Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Ionization, frequency

Photoionization ti me-of-fli ght mass spectrometry is almost exclusively the method used in chemical reaction studies. The mass spectrometers, detectors and electronics are almost identical. A major distinction is the choice of ionizing frequency and intensity. For many stable molecules multi photon ionization allowed for almost unit detection efficiency with controllable fragmentation(20). For cluster systems this has been more difficult because high laser intensities generally cause extensive dissociation of neutrals and ions(21). This has forced the use of single photon ionization. This works very well for low i oni zati on potential metals ( < 7.87 eV) if the intensity is kept fairly low. In fact for most systems the ionizing laser must be attenuated. A few very small... [Pg.52]

Though the original work is difficult to understand very good reviews about the van der Waals interaction between macroscopic bodies have appeared [114,120], In the macroscopic treatment the molecular polarizability and the ionization frequency are replaced by the static and frequency dependent dielectric permittivity. The Hamaker constant turns out to be the sum over many frequencies. The sum can be converted into an integral. For a material 1 interacting with material 2 across a medium 3, the non-retarded5 Hamaker constant is... [Pg.88]

Here, n is the refractive index and is the ise the mean ionization frequency of the material. Typically this is ve k, 3x1015 Hz. If the absorption frequencies of all three materials are assumed to be the same, we obtain the following approximation for the non-retarded Hamaker constant ... [Pg.89]

I = ionization frequency (formation of an electron and a positive ion). Further, electrical neutrality is assumed ... [Pg.537]

The polarizabilities and the medium permittivity are, however, real quantities. Its exact form depends on the molecular details of the three substances. For the case of isolated (gas phase) molecules, e iiv) = 1, and if there is only one electronic absorption frequency, the ionization frequency Vj, the excess polarizabilities are equal to the total ones, which are shown to be... [Pg.81]

IP the spectrum in Fig.5(a) was measured. Upon reducing the ionization frequency so that hvj +hv2 lies within the "tail region but below the "step" in Fig.4, the spectrum shown in Fig.5(b) was obtained. This spectrum corresponds to the subpopulation of the n=5 clusters which exhibit the dashed PIE curve in Fig.4. Since the spectrum in Fig.5(a) is a composite of both subpopulations, subtraction of Fig.5(b) yields the spectrum of the subpopulation corresponding to the sharp step region in Fig.4 [bottom curve (a-b) in Fig.5]. Thus, Fig.5(a) decomposes cleanly into a spectrum with narrow bands (Av 5 cm" ) with virtually no residual broad background after subtraction, and a broad (Av 70 cm ) spectrum with... [Pg.391]

Fig.5.SI-R2PI spectra of Car Arg clusters at different ionization frequencies(see Fig. 4). (a) above step (b) below step in tail region of ionization threshold ... Fig.5.SI-R2PI spectra of Car Arg clusters at different ionization frequencies(see Fig. 4). (a) above step (b) below step in tail region of ionization threshold ...
X-rays may be detected either photographically or with an ionization counter. They have great penetrating power which increases with their frequency, and owing to this are used to photograph the interior of many solid objects, notably the human body and in monitoring for faults in construction. [Pg.429]

A-B relative or external motion undergo free-free transitions (E., E. + dE.) (Ej Ej+ dE within the translational continuum, while the structured particles undergo bound-bound (excitation, de-excitation, excitation transfer) or bound-free (ionization, dissociation) transitions = (a, 3) ->/= (a, (3 ) in their internal electronic, vibrational or rotational structure. The transition frequency (s ) for this collision is... [Pg.2011]

In contrast to the ionization of C q after vibrational excitation, typical multiphoton ionization proceeds via the excitation of higher electronic levels. In principle, multiphoton ionization can either be used to generate ions and to study their reactions, or as a sensitive detection technique for atoms, molecules, and radicals in reaction kinetics. The second application is more common. In most cases of excitation with visible or UV laser radiation, a few photons are enough to reach or exceed the ionization limit. A particularly important teclmique is resonantly enlianced multiphoton ionization (REMPI), which exploits the resonance of monocluomatic laser radiation with one or several intennediate levels (in one-photon or in multiphoton processes). The mechanisms are distinguished according to the number of photons leading to the resonant intennediate levels and to tire final level, as illustrated in figure B2.5.16. Several lasers of different frequencies may be combined. [Pg.2135]

Electrons from a spark are accelerated backward and forward rapidly in the oscillating electromagnetic field and collide with neutral atoms. At atmospheric pressure, the high collision frequency of electrons with atoms induces chaotic electron motion. The electrons gain rapidly in kinetic energy until they have sufficient energy to cause ionization of some gas atoms. [Pg.395]

