Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Plasma source spectrometers

Walsh J.N. and Howie R.A., 1980, An evaluation oif the performance of an inductively coupled plasma source spectrometer for the determination of major and trace constituents, , of silicate rocks and minerals. Mineral. Mag., 47, 967-974. [Pg.341]

The weakness of MC-ICPMS lies in the inefficiency by which ions are transferred from the plasma source into the mass spectrometer. Therefore, despite very high ionization efficiencies for nearly all elements, the overall sensitivity (defined as ionization plus transmission efficiencies) of first generation MC-ICPMS instruments is of the order of one to a few permil for the U-series nuclides. For most, this is comparable to what can be attained using TIMS. [Pg.40]

Plasma sources were developed for emission spectrometric analysis in the late-1960s. Commercial inductively coupled and d.c. plasma spectrometers were introduced in the mid-1970s. By comparison with AAS, atomic plasma emission spectroscopy (APES) can achieve simultaneous multi-element measurement, while maintaining a wide dynamic measurement range and high sensitivities and selectivities over background elements. As a result of the wide variety of radiation sources, optical atomic emission spectrometry is very suitable for multi-element trace determinations. With several techniques, absolute detection limits are below the ng level. [Pg.614]

In a typical MIP-MS instrument, the ICP portion is replaced with one of a variety of microwave discharge sources, usually a fairly standardised (modified) Beenakker cavity connected to a microwave generator. The analytical MIP at intermediate power (<500 W) is a small and quiet plasma source compared with the ICP. The mass spectrometer needs no major modifications for it to be interfaced with the MIP. With MIP used as a spectroscopic radiation source, typically consisting of a capillary (1mm i.d.), a power of 30-50 W and a gas flow below 1 L min 1, multi-element determinations are possible. By applying electrodeposition on graphite electrodes, ultratrace element determinations are within reach, e.g. pg amounts of Hg. [Pg.624]

The main advantages of plasma-source mass spectrometry (PS-MS) over other analytical techniques, such as PS-AES and ETAAS, are the possibilities of quantitative isotope determination and isotope dilution analysis the rapid spectral scanning capability of the mass spectrometer and semiquantitative determinations to within a factor of two or three. Several labelling methods are used for the quantification of analytes present in complex mixtures. In these methods, the sample is spiked... [Pg.649]

The outline of the construction of a typical plasma emission spectrometer is to be seen in Figure 8.10. The example shown has an inductively coupled plasma, excitation source, but the outline would be similar were a dc source to be fitted. Different combinations of prisms and diffraction gratings may be used in the dispersion of the emitted radiation, and in the presentation of the analytical signal. Instruments are computerized in operation and make use of automatic sample handling. Sophisticated data handling packages are employed routinely to deal with interferences, and to provide for clarity in data output. [Pg.299]

The recent and meteoric development of plasma source mass spectrometers equipped with high stability sector-field magnets and multiple collectors (MC-ICP-MS) has lead to a... [Pg.156]

The observed range of natural variations of 5 Ca is about 4 to 5%o in terrestrial materials and up to 50%o in high temperature condensate minerals in carbonaceous chondrites. The typical reproducibility of measurements is about +0.15%o. Broader application of Ca isotope measurements in geochemistry may be possible, particularly if the reproducibility can be improved to 0.05%o to 0.03%o. There is hope that this can be achieved either with inductively coupled plasma source mass spectrometry (Halicz et al. 1999) or with a new generation of multi-collector thermal ionization mass spectrometers (Heuser et al. 2002). [Pg.284]

Inductively coupled plasma (ICP) mass spectrometers use a hot plasma to ionize the sample. The plasma is generated by electromagnetic induction. The plasma source operates at... [Pg.527]

Inductively coupled-plasma mass spectrometers (ICPMS) are relatively new to cosmochemistry, although they have been widely used in other fields. The plasma source makes most of the periodic table accessible to measurement, so several radioactive isotope systems that used to be impractical to use for chronology are now routinely used (see Chapters 8 and 9). [Pg.532]

Figure 5.1 Main parts of an inductively coupled plasma mass spectrometer sample introduction systems (left column), e.g., Meinhard or MicroMist nebulizer with cyclonic spray chamber, ultrasonic nebulizer, microconcentric nebulizer and laser ablation system (all from CETAC Technologies), ion source (middle column) and several types of mass spectrometers, (a) Agilent 7500 from Agilent, (b) Platform from CV Instruments, or (c) Element from Thermo Fisher Scientific. (Parts of this figure were reproduced with permission from CETAC Technologies, Agilent, CV Instruments and Thermo Tisher Scientific, respectively.)... Figure 5.1 Main parts of an inductively coupled plasma mass spectrometer sample introduction systems (left column), e.g., Meinhard or MicroMist nebulizer with cyclonic spray chamber, ultrasonic nebulizer, microconcentric nebulizer and laser ablation system (all from CETAC Technologies), ion source (middle column) and several types of mass spectrometers, (a) Agilent 7500 from Agilent, (b) Platform from CV Instruments, or (c) Element from Thermo Fisher Scientific. (Parts of this figure were reproduced with permission from CETAC Technologies, Agilent, CV Instruments and Thermo Tisher Scientific, respectively.)...
Recently, a very important development has been made to enhance the ease of manipulation and the range of applicability of emission spectroscopy to trace metal analysis. This development is the plasma source which can be employed as an. accessory source in most direct reading emission spectrometers in place of the arc or spark or may be incorporated directly in the design of the spectrometer by the manufacturer. This development has been discussed in detail in the recent literature ( 9,53-58). [Pg.372]

Russ, G.R III, and Bazan, J.M. (1987) Isotopic ratio measurements with an inductively coupled plasma source mass spectrometer. Spectrochim. Acta 42B, 49-62. [Pg.327]


See other pages where Plasma source spectrometers is mentioned: [Pg.259]    [Pg.665]    [Pg.259]    [Pg.665]    [Pg.353]    [Pg.369]    [Pg.317]    [Pg.628]    [Pg.631]    [Pg.40]    [Pg.55]    [Pg.665]    [Pg.472]    [Pg.526]    [Pg.620]    [Pg.293]    [Pg.129]    [Pg.161]    [Pg.110]    [Pg.113]    [Pg.118]    [Pg.149]    [Pg.20]    [Pg.28]    [Pg.29]    [Pg.65]    [Pg.95]    [Pg.120]    [Pg.131]    [Pg.158]    [Pg.358]    [Pg.381]    [Pg.293]    [Pg.311]    [Pg.61]    [Pg.67]    [Pg.32]   
See also in sourсe #XX -- [ Pg.259 ]




SEARCH



Plasma source ion trap mass spectrometer

Plasma sources

© 2024 chempedia.info