Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Galvanic insert

To meet the accuracy requirements that are placed on injection molds, it can be useful to perform the machining of the outer contour of the galvanized structured mold inserts before the model is pulled out of the galvanized insert. [Pg.534]

For inlet or outlet end erosion-corrosion, either extend tube ends 3 or 4 inches into the water box or install sleeves, inserts, or ferrules into the tube ends. These should be a minimum of 5 inches long. The ferrules may be nonmetallic or erosion-resistant metals, such as stainless steel, if galvanically compatible. The end of the ferrule should be feathered to prevent turbulence. [Pg.249]

Care must be exercised when installing stainless steel inserts in the inlet or exit end of copper or copper-alloy tubes, since galvanic corrosion can occur at the tuhe-insert junction. [Pg.366]

The electrolysis protection process using impressed current aluminum anodes allows uncoated and hot-dipped galvanized ferrous materials in domestic installations to be protected from corrosion. If impressed current aluminum anodes are installed in water tanks, the pipework is protected by the formation of a film without affecting the potability of the water. With domestic galvanized steel pipes, a marked retardation of the cathodic partial reaction occurs [15]. Electrolytic treatment alters the electrolytic characteristics of the water, as well as internal cathodic protection of the tank and its inserts (e.g., heating elements). The pipe protection relies on colloidal chemical processes and is applied only to new installations and not to old ones already attacked by corrosion. [Pg.456]

Figure 2.5. Gasket insertion between pipe llangcs for sealing purposes and to minimize galvanic corrosion between dissimilar piping metals. Figure 2.5. Gasket insertion between pipe llangcs for sealing purposes and to minimize galvanic corrosion between dissimilar piping metals.
The difference in electrical potential between two electrodes is the cell potential, designated E and measured in volts (V). The magnitude of E increases as the amount of charge imbalance between the two electrodes increases. For any galvanic cell, the value of E and the direction of electron flow can be determined experimentally by inserting a voltmeter in the external circuit. [Pg.1381]

In the simplest case, analyte is an electroactive species that is part of a galvanic cell. An electroactive species is one that can donate or accept electrons at an electrode. We turn the unknown solution into a half-cell by inserting an electrode, such as a Pt wire, that can transfer electrons to or from the analyte. Because this electrode responds to analyte, it is called the indicator electrode. We then connect this half-cell to a second half-cell by a salt bridge. The second half-cell has a fixed composition, so it has a constant potential. Because of its constant potential, the second half-cell is called a reference electrode. The cell voltage is the difference between the variable potential of the analyte half-cell and the constant potential of the reference electrode. [Pg.299]

Suppose you want to measure the relative amounts of Fe2+ and Fe3+ in a solution. You can make this solution part of a galvanic cell by inserting a Pt wire and connecting the cell to a constant-potential half-cell by a salt bridge, as shown in Figure 15-1. [Pg.299]

If, however, solid electrolytes remain stable when in direct contact with the reacting solid to be probed, direct in-situ determinations of /r,( ,0 are possible by spatially resolved emf measurements with miniaturized galvanic cells. Obviously, the response time of the sensor must be shorter than the characteristic time of the process to be investigated. Since the probing is confined to the contact area between sensor and sample surface, we cannot determine the component activities in the interior of a sample. This is in contrast to liquid systems where capillaries filled with a liquid electrolyte can be inserted. In order to equilibrate, the contacting sensor always perturbs the system to be measured. The perturbation capacity of a sensor and its individual response time are related to each other. However, the main limitation for the application of high-temperature solid emf sensors is their lack of chemical stability. [Pg.399]

However, electrochemical cells are most conveniently considered as two individual half reactions, whereby each is written as a reduction in the form indicated by Equations2.ll and 2.12. When this is done and values of the appropriate quantities are inserted, a potential can be calculated for each half cell of the electrode system. Then the reaction corresponding to the half cell with the more positive potential will be the positive terminal in a galvanic cell, and the electromotive force of that cell will be represented by the algebraic difference between the potential of the more-positive half cell and the potential of the less-positive half cell ... [Pg.39]

Where an intrinsically safe circuit is earthed by construction at the place of the field device (e.g. this is frequently the case using thermocouples), an Ex i-isolator with galvanic isolation shall be inserted. The reason is that intrinsically safe circuits shall be earthed at one single point only. [Pg.373]

The classic Ex i-isolator is an associated apparatus for installation in a safe area. In this case, too, Zener diodes and resistors are used for voltage and current limitation. However, the components may be rated lower, since the components inserted for galvanic isolation are able to transfer only a limited (e.g. transformers) or absolutely no electric power (e.g. optocouplers). For the maximum voltage Um, the above-mentioned is valid concerning safety barriers. [Pg.373]

The best material for a fusion apparatus (see Fig. 14a) in the laboratory is copper, which saves on gas because of its good conductivity and which is therefore economical in operation. (It is important that the melt does not come in contact with two diflFerent metals, because then a galvanic cell is formed which causes deleterious oxidation and reduction reactions.) The high fusion temperature makes it essential that the stirrer sweep over the whole surface of the fusion vessel (see sketch). The thermometer is inserted in a copper tube closed at the end with hard solder and filled with dry cylinder oil to such a depth that at least 10 cm. of the thermometer is immersed. It may be practical also to insert the thermometer in the hollow shaft of the stirrer (Fig. 14b). [Pg.58]


See other pages where Galvanic insert is mentioned: [Pg.527]    [Pg.527]    [Pg.528]    [Pg.528]    [Pg.528]    [Pg.530]    [Pg.532]    [Pg.534]    [Pg.536]    [Pg.538]    [Pg.540]    [Pg.542]    [Pg.544]    [Pg.527]    [Pg.527]    [Pg.528]    [Pg.528]    [Pg.528]    [Pg.530]    [Pg.532]    [Pg.534]    [Pg.536]    [Pg.538]    [Pg.540]    [Pg.542]    [Pg.544]    [Pg.1829]    [Pg.357]    [Pg.41]    [Pg.221]    [Pg.41]    [Pg.192]    [Pg.195]    [Pg.347]    [Pg.353]    [Pg.358]    [Pg.1588]    [Pg.1703]    [Pg.133]    [Pg.1]    [Pg.190]    [Pg.723]    [Pg.2292]    [Pg.389]    [Pg.531]   
See also in sourсe #XX -- [ Pg.502 ]




SEARCH



Galvanized Inserts and Molds

© 2024 chempedia.info