Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Flicker noise limited

Photon shot-noise or preamplifier shot-noise limitations, or source-flicker-noise limitations but utilizing source compensation" techniques... [Pg.8]

Finally, it should be noted that with our present system and certain samples, background or analyte photon signal fluxes are large enough that measurements are flicker noise limited (noise proportional to the signal). In this case, the IDA already provides detection limits equivalent to that obtained with PMT detection. [Pg.169]

Finally, zinc Is an example of an element whose absorption wavelength Is lower than 230 nm, the range where flame transmission noise dominates. Let us look a bit more closely at the case of zinc. The Instrument performance can be characterized using a precision plot, shown In Figure 2, where the relative standard deviation of concentration Is plotted on the vertical axis and concentration on the horizontal axis. The detection limit Is defined by the Intersection of the precision curve with the RSD which represents the criterion used to define detection, about 30% RSD for k - 3. Note that flame transmission flicker noise limits detection. [Pg.121]

Because process mixtures are complex, specialized detectors may substitute for separation efficiency. One specialized detector is the array amperometric detector, which allows selective detection of electrochemically active compounds.23 Electrochemical array detectors are discussed in greater detail in Chapter 5. Many pharmaceutical compounds are chiral, so a detector capable of determining optical purity would be extremely useful in monitoring synthetic reactions. A double-beam circular dichroism detector using a laser as the source was used for the selective detection of chiral cobalt compounds.24 The double-beam, single-source construction reduces the limitations of flicker noise. Chemiluminescence of an ozonized mixture was used as the principle for a sulfur-selective detector used to analyze pesticides, proteins, and blood thiols from rat plasma.25 Chemiluminescence using bis (2,4, 6-trichlorophenyl) oxalate was used for the selective detection of catalytically reduced nitrated polycyclic aromatic hydrocarbons from diesel exhaust.26... [Pg.93]

However, the disadvantages of optical rotation detectors may be limited by shot or flicker noise, which are dependent on the optical and mechanical properties of the system or by noise in the detector electronics. Generally, the usefulness of this technique has been limited by the lack of sensitivity of commercially available instruments. [Pg.1075]

Here, up represents the noise of the photoelectrons. When the photon flux is n, Up x VW up, is the dark current noise of the photomultiplier and is proportional to the dark current itself, up is the flicker noise of the source and is proportional to the signal and uA is the amplifier noise resulting from electronic components. The last contribution can usually be neglected, whereas up is low for very stable sources (e.g., glow discharges) or can be compensated for by simultaneous line and background measurements. As up, x Ip, one should use detectors with low dark current, then the photon noise of the source limits the power of detection. [Pg.47]

As the sample introduction into the plasma is very stable, RSDs in ICP-AES are about 1% and the limiting noise is proportional to this, including flicker noise and nebulizer noise. [Pg.224]

Whereas the plasma background signal (except at high levels) is predominantly shot noise limited, the net analyte signal is dominated by flicker noise (except at the detection limit region) as demonstrated by the signal to noise relationship data in Tables V and VI. Source flicker noise dependence of the net analyte intensity is a multiparameter phenomenon which may stem from 120 Hz power beats of the plasma, variations in gas flow rates or variations in the rate of nebulization. [Pg.101]

The multichannel nature of the SPD could be used to reduce the effect of flicker noise by utilizing the classic internal reference technique. The intensities of a few carefully selected lines are simultaneously integrated with the analyte lines and used as introduction-evaporization-atomization-excitation indicators. This approach can significantly reduce flicker noise and thus improve the S/N performance of the system as shown in Table VII. As expected, at low concentration levels, where the system becomes RFI noise limited, this method offers a very limited S/N improvement. [Pg.101]

For optimum noise performance the bandwidth of a sensor and its evaluation electronics should be limited to a value no larger than that required by the application. For example, an accelerometer with a noise floor of 1 mG (1 G = 9.8 m/s2) in a 1 kHz bandwidth will resolve 500 pG when its bandwidth is reduced to 250 Hz. Further bandwidth reduction results in additional improvement up to the point where a different noise source dominates (e.g., flicker noise). [Pg.251]

One source of noise of ihls type is the slow drift in the radiant output of the source. This type of noise can be called source flicker noise (Section 5B 2). The effects of nuctuations in the intensity of a source can be minimized by the use of a constani-voUagc power supply or a feedback system in which the source intensity is maintained at a constant level. Modern double-beam spectrophotometers (Sections I3D-2and I3D-3) can also help cancel the effect of flicker noise. With many instruments, source flicker noise does not limit performance. [Pg.346]

All of the flicker noises can be effectively eliminated by the use of double-beam optics in conjunction with a background correction system such as Zeeman splitting or a well-aligned (or wavelength-modulated) continuum source. Thus the ultimate limiting noise in atomic absorption is source shot noise, which can be reduced (relative to total source intensity or I, ) by increasing the source intensity, up to the point of optical saturation. [Pg.119]

If a drifting baseline is present and possibly corrected, the formulae become rather complicated and are not directly usable in daily practice. An extension to other kinds of noise, i.e. the more realistic 1/f or flicker noise leads to even more complicated formulae. Nevertheless, it is possible to determine quantitatively detection limits in case of the application of some specific baseline drift correction procedure, if the measurement conditions are well defined and stable and good models of the noise and drift are known from an a priori analysis. Moreover, a better insight in several effects influencing the remaining uncertainty after drift correction is obtained. [Pg.148]

The possibility of modulating the emitted wavelength of DLs at GHz frequencies by the modulation of the diode current allows one to reduce the low-frequency (flicker) noise in the baseline, which again improves the detection limit. Emission wavelength modulation also permits to correct for unspecific background absorption, thereby improving the selectivity of the technique. [Pg.440]

Because DNA translocation events have main power spectral densities with a bandwidth of 10 kHz and R is also negligible compared to Rj, is approximated to Ri / 1 + ]2%f-Cj Rj ). As a result, at low frequencies, can be simplified to AkT/Rj, which is proportional to the conductance of the nanopore. Nanopore flicker noise is given by (a x F)/ (Nc xf). Here, I, a, and are the direct current, the Hooge parameter, and the number of charge carriers [18], respectively, all of which are associated with the KCl buffer concentration and nanopore material. To increase Rj, a nanopore with a narrow diameter should be adopted. For instance, the a-hemolysin pore with a limiting diameter of 1.5 nm has a resistance of 3 GO, in 0.3 M KCl or 1 GQ in 1 M KCl. [Pg.625]


See other pages where Flicker noise limited is mentioned: [Pg.99]    [Pg.467]    [Pg.113]    [Pg.115]    [Pg.99]    [Pg.467]    [Pg.113]    [Pg.115]    [Pg.239]    [Pg.135]    [Pg.43]    [Pg.112]    [Pg.135]    [Pg.135]    [Pg.60]    [Pg.246]    [Pg.300]    [Pg.101]    [Pg.251]    [Pg.206]    [Pg.266]    [Pg.20]    [Pg.115]    [Pg.118]    [Pg.118]    [Pg.119]    [Pg.119]    [Pg.120]    [Pg.121]    [Pg.123]    [Pg.224]    [Pg.630]    [Pg.246]    [Pg.300]   
See also in sourсe #XX -- [ Pg.113 ]




SEARCH



Flicker

Flicker/flickering

Limiting noises

Noise limits

© 2024 chempedia.info