Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Mass spectrometrist

The two major difficulties facing the analyst/mass spectrometrist concern firstly how to get the whole of the sample into the plasma flame efficiently and secondly how to do so without destabilizing or extinguishing the flame. Although plasma flames operate at temperatures of 6000 to 8000 K, the mass of gas in the flame is very small, and its thermal capacity is correspondingly small (Figure 15.1). [Pg.97]

Two further expressions are used in discussions on isotope ratios. These are the atom% and the atom% excess, which are defined in Figure 48.6 and are related to abundance ratios R. It has been recommended that these definitions and some similar ones should be used routinely so as to conform with the system of international units (SI). While these proposals will almost certainly be accepted by mass spectrometrists, their adoption will still leave important data in the present format. Therefore, in this chapter, the current widely used methods for comparison of isotope ratios are fully described. The recommended Sl-compatible units such as atom% excess are introduced where necessary. [Pg.361]

EPA. Environmental Protection Agency (U.S. agency responsible for many methods of analysis used by mass spectrometrists)... [Pg.445]

Detailed discussions of some of the remaining peaks in Figures 7 and 8 and in the mass spectra of 10a and the D20-exchanged analogs is of more interest to the mass spectrometrist than to the carbohydrate chemist. The probable origins of these peaks will be discussed here, however, because there will be occasions when the carbohydrate chemist must dig into a spectrum in order to satisfy himself that he has interpreted the spectrum in terms of a correct structure. [Pg.230]

Sensitivities achieved in f.a.b. analysis are both operator- and sample-dependent. Experienced mass spectrometrists working with well purified samples use between 0.1 and 5 jig of sample when analyzing derivatives, and between 1 and 10 (ig when analyzing native compounds. The higher the molecular weight, the greater the quantity of sample needed. [Pg.33]

The traditional mass spectrometry laboratory is equipped with top-of-the-line instrumentation, run by a mass spectrometrist for high-performance, nonroutine analyses. The use of mass spectrometry has recently... [Pg.349]

Various analytical methods have made quantum leaps in the last decade, not least on account of superior computing facilities which have revolutionised both data acquisition and data evaluation. Major developments have centred around mass spectrometry (as an ensemble of techniques), which now has become a staple tool in polymer/additive analysis, as illustrated in Chapters 6 and 7 and Section 8.5. The impact of mass spectrometry on polymer/additive analysis in 1990 was quite insignificant [100], but meanwhile this situation has changed completely. Initially, mass spectrometrists have driven the application of MS to polymer/additive analysis. With the recent, user-friendly mass spectrometers, additive specialists may do the job and run LC-PB-MS or LC-API-MS. The constant drive in industry to increase speed will undoubtedly continuously stimulate industrial analytical scientists to improve their mass-spectrometric methods. [Pg.734]

Mass spectrometry combines exquisite sensitivity with a precision that often depends more on the uncertainties of sampling and sample preparation than on those of the method itself. Mass spectrometry is a supreme identification and recognition method in polymer/additive analysis through highly accurate masses and fragmentation patterns quantitation is its weakness. Direct mass spectrometry of complex polymeric matrices is feasible, yet not often pursued. Solid probe ToF-MS (DI-HRMS) is a breakthrough. Where used routinely, mass spectrometrists are usually still in charge. At the same time, however, costs need to be watched. [Pg.734]

The extension of analytical mass spectrometry from electron ionization (El) to chemical ionization (Cl) and then to the ion desorption (probably more correctly ion desolvation ) techniques terminating with ES, represents not only an increase of analytical capabilities, but also a broadening of the chemical horizon for the analytical mass spectrometrist. While Cl introduced the necessity for understanding ion—molecule reactions, such as proton transfer and acidities and basicities, the desolvation techniques bring the mass spectrometrist in touch with ions in solution, ion-ligand complexes, and intermediate states of ion solvation in the gas phase. Gas-phase ion chemistry can play a key role in this new interdisciplinary integration. [Pg.315]

Mass Spectral Library[35] and the Wiley Registry of Mass Spectral Data[36] are now available to help mass spectrometrists in the identification of unknowns. [Pg.67]

Technique selection Since you are a mass spectrometrist, you decide a liquid-solid extraction of the solid followed by MS analysis can identify impurities in the carbon. [Pg.844]

Commentary To solve this problem, the mass spectrometrist needed to work outside of his graduate school specialty (his comfort zone ). This required him to call upon other resources, like the polymer chemist. The vital importance of communication (between the analyst, the technical director, the PI and DA plant engineers, and the polymer chemist) was essential to defining the problem, and ultimately solving it. [Pg.846]

To apply this equation to calculate the difference between the standard enthalpies of formation of AB and AB+, we need to know the values of Af//°(e ,g), A, and A. The heat capacities of AB and AB+ are easily evaluated lfom statistical mechanics calculations, provided that their structures and vibrational frequencies are available. Usually, A 0. However, with regard to AfH°(e, g) and A, the apparently simple task of assigning their values has been a source of controversy involving two scientific communities the calorimetrists and the mass spectrometrists. [Pg.48]

