Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Photometric units

A summary of symbols and units that are used in radiometry is shown in Table B.l. A summary of symbols and units that are used in photometry is shown in Table B.2. The photometric quantities use the same symbol except that the subscript v is added. A good introduction to the theory of radiometry is given by Poynton (2003), Jahne (2002), Horn (1986), and Levi (1993). See also International Commission on Illumination (1983). The radiometric units refer to physical quantities whereas the photometric units refer to perceptual quantities. In photometry, the physical quantities are weighted by the spectral sensitivity of the human visual system. Let Q(X) be the energy at wavelength X. Then, the corresponding photometric quantity is given by... [Pg.331]

Photometry The measurement of quantities associated with light, i.e., based on the average apparent intensity of a light source as viewed by a normal light-adapted human eye. Photometric units report light intensity in terms of the illuminance, e.g., candlepower (lumen/ft2) or lux (lumen/m2). Photometric units are only appropriate for visible radiation. [Pg.374]

Fig. 3). For this reason, always radiometric units must be used in photosynthesis instead of photometric units. The radiometric analogue of the lumen (luminous flux) is the watt (radiant flux). At each wavelength both are coupled according to the standard observer curve at 555 nm 1 W of radiant flux corresponds to a luminous flux of 683 lumens by definition. As a comparison, at 650 nm, 1 W corresponds to 73 lumens only. [Pg.196]

The optical, electrochemical, and electrical properties discussed up to now are properties of distinct materials, but in a multicomponent system the overall efficiency depends on many factors. This means that only the appropriate combination of materials in a device determines whether it is efficient or not. But for the application the device efficiency is the most important parameter. For OLEDs, several efficiency values are distinguished. Since they are related to display applications, the photometric unit system weighted by the photosensitivity of the human eye is used. The luminance measures the brightness of a radiating area in cd/m the luminous efficiency is given in cd/A and the power efficiency in Im/W [64]. [Pg.99]

The completion of the wash-out operations should be checked as carefully as the chemical transformations on polymer, since remaining by-products would cause side reactions in subsequent steps of the synthesis. In our group, this is simply and sensitively performed by the continuously monitoring control system, which operates during the whole Merrifield synthesis. On the recordings, completed wash-out operations after each chemical reaction on polymer are indicated by a straight zero curve, since reference and sample flow cell of the photometric unit in this situation contain nothing but pure solvent. [Pg.62]

Intensity of UV radiation is measured in radiometric units (Tables 10.1,10.2), whereas the intensity of visible light is measured in photometric units (Table 10.1) by radiometers [1762] ... [Pg.441]

Radiometers consists of filters, a sensor (silicon radiation detectors, which are sensitive in the range 200-1100 nm, or vacuum photodiodes with specific photocathode materials) and an input optical attachment such a cosine correction diffuser or fibre optic assembly [1762, 1770]. The current of the sensor is converted to a voltage by the converter module . This output voltage (0-5 V) can monitor the relative intensity of emitted radiation of 0-100%. This value can simply be displayed using an analogue or digital instrument, calibrated in any radiometric and/or photometric units. [Pg.441]

Radiometric units (subscript e) Photometric units (subscript v) ... [Pg.442]

The network is implemented by the participating countries through WHO. Sulfur Dioxide (SO,) and suspended particulate matter (SPN) data are routinely reported from each of the participating countries to WHO for entry into the global data base, which is maintained by the United States Environmental Protection Agency (U.S. EPA) at Research Triangle Park, North Carolina. Methods of collection and analysis varies by country. SPN is reported either as a gravimetric or a photometric (transmittance or reflectance) measurement converted to mass units. Since the accuracy of the conversion of the photometric measurement to mass units is... [Pg.163]

In many instruments the meter read-out is calibrated in absorbance units using a logarithmic scale while other instruments retain the convenience of a linear scale but convert the signal from the detector to a logarithmic one by electronic or mechanical means. It is essential when using a photometric instrument to know if it is calibrated in absorbance or transmittance units. [Pg.50]

The amount transformation process is illustrated with data for chlorpyrifos in the flame photometric detector, phosphorus mode, and shown in Table VI. Level 1 transformations were calculated where the amount power was increased by 0.03 units for each step. At an amount power of 0.20 the F statistic of 32.7 showed a minimum but at a confidence level of 95% did not satisfy the F test for linearity. Power steps changed by only 0.01 and 0.001 units in the vicinity of the minimum were then calculated as shown in levels 2 and 3. The best linearity was found in this case at a power transformation of 0.182 although the F statistic of 8.33 did not indicate linearity when compared with the critical F of 2.99 at P=.95. Calculations at these second and third levels were not always necessary and even when performed did not always lead to a satisfactory condition of linearity. [Pg.148]

The paH response is tested by means of the paH values as determined by the indicator method the electromodve force of the cell immersed in buffer solutions whose pan is known is measured and the pan is spectro-photometrically determined and then plotted against E (Fig. 11). It can be seen that for this ethylene glycol-glass electrode the practical response is in good agreement with the theoretical one between pan 2 and 9 and for -1-21, -Hi, and — 19°C. The reproducibility of the determinations, estimated by the use of two different assemblies of electrodes, is better than 1.0 mV and the uncertainty of the pon determination is estimated at 0.1 pan unit. [Pg.298]


See other pages where Photometric units is mentioned: [Pg.332]    [Pg.200]    [Pg.187]    [Pg.169]    [Pg.170]    [Pg.125]    [Pg.143]    [Pg.141]    [Pg.2714]    [Pg.320]    [Pg.371]    [Pg.1638]    [Pg.216]    [Pg.332]    [Pg.200]    [Pg.187]    [Pg.169]    [Pg.170]    [Pg.125]    [Pg.143]    [Pg.141]    [Pg.2714]    [Pg.320]    [Pg.371]    [Pg.1638]    [Pg.216]    [Pg.76]    [Pg.79]    [Pg.387]    [Pg.238]    [Pg.99]    [Pg.50]    [Pg.762]    [Pg.825]    [Pg.297]    [Pg.342]    [Pg.608]    [Pg.175]    [Pg.160]    [Pg.13]    [Pg.178]    [Pg.625]    [Pg.145]    [Pg.273]    [Pg.124]    [Pg.135]    [Pg.82]    [Pg.213]    [Pg.224]    [Pg.225]   
See also in sourсe #XX -- [ Pg.331 ]




SEARCH



Photometric

© 2024 chempedia.info