Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Open tubular columns plate heights

Mass transfer resistance in the mobile phase is more difficult to calculate because it requires an exact knowledge of the flow profile of the mobile phase. This is only known exactly for open tubular columns for which the contribution of mass transfer resistance in the mobile phase to the total plate height can be described by equation (1.27)... [Pg.14]

The possibility of obtaining significant improvements in performance by using semi-packed and open tubular columns is clearly illustrated by the values for the separation impedance in Table 1.17. Variation of the reduced plate height with the reduced velocity for an open tubular column is given by equation (1.82), assuming that the resistance to mass transfer in the stationary phase can be neglected... [Pg.44]

The efficiency of an open tubular column can be measured in several ways the most widely used methods are the number of theoretical plates (n), the nu3aber of effective theoretical plates (N), the height equivalent to a theoretical plate (HETP) or effective plate, the coating efficiency and the separation number (SM). No single method is ideal, standardization is lacking and... [Pg.85]

The general approach for kinetic optiaization of open i tubular columns has been to adopt the familiar Golay equation T (equation 1.34) and to assuae that the aobile phase can be approximated by an incompressible fluid with ideal gas properties, (44-50). Circumstances that are approximate at best but serve adequately to demonstrate some of the fundamental characteristics of open tubular columns operated at low fluid densities. The column plate height equation can be written in the form given in M equation (6.1)... [Pg.310]

RELATIVE CONTRIBUTION (PERCENTAGE BASIS) OF MASS TRANSFER RESISTANCE IN THE MOBILE AMD STATIONARY PHASE TO COLUMN PLATE HEIGHT FOR A SERIES OF 0.32 mm I.D. OPEN TUBULAR COLUMNS USING UMDECAME AT 130 C AS THE TEST SOLUTE... [Pg.535]

Upon substitution of the reduced parameters given above the separation time for a packed column and an open tubular column would be Identical if d 1.73 dp given the current limitations of open tubular column technology the column diameter cannot be reduced to the point %diere these columns can compete with packed columns for fast separations. This is illustrated by the practical txanple in Figure 6.3 (57). Ihe separation speed cannot be Increased for an open tubular column by increasing the reduced velocity since the reduced plate height is increased... [Pg.823]

In packed columns, all three terms contribute to band broadening. For open tubular columns, the multiple path term, A, is 0, so bandwidth decreases and resolution increases. In capillary electrophoresis (Chapter 26), both A and C go to 0, thereby reducing plate height to submicron values and providing extraordinary separation powers. [Pg.518]

Particles in a packed column resist flow of the mobile phase, so the linear flow rate cannot be very fast. For the same length of column and applied pressure, the linear flow rate in an open tubular column is much higher than that of a packed column. Therefore, the open tubular column can be made 100 times longer than the packed column, to give a similar pressure drop and linear flow rate. If plate height is the same, the longer column provides 100 times more theoretical plates, yielding VTOO = 10 times more resolution. [Pg.520]

Plate height is reduced in an open tubular column because band spreading by multiple flow paths (Figure 23-19) cannot occur. In the van Deemter curve for the packed column in Figure 23-15. the A term accounts for half of the plate height at the most efficient flow rate (minimum H) near 30 mL/min. If A were deleted, the number of plates on the column would be doubled. To obtain high performance from an open tubular column, the radius of the column must be small and the stationary phase must be as thin as possible to ensure rapid exchange of solute between mobile and stationary phases. [Pg.520]

Capillary electrophoresis provides unprecedented resolution. When we conduct chromatography in a packed column, peaks are broadened by three mechanisms in the van Deemter equation (23-33) multiple flow paths, longitudinal diffusion, and finite rate of mass transfer. An open tubular column eliminates multiple paths and thereby reduces plate height and improves resolution. Capillary electrophoresis reduces plate height further by knocking out the mass transfer term that comes from the finite time needed for solute to equilibrate... [Pg.604]

Reduced plate height is h = H/dp with packed columns h = H/dt. with open tubular columns Reduced length of column is Z = L/dp or L/dc. [Pg.6]

Carrier gas flow should be optimised for a particular column and a particular carrier gas. This is most important for open tubular columns. Fig. 5 shows the relationship between efficiency expressed as the height equivalent of a theoretical plate versus carrier gas velocity (Van Deemter plot) for a 28 m by 0.25 mm internal diameter wall-coated open tubular column of Carbowax 20M. [Pg.191]

Relative contribution (%) of resistance to mass transfer in the mobile and stationary phases to the column plate height for undecane at 130°C for a 0.32 mm internal diameter open tubular columns in gas chromatography... [Pg.35]


See other pages where Open tubular columns plate heights is mentioned: [Pg.615]    [Pg.394]    [Pg.16]    [Pg.23]    [Pg.531]    [Pg.534]    [Pg.542]    [Pg.563]    [Pg.818]    [Pg.822]    [Pg.522]    [Pg.523]    [Pg.538]    [Pg.167]    [Pg.179]    [Pg.324]    [Pg.325]    [Pg.137]    [Pg.30]    [Pg.31]    [Pg.34]    [Pg.35]    [Pg.41]    [Pg.81]    [Pg.85]    [Pg.128]    [Pg.585]    [Pg.688]    [Pg.690]    [Pg.278]    [Pg.259]    [Pg.89]    [Pg.463]   
See also in sourсe #XX -- [ Pg.688 ]




SEARCH



Column plate

Column plate height

Height column

Open columns

Open tubular

Open tubular columns

Plate height

Tubular plates

© 2024 chempedia.info