrated for a triple quadrupole instrument configuration. The top panel shows the attributes of the popular and prevalent product ion CID experiment. The first mass filter is held at a constant m/z value transmitting only ions of a single mlz value into the collision region. Conversion of a portion of translational energy into internal energy in the collision event results in excitation of the mass-selected ions, followed by unimolecular dissociation. The spectrum of product ions is recorded by scanning the second mass filter (commonly referred to as Q3 ). The center panel illustrates the precursor ion CID experiment. Ions of all mlz values are transmitted sequentially into the collision region as the first analyzer (Ql) is scanned. Only dissociation processes that generate product ions of a specific mlz ratio are transmitted by Q3 to the detector. The lower panel shows the constant neutral loss CID experiment. Both mass analyzers are scanned simultaneously, at the same rate, and at a constant mlz offset. The mlz offset is selected on the basis of known neutral elimination products (e.g., H20, NH3, CH3COOH, etc.) that may be particularly diagnostic of one or more compound classes that may be present in a sample mixture. The utility of the two compound class-specific scans (precursor ion and neutral loss) is illustrated in Fig. 11.17. Fig. 11.16. Representation of three tandem mass spectrometry (MS/MS) scan modes illustrated for a triple quadrupole instrument configuration. The top panel shows the attributes of the popular and prevalent product ion CID experiment. The first mass filter is held at a constant m/z value transmitting only ions of a single mlz value into the collision region. Conversion of a portion of translational energy into internal energy in the collision event results in excitation of the mass-selected ions, followed by unimolecular dissociation. The spectrum of product ions is recorded by scanning the second mass filter (commonly referred to as Q3 ). The center panel illustrates the precursor ion CID experiment. Ions of all mlz values are transmitted sequentially into the collision region as the first analyzer (Ql) is scanned. Only dissociation processes that generate product ions of a specific mlz ratio are transmitted by Q3 to the detector. The lower panel shows the constant neutral loss CID experiment. Both mass analyzers are scanned simultaneously, at the same rate, and at a constant mlz offset. The mlz offset is selected on the basis of known neutral elimination products (e.g., H20, NH3, CH3COOH, etc.) that may be particularly diagnostic of one or more compound classes that may be present in a sample mixture. The utility of the two compound class-specific scans (precursor ion and neutral loss) is illustrated in Fig. 11.17.
What are the six basic elements of a mass spectrometer (sample introduction, ion source, mass analyzer, detector, signal processing and computer). [Pg.399]

Mass bias, or the instrumental mass fractionation, is the variable transmission of the ion beam into the mass spectrometer. A variety of phenomena create conditions that lead to variable transmission of ion beams. For modem instmments, the transmission in the flight tube and the efficiency of ion conversion to electrons at the collector are almost quantitative. Most fractionation processes, therefore, take place within the source, namely in the area where the analyte is introduced into the mass spectrometer and ionized, or at the interface between the source and the mass analyzer. [Pg.114]

Tandem mass spectrometry summarizes the numerous techniques where mass-selected ions (MSI) are subjected to a second mass spectrometric analysis (MS2). [111,112] Dissociations in transit through the mass analyzer may either occur spontaneously metastable. Chaps. 2.7.1, 2.8.2) or can result from intentionally supplied additional activation, i.e., typically from collisions with neutrals. Below, we will chiefly discuss the collision process as such and its consequences for the further fate of the ions. Instrumental aspects of tandem MS are included in Chap. 4 and applications are presented in Chaps. 7-12. [Pg.53]

The properties of an ideal mass analyzer are well described, [2] but despite the tremendous improvements made, still no mass analyzer is perfect. To reach a deeper insight into the evolution of mass spectrometers the articles by Beynon, [3] Habfast and Aulinger, [4,5] Brunnee [6,7], Chapman et al. [8] and McLuckey [9] are recommended for further reading. In recent years, miniature mass analyzers have gained interest for in situ analysis, [10] e.g., in environmental [11] or biochemical applications, [12] for process monitoring, for detection of chemical warfare agents, for extraterrestrial applications, [13] and to improve Space Shuttle safety prior to launch. [14]... [Pg.112]


See other pages where Process mass analyzers is mentioned: [Pg.1029]    [Pg.1029]    [Pg.919]    [Pg.1029]    [Pg.1029]    [Pg.919]    [Pg.3]    [Pg.38]    [Pg.46]    [Pg.135]    [Pg.283]    [Pg.371]    [Pg.391]    [Pg.561]    [Pg.17]    [Pg.37]    [Pg.40]    [Pg.494]    [Pg.992]    [Pg.997]    [Pg.998]    [Pg.329]    [Pg.42]    [Pg.236]    [Pg.75]    [Pg.142]    [Pg.1083]    [Pg.56]    [Pg.15]    [Pg.101]    [Pg.402]    [Pg.341]    [Pg.361]    [Pg.363]    [Pg.381]    [Pg.383]    [Pg.395]    [Pg.11]    [Pg.232]    [Pg.122]    [Pg.47]    [Pg.127]    [Pg.431]   
See also in sourсe #XX -- [ Pg.2 , Pg.322 ]




SEARCH



Mass analyzer

Process analyzers

© 2024 chempedia.info