Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Electronic Standards for Transmission

M2—Electronic Standards for Transmission of Regulatory Information (ESTRI)... [Pg.223]

Electronic Standards for Transmission of Regulatory Information European Union antigen-binding fragment fluorescence-activated cell sorter fetal bovine serum constant fragment... [Pg.437]

EGE (epidermal growth factor) 238, 463 EGER (epidermal growth factor receptor) 59, 105, 623 Elastase 380 Elcatonin 476 Elcitonin Inj. 20S 476 Electromechanical (NOGA) mapping 155 Electron spray ionization (ESI) 1564 Electronic standards for transmission of regulatory information (ESTRI) 1699 Electrophoretic methods... [Pg.1853]

When a laboratory reports data from anal5d ical instruments electronically to the EPA, those data must be submitted on standard magnetic media—tapes or diskettes—and conform to all requirements of EPA order 2180.2, Data Standards for Electronic Transmission of Laboratory Measurement Results. ... [Pg.151]

The particles continue to fly into the sampling chamber through an orifice between the reaction and the sampling chamber. The pressure of the sampling chamber is 9.5 X 10-4 torr. The particles are collected in a form that is convenient for characterization or application. For example, the particles are collected on a microgrid for transmission electron microscope (TEM) observation and on a polyimid-film for MOssbauer and x-ray diffraction studies. A standard passivation treatment, namely, slow introduction of O2 gas followed by the introduction of dry air to the chamber, is made. [Pg.121]

Ejected electron analyzers can be calibrated at lower energies (<25 eV) using UV photoelectron spectroscopy and comparison with quantitative photoelectron spectra. The intensity ratios provide a relative transmission function (7 ) directly. Quantitative (relative) photoelectron spectra have been reported by Hotop and Niehaus79 at an ejection angle of 90°, and these results have been used by Yee et al.66 to calibrate a 127° analyzer for which the correction curve has already been shown in Fig. 3. More recently Gardner and Samson80 reported quantitative (relative) photoelectron spectra that can be used as a standard for analyzer... [Pg.30]

Ferromagnetic resonance (PMR) spectra of cobalt mcial were recorded at X-band frequencies on a Varian E-4 spectrometer. DPPH and weak pitch were used as standards for g factor. Transmission electron microscopy (TEM) was done with JEOL 200CX microscope using 160 KeV electrons. [Pg.222]

The CIC connectivity standards are represented simply as the two interfaces between the POCT devices and information systems (Figure 12-11). The device interface passes patient results and QC information between the POCT instrument and devices, such as docking stations, concentrators, terminal servers, and point-of-care data managers. The latter have to be linked to a variety of information systems via the observation reporting interface or electronic data interface, for transmission of ordering information and patient results. [Pg.309]

X-ray photoelectron spectra were recorded with a VG Escalab 200R electron spectrometer equipped with a Mg Ka X-ray source (hv = 1253.6 eV) and a hemispherical electron analyzer operating at constant transmission energy (50 eV). No reduction treatment of ICP catalysts was carried out before XPS measurements. The C Is peak at 284.6 eV was used as an internal standard for peak position measurement. [Pg.161]

Because of the low penetrating power of electrons, specimens for examination in transmission have to be very thin, depending on the circumstances tens or hundreds of nanometres in thickness. Surface features on thick specimens can be examined by making a thin replica. A standard technique is to evaporate a thin film of carbon on to the surface of interest, to shadow it by depositing an even thinner layer of platinum at a different angle, and then to remove the deposited replica, for example, by dissolution of the original specimen. [Pg.129]

Hettinger, B. J., BrazUe, R. P., Health Level Seven (HL7) standard for healthcare electronic data transmissions. Comput. Nurs. 12(1) 13-16 (1994)... [Pg.353]

The technique of transmission spectