Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Field frequency

Other types of mass spectrometer can use point, array, or both types of ion detection. Ion trap mass spectrometers can detect ions sequentially or simultaneously and in some cases, as with ion cyclotron resonance (ICR), may not use a formal electron multiplier type of ion collector at all the ions can be detected by their different electric field frequencies in flight. [Pg.212]

In the inset of Fig. 9 we show the mean field frequency 0 = 0// as a function of density for T = 1. At this temperature the system undergoes a phase transition from a paramagnetic to a ferromagnetic fluid at a density whose mean field value is p mf = 0-4- For densities below this value we obtain 0 = cjq, which agrees with the frequency value of the low-order virial expansion (see Eq. (34)). For p > Pc,mF) increases with the density due to increase of the magnetization. [Pg.104]

Besides the deviation mentioned above, the main problem with the dynamical information from the MF approximation is that it contains only one positive frequency and so the resulting real-time correlations cannot be damped or describe localizations on one side of the double well due to interference effects, as one expects for real materials. Thus we expect that the frequency distribution is not singly peaked but has a broad distribution, perhaps with several maxima instead of a single peak at an average mean field frequency. In order to study the shape of the frequency distribution we analyze the imaginary-time correlations in more detail. [Pg.104]

Cyclitol Spectra at 220 MHz with the Superconducting Solenoid. In 1964, Nelson and Weaver (34) at Varian Associates constructed a superconducting solenoid with which proton spectra can be observed at 51.7 kilogauss (220 MHz.) or even higher fields. Other nuclei have been observed at suitable field/frequency combinations. [Pg.57]

Fast Fourier transform (FFT) A procedure for carrying out Fourier transformation at high speed, with a minimum of storage space being used. Field frequency lock The magnitude of the field Bq is stabilized by locking onto the fixed frequency of a resonance in the solution, usually of the solvent. [Pg.414]

Finally, the intrinsic features of the ultrasonic field, frequency and power, should also be taken into account in the design of the experimental arrangement. It is obvious that the mechanical and chemical effects derived from a low frequency field are quite different than those provided by high frequency fields, and these features should match with features of the electrode materials such dimension, structure and physical and chemical properties [30]. [Pg.109]

Figure 9.8 Second-harmonic generation from thin film (or surface). Fundamental beam at frequency > i p or. v polarized) is incident on film at angle 0. The p- and. s-polarized second-harmonic fields (frequency 2 co) are generated in reflection and transmission. [Pg.533]

As mentioned above, in an ENDOR experiment the rf field is swept while the static magnetic field is held at a constant position in the EPR spectrum. For slow sweep rates and narrow EPR lines a device would be desirable which is able to stabilize the ratio of the microwave frequency to the static magnetic field. The applicaiton of a commercially available field/frequency lock system is restricted to a region of 6 mT about the DPPH resonance field33). In metal complexes with strongly anisotropic EPR spectra, however,... [Pg.7]

Let us now turn our attention to the application of the sound wave to a liquid since this is the medium of importance to the practising chemist. The sound wave is usually introduced to the medium by either an ultrasonic bath or an ultrasonic horn (see Chapter 7). In either case, an alternating electrical field (generally in the range 20-50 kHz) produces a mechanical vibration in a transducer, which in turn causes vibration of the probe (or bottom of the bath) at the applied electric field frequency. The horn (or bath bottom) then acts in a similar manner to one prong of a tuning fork. As in the case of air, the molecules of the liquid, under the action of the applied acoustic field, will vibrate about their mean position and an acoustic pressure (P = P sin 2k ft) will be superimposed upon the already ambient pressure (usually hydrostatic, Pjj) present in the liquid. The total pressure, P, in the liquid at any time, t, is given by Eq. 2.4. [Pg.30]

A disadvantage of the LR-CPMG detection method is its total insensitivity to field/frequency offset which must be adjusted before a profile measurement and cannot be corrected by means of a simple procedure during an automatic profile measurement. This requires a higher degree of longterm field stability (including any thermal effects) than the other methods. Despite the insensitivity of the technique, in fact, the field may not be allowed to drift too far from resonance where the RF pulses would lose their efficiency (excursions up to about 5 kHz are, however, quite tolerable). [Pg.459]

The proton noise-decoupled 13c-nmr spectra were obtained on a Bruker WH-90 Fourier transform spectrometer operating at 22.63 MHz. The other spectrometer systems used were a Bruker Model HFX-90 and a Varian XL-100. Tetramethylsilane (TMS) was used as internal reference, and all chemical shifts are reported downfield from TMS. Field-frequency stabilization was maintained by deuterium lock on external or internal perdeuterated nitromethane. Quantitative spectral intensities were obtained by gated decoupling and a pulse delay of 10 seconds. Accumulation of 1000 pulses with phase alternating pulse sequence was generally used. For "relative" spectral intensities no pulse delay was used, and accumulation of 200 pulses was found to give adequate signal-to-noise ratios for quantitative data collection. [Pg.237]

As already introduced in section I of this chapter, in a CARS process (Figures 7.9a-c see also Figure 7.1c), a Raman transition between two vibrational energy levels of a molecule is coherently driven by two optical laser fields (frequencies co and co) and subsequently probed by interaction with a third field at frequency co, . This generates the anti-Stokes signal at the blue-shifted frequency cars = p- The... [Pg.179]

For quantitative estimation, a sealed reusable capillary tube, with a known quantity of sodium salt of trimethylsilyl propionic acid (TSP) dissolved in 35 pi of D20, is inserted into the NMR tube while obtaining NMR spectra. The internal standard TSP is used as a chemical shift reference as well as a quantitative standard for the estimation of metabolites, and D20 is used as the field-frequency-lock . Spectra are acquired at room temperature. Typical spectra acquired at room temperature of human bile and standard glycine- and taurine-conjugated BAs are shown in Fig. 5.4.16. [Pg.653]


See other pages where Field frequency is mentioned: [Pg.1439]    [Pg.2988]    [Pg.2988]    [Pg.195]    [Pg.205]    [Pg.105]    [Pg.114]    [Pg.271]    [Pg.188]    [Pg.161]    [Pg.124]    [Pg.278]    [Pg.73]    [Pg.55]    [Pg.55]    [Pg.55]    [Pg.319]    [Pg.189]    [Pg.398]    [Pg.8]    [Pg.113]    [Pg.84]    [Pg.347]    [Pg.394]    [Pg.67]    [Pg.153]    [Pg.154]    [Pg.118]    [Pg.144]    [Pg.182]    [Pg.224]    [Pg.163]    [Pg.213]    [Pg.217]    [Pg.569]    [Pg.569]    [Pg.572]    [Pg.286]   
See also in sourсe #XX -- [ Pg.135 ]

See also in sourсe #XX -- [ Pg.359 , Pg.360 ]




SEARCH



© 2024 chempedia.info