Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Gas sample injection

The simplest mode of IGC is the infinite dilution mode , effected when the adsorbing species is present at very low concentration in a non-adsorbing carrier gas. Under such conditions, the adsorption may be assumed to be sub-monolayer, and if one assumes in addition that the surface is energetically homogeneous with respect to the adsorption (often an acceptable assumption for dispersion-force-only adsorbates), the isotherm will be linear (Henry s Law), i.e. the amount adsorbed will be linearly dependent on the partial saturation of the gas. The proportionality factor is the adsorption equilibrium constant, which is the ratio of the volume of gas adsorbed per unit area of solid to its relative saturation in the carrier. The quantity measured experimentally is the relative retention volume, Vn, for a gas sample injected into the column. It is the volume of carrier gas required to completely elute the sample, relative to the amount required to elute a non-adsorbing probe, i.e. [Pg.35]

TMA, its N-oxide and related aliphatic amines like methylamine and dimethylamine in urine may be quantified using head-space gas chromatography [28] or direct injection of the head-space gas into the gas sample injection port of a mass spectrometer [27]. These methods take advantage of the volatility of the amines and evaluate the amine-rich head-space gas generated above the sample by direct injection. The... [Pg.787]

For the isotope dilution, mass spectrometry method samples are injected directly into the gas sample injection port of the mass spectrometer [27]. These techniques do not allow concurrent analysis of TMA and TMA N-oxide in the sample. TMA N-oxide is quantified indirectly by measuring the increase in TMA after chemical reduction. [Pg.788]

When the samples are spiked with [2H9]-TMA instead of isopropylamine, the head-space gas can also be injected directly into the gas sample injection port of a mass spectrometer for TMA quantification [27]. Electron impact mass spectra are collected over the mass range m/z 10-500 at a scanning rate of once per 3 s. The ion intensities of 20 consecutive scans are averaged and the ratio of the ions at m/z 59 and 68 is determined for TMA [27]. [Pg.789]

For hydrocarbons, using a 5-cnu gas sample injected into the chromatograph, McAuliffe indicates... [Pg.407]

Both manual procedures and automated devices for gas sample injection have been developed manual injection using gas-tight syringes [93], sample injection using a loop in a headspace autosampler with electropneumatic systems [2], and automated injections by means of an autosampler equipped with gas-tight syringes and by means of a static-headspace autosampler equipped with a trap to preconcentrate and focus the VOCs [94-96]. [Pg.617]

The inlet gas line should have taps for gas sampling, temperature measurement, pressure measurement, and for an injection nozzle for... [Pg.180]

Introduce a 0.30 pL portion of the solvent extract into the gas chromatograph. It is found that solutions of concentrations greater than 0.3 M are unsuitable as they deposit solid and thus cause a blockage of the 1 jj.L microsyringe used for the injection of the sample. The syringe is flushed several times with the sample solution, filled with the sample to the required volume, excess liquid wiped from the tip of the needle and the sample injected into the chromatograph. [Pg.249]

There are several types of sample introduction systems available for GC analysis. These include gas sampling valves, split and splitless injectors, on-column injection systems, programmed-temperature injectors, and concentrating devices. The sample introduction device used depends on the application. [Pg.9]

Step- and impulse-response methods. Chemisorption can conveniently be measured under flow conditions using transient techniques, in particular, step-response and impulse-response measurements. After pretreatment, pulses of probe gas are injected into a carrier gas stream passing through the reactor that contains the pre-treated. sample. The response is detected at the reactor exit. [Pg.107]

Gas samples were taken by releasing a portion of the interior gas mixture into a sample chamber which uses a small balloon to maintain a low positive pressure against a septum. This chamber was purged three times with the gas mixture before removing a sample for injection into the gas partitioner by means of a Pressure-Lok syringe. [Pg.127]

The repeatability of gas sample separation is presented in Fig. 6.12. The separation peaks of the mixed gaseous analytes match those for each individual analyte quite well, irrespective of the analyte concentration, which correlates to the peak height (Fig. 6.12a, b). Although manual sample injection may introduce a certain degree of variation in retention time measurement, using a gas marker can significantly improve such measurement, as demonstrated in Fig. 6.12c, in which decane is used as a marker. [Pg.137]

Field desorption (FD) was introduced by Beckey in 1969 [76]. FD was the first soft ionization method that could generate intact ions from nonvolatile compounds, such as small peptides [77]. The principal difference between FD and FI is the sample injection. Rather than being in the gas phase as in FI, analytes in FD are placed onto the emitter and desorbed from its surface. Application of the analyte onto the emitter can be performed by just dipping the activated emitter in a solution. The emitter is then introduced into the ion source of the spectrometer. The positioning of the emitter is cmcial for a successful experiment, and so is the temperature setting. In general, FI and FD are now replaced by more efficient ionization methods, such as MALDI and ESI. For a description of FD (and FI), see Reference 78. [Pg.27]

Once into the 21st century, hyphenated instrumentation (i.e., those that couple two instruments together) became prevalent in laboratories. This is the combination of two or more, often different, instruments. In simple terms, the purpose is to first separate the analyte of interest and then to identify it. This takes place using a sample injected into the combined instruments. The most common of the hyphenated instruments is the gas chromatograph, the output of which is fed into a mass spectrometer to produce a gas chromatography-mass spectrometry (GC-MS) [35],... [Pg.32]

GC is a powerful, rapid method for separating mixtures of gases and compounds with boiling points below 400°C.2 The sample, when introduced into a gas chromatograph, must either already be a gas or be immediately turned into a gas on injection. A complete chromatogram is usually obtained in less than 20 minutes, although some analyses can take much longer. [Pg.274]


See other pages where Gas sample injection is mentioned: [Pg.270]    [Pg.646]    [Pg.270]    [Pg.646]    [Pg.577]    [Pg.165]    [Pg.306]    [Pg.239]    [Pg.10]    [Pg.49]    [Pg.139]    [Pg.830]    [Pg.420]    [Pg.421]    [Pg.131]    [Pg.416]    [Pg.417]    [Pg.627]    [Pg.643]    [Pg.645]    [Pg.648]    [Pg.202]    [Pg.438]    [Pg.245]    [Pg.46]    [Pg.49]    [Pg.219]    [Pg.465]    [Pg.77]    [Pg.94]    [Pg.136]    [Pg.315]    [Pg.142]    [Pg.391]    [Pg.460]    [Pg.64]    [Pg.418]   
See also in sourсe #XX -- [ Pg.493 ]




SEARCH



GAS INJECTION

Injecting sample

Sample injection

© 2024 chempedia.info