Photoelectron spectroscopy involves the ejection of electrons from atoms or molecules following bombardment by monochromatic photons. The ejected electrons are called photoelectrons and were mentioned, in the context of the photoelectric effect, in Section 1.2. The effect was observed originally on surfaces of easily ionizable metals, such as the alkali metals. Bombardment of the surface with photons of tunable frequency does not produce any photoelectrons until the threshold frequency is reached (see Figure 1.2). At this frequency, v, the photon energy is just sufficient to overcome the work function

[Pg.289]

Ionization energies can be computed to about 0.2 eV rotational barriers to about 0.5 kcal/mol dipole moments to about 0.5 D barriers to inversion to about 2.5 kcal/mol infrared frequencies can be computed with about a 15% error (usuaHy too high) and protonation energies are accurate to about 1 piCunit. [Pg.162]

The electron-ion coUision frequency can be approximated by the Spit2er-Harm (39) formulation for a completely ionized gas ... [Pg.420]

Multiphoton Absorption and Ionization. High laser powers can induce the simultaneous absorption of two or more photons that together provide the energy necessary to excite a transition this transition may be one that is forbidden as a single-photon process (8,297). Such absorption can be made Doppler-free by propagating two laser beams of frequency V in opposite directions, so the Doppler shifts cancel and a two-photon transition occurs at 2v for any absorber velocity. The signal is strong because aU absorbers contribute, and peak ampHtudes are enhanced by, which may... [Pg.321]

Charge-Transfer Compounds. Similat to iodine and chlorine, bromine can form charge-transfer complexes with organic molecules that can serve as Lewis bases. The frequency of the iatense uv charge-transfer adsorption band is dependent on the ionization potential of the donor solvent molecule. Electronic charge can be transferred from a TT-electron system as ia the case of aromatic compounds or from lone-pairs of electrons as ia ethers and amines. [Pg.284]

Spark Source Mass Spectrometry (SSMS) is a method of trace level analysis—less than 1 part per million atomic (ppma)—in which a solid material, in the form of two conducting electrodes, is vaporized and ionized by a high-voltage radio frequency spark in vacuum. The ions produced from the sample electrodes are accelerated into a mass spectrometer, separated according to their mass-to-charge ratio, and collected for qualitative identification and quantitative analysis. [Pg.45]

The large variability in elemental ion yields which is typical of the single-laser LIMS technique, has motivated the development of alternative techniques, that are collectively labeled post-ablation ionization (PAI) techniques. These variants of LIMS are characterized by the use of a second laser to ionize the neutral species removed (ablated) from the sample surface by the primary (ablating) laser. One PAI technique uses a high-power, frequency-quadrupled Nd-YAG laser (A, = 266 nm) to produce elemental ions from the ablated neutrals, through nonresonant multiphoton ionization (NRMPI). Because of the high photon flux available, 100% ionization efflciency can be achieved for most elements, and this reduces the differences in elemental ion yields that are typical of single-laser LIMS. A typical analytical application is discussed below. [Pg.588]

SNMS sensitivity depends on the efficiency of the ionization process. SNs are post-ionized (to SN" ) either hy electron impact (El) with electrons from a hroad electron (e-)heam or a high-frequency (HF-) plasma (i.e. an e-gas), or, most efficiently, hy photons from a laser. In particular, the photoionization process enables adjustment of the fragmentation rate of sputtered molecules by varying the laser intensity, pulse width, and/or wavelength. [Pg.123]


See other pages where Ionization, frequency is mentioned: [Pg.236]    [Pg.5]    [Pg.160]    [Pg.202]    [Pg.130]    [Pg.79]    [Pg.267]    [Pg.157]    [Pg.651]    [Pg.201]    [Pg.387]    [Pg.390]    [Pg.236]    [Pg.5]    [Pg.160]    [Pg.202]    [Pg.130]    [Pg.79]    [Pg.267]    [Pg.157]    [Pg.651]    [Pg.201]    [Pg.387]    [Pg.390]    [Pg.1199]    [Pg.2060]    [Pg.29]    [Pg.89]    [Pg.363]    [Pg.135]    [Pg.337]    [Pg.347]    [Pg.307]    [Pg.204]    [Pg.3]    [Pg.17]    [Pg.116]    [Pg.184]    [Pg.185]    [Pg.197]   
See also in sourсe #XX -- [ Pg.79 ]




SEARCH



© 2024 chempedia.info