Particle-Beam Techniques

Further laser-rf studies using the described scheme have been performed for a considerable number of species, namely, for Sm isotopes, for [Pg.19]

) iDy Dy and Tr, numerous aims of these investigations were hyperfine structure interaction constants, nuclear moments, effects of configuration interaction and core polarization in the electron shell of atoms, molecular data like effective spin rotation constants, etc. The combination of experimental data and advanced theoretical procedures, like relativistic self-consistent field calculations, provided many fruitful results regarding the reliability of electronic wave functions of atoms. [Pg.19]

Rosner and collaborators applied the laser-microwave scheme discussed above not only to atoms but for the first time also to ions, particularly to hyperfine structure measurements of Xe. In this work the ion beam passed a coaxial waveguide where the rf transitions were induced. Since the ions moved in opposite direction but parallel to the rf wave propagation, the resonance frequencies had to be corrected for the Doppler shift. [Pg.19]

In our laboratory a series of laser-microwave studies has been performed on metastable and short-lived excited states of heliumlike Li. Singly ionized lithium, like all members of the two-electron He isoelectronic sequence, belongs to the fundamental systems in atomic physics. Many of its spectroscopic and quantum-mechanical characteristics have been calcu- [Pg.20]

Figaro 12. Energy level scheme of the Isls Si and Islp P terms of The diagram includes an example of a laser-microwave pumping cycle. [Pg.21]


This technique is complementary to the thermospray technique. Relative advances of the particles beam technique over thermospray include library searchable electron impact spectra, improved reproducibility, easier use and increased predictability over a broad range of compounds. But since a particle beam requires same sample volatility, very large and polar compounds such as proteins may not provide satisfactory results using particle beam liquid chromatography-mass spectrometry. Additionally, certain classes of compounds such as preformed ions, azo dyes and complex sugars may not yield satisfactory electron impact spectra, but can be run on thermospray. In other words, both liquid chromatography-mass spectrometry techniques complement each other s limitations and the analyst may want to add both to address a broader range of samples. [Pg.56]

This method is still in use but is not described in this book because it has been superseded by more recent developments, such as particle beam and electrospray. These newer techniques have no moving parts, are quite robust, and can handle a wide variety of compound types. Chapters 8 through 13 describe these newer ionization techniques, including electrospray, atmospheric pressure ionization, plasmaspray, thermospray, dynamic fast-atom bombardment (FAB), and particle beam. [Pg.263]

The development of methods of analysis of tria2ines and thek hydroxy metabohtes in humic soil samples with combined chromatographic and ms techniques has been described (78). A two-way approach was used for separating interfering humic substances and for performing stmctural elucidation of the herbicide traces. Humic samples were extracted by supercritical fluid extraction and analy2ed by both hplc/particle beam ms and a new ms/ms method. The new ms /ms unit was of the tandem sector field-time-of-flight/ms type. [Pg.246]

The particle-beam interface has been developed primarily to provide El spectra from HPLC eluates but may be combined with other ionization techniques such as CL If quantitative studies are being undertaken, a detailed study of experimental conditions should be undertaken. Isotope-dilution methodology is advocated for the most accurate results. [Pg.151]

Various forms of off- and on-line AES/AAS can achieve element specific detection in IC. The majority of atomic emission techniques for detection in IC are based on ICP. In the field of speciation analysis both IC-ICP-AES and IC-ICP-MS play an important role. Besides the availability of the ICP ion source for elemental MS analysis, structural information can be provided by interfaces and ion sources like particle beam or electrospray. [Pg.272]

Flow limitations restrict application of the DFI interface for pSFC-MS coupling. pSFC-DFI-MS with electron-capture negative ionisation (ECNI) has been reported [421], The flow-rate of eluent associated with pSFC (either analytical scale - 4.6 mm i.d. - or microbore scale 1-2 mm, i.d.) renders this technique more compatible with other LC-MS interfaces, notably TSP and PB. There are few reports on workable pSFC-TSP-MS couplings that have solved real analytical problems. Two interfaces have been used for pSFC-EI-MS the moving-belt (MB) [422] and particle-beam (PB) interfaces [408]. pSFC-MB-MS suffers from mechanical complexity of the interface decomposition of thermally labile analytes problems with quantitative transfer of nonvolatile analytes and poor sensitivity (low ng range). The PB interface is mechanically simpler but requires complex optimisation and poor mass transfer to the ion source results in a limited sensitivity. Table 7.39 lists the main characteristics of pSFC-PB-MS. Jedrzejewski... [Pg.482]

Electrospray has been successful for numerous azo dyes that are not ionic salts. Several anthraquinone dyes have been analysed by LC-ESI-MS [552]. Electrospray achieves the best sensitivity for compounds that are precharged in solution (e.g. ionic species or compounds that can be (de)protonated by pH adjustment). Consequently, LC-ESI-MS has focused on ionic dyes such as sulfonated azo dyes which have eluded analysis by particle-beam or thermospray LC-MS [594,617,618]. Techniques like LC-PB-MS and GC-MS, based on gas-phase ionisation, are not suitable for nonvolatile components such as sulfonated azo dyes. LC-TSP-MS on... [Pg.514]

Table 7.83 lists the main characteristics of TLC-FAB-MS/LSIMS. A key difference between EI/CI and FAB/LSIMS/LD is the fact that sampling in FAB and LSIMS is from a specified location that corresponds to the impact footprint of the primary particle beam. The natural compatibility of FAB, LSIMS and LD with the direct mass-spectrometric analysis of TLC plates is readily apparent. Most mass-spectrometric measurements are destructive in nature, but FAB and LSIMS are surface-sensitive techniques in which the material actually consumed in the analysis is sputtered only from the top few microns of the sample spot. The underlying bulk is not affected, and can be used for further probing. The major limitation of TLC-FAB depends on the capability of the compounds to produce a good spectrum. [Pg.540]

With charged-particle microprobes, the samples must be stained and thinned to improve both contrast and signal-to-noise ratio coated with a thin conducting layer to reduce charging effects and improve spatial resolution and be in vacuum to maintain the charged-particle beams. Finally, information on the chemical state of the detected elements is difficult to obtain using techniques based on charged particles. [Pg.640]

If thick samples are placed in the specimen chamber for analysis, the particles are slowed down and eventually stopped in the sample, so the calculation of the X-ray yield and their absorption is more complicated. Some objects may be too large to be placed in the specimen chamber, in which case the external beam technique is employed. The particle beam passes through a window at the end of the beam-line into the air where an object of any size (e.g. an archaeological artefact) may be analyzed. [Pg.101]

In recent decades the hyphenated technique, HPLC-MS, also become the method of preference in HPLC practice. Various techniques have been developed and applied for MS detection such as thermo-spray interface, atmospheric ionization (API), electrospray or ionspray ionization (ESI or ISI), and particle beam ionization. [Pg.43]


See other pages where Particle-Beam Techniques is mentioned: [Pg.596]    [Pg.720]    [Pg.114]    [Pg.121]    [Pg.121]    [Pg.123]    [Pg.321]    [Pg.17]    [Pg.258]    [Pg.276]    [Pg.8]    [Pg.596]    [Pg.720]    [Pg.114]    [Pg.121]    [Pg.121]    [Pg.123]    [Pg.321]    [Pg.17]    [Pg.258]    [Pg.276]    [Pg.8]    [Pg.800]    [Pg.1264]    [Pg.77]    [Pg.80]    [Pg.399]    [Pg.403]    [Pg.403]    [Pg.336]    [Pg.224]    [Pg.224]    [Pg.351]    [Pg.368]    [Pg.384]    [Pg.506]    [Pg.515]    [Pg.540]    [Pg.26]    [Pg.210]    [Pg.402]    [Pg.375]    [Pg.55]    [Pg.153]    [Pg.169]    [Pg.155]   