The reasoning we have just used to derive equation 4.6 was taken from the introductory chapter of a popular data compilation [62], However, as pointed out by the authors of this database, the true historical origin of the ion convention is less complicated Although the mass spectrometrists were willing to accept that AfH°(e, g) = 0 at all temperatures, they justly (see following discussion) felt uneasy about the value assigned to X. To avoid this uncertainty, they have postulated A = 0 and obtained equation 4.6. [Pg.48]

From the 1950s to the present mass spectrometry has changed tremendously and still is changing. [12,13] The pioneering mass spectrometrist had a home-built rather than a commercial instmment. This machine, typically a magnetic sector instrument with electron ionization, delivered a few mass spectra per day, provided sufficient care was taken of this delicate device. If the mass spectrometrist knew this particular instrument and understood how to interpret El spectra he or she had a substantial knowledge of mass spectrometry of that time. [14-18]... [Pg.1]

Well, mass spectrometry is somewhat different. The problems usually start with the simple fact that most mass spectrometrists do not like to be called mass spec-troscopists. [Pg.2]

Rule First of all, never make the mistake of calling it mass spectroscopy. Spectroscopy involves the absorption of electromagnetic radiation, and mass spectrometry is different, as we will see. The mass spectrometrists sometimes get upset if you confuse this issue. [21]... [Pg.2]

Note Some mass spectrometrists use the unit thomson [Th] (to honor J. J. Thomson) instead of the dimensionless quantity m/z. Although the thomson is accepted by some journals, it is not a SI unit. [Pg.5]

Note In particular mass spectrometrists in the biomedical field of mass spectrometry tend to use the dalton [Da] (to honor J. Dalton) instead of the unified atomic mass [u]. The dalton also is not a SI unit. [Pg.5]

As with acylium ions and carbenium ions before, the series of homologous im-monium ions is part of the mass spectrometrist s tool box. They can easily be recognized in the mass spectra and have even-numbered m/z values (Tab. 6.6). In the El spectrum of iV-ethyl-iV-methyl-propanamine the series is completely present from m/z 30 up to m/z 100. [Pg.238]

Perhaps the first suggestion of INCs came from Rylander and Meyerson. [172,173] The concept that the decomposition of oxonium and immonium ions involve INCs (Chap. 6.11.2) was successfully put forth by Bowen and Williams, [143,157,158,167,174] and the analogies to solvolysis were described by Morton. [168] Nonetheless, mass spectrometrist were too much used to strictly unimolecular reactions to assimilate such a concept without stringent proof. [Pg.300]

Random rearrangements of hydrocarbons were noted by the early mass spectrometrists in the petroleum industry. For example,... [Pg.15]

It is important to clarify a confusing issue that is commonly encountered in mass spectrometry and clinical chemistry. Mass spectrometrists define selectivity and sensitivity in quite different terms than do clinical chemists. Analytical selectivity and sensitivity are terms that should help clarify the situation. The measure of sensitiv-... [Pg.800]

In general, all mass spectrometers share at least three distinct structures the source, the analyzer, and the detector. Differences in these three structures identify the multitude of MS systems. The source is perhaps the most crucial element of the mass spectrometer. Therefore, the selection of the source primarily differentiates the various MS systems. Although specific analyzers or detectors may be preferable for a particular MS application, mass spectrometrists will often refer to different systems solely by the source. Only occasionally, when a time-of-flight mass analyzer is used, for example, analysts refer to the method by the type of the mass analyzer. [Pg.706]

Mass Unit The unified atomic mass unit, or u, is the fundamental unit of mass for most mass spectrometrists. The Dalton, or Da, is also generally accepted and is commonly used in descriptions of large, biological molecules. The mass unit is defined as one-twelfth of the mass of carbon-12. Atomic mass unit, or amu, is technically incorrect but still commonly used. The unit Thomson (Th) has been used as a unit of m jz. However, Th is not accepted by most mass spectrometry journals and the International Union of Pure and Applied Chemistry (IUPAC). Therefore, m/z used for labeling the x-axis of mass spectra is unit less. [Pg.13]

Among the questions of importance to cosmology are the elemental composition of stars and other galactic matter and the isotopic compositions of those elements. Investigations of this type have covered several decades and represent a nice collaboration between theoretical astrophysicists and mass spectrometrists [80]. Thermal ionization has played a role in analysis, both isotopic and, through isotope dilution, of-concentration, of many of the elements and helped resolve some of the anomalies that were present in the, results of early work. Isotope dilution is inherently a precise method of quantification and was able to reduce... [Pg.22]


See other pages where Mass spectrometrist is mentioned: [Pg.20]    [Pg.349]    [Pg.350]    [Pg.350]    [Pg.350]    [Pg.350]    [Pg.351]    [Pg.384]    [Pg.27]    [Pg.279]    [Pg.370]    [Pg.117]    [Pg.8]    [Pg.331]    [Pg.398]    [Pg.524]    [Pg.524]    [Pg.166]    [Pg.110]    [Pg.476]    [Pg.193]    [Pg.49]   
See also in sourсe #XX -- [ Pg.6 ]




SEARCH



© 2024 chempedia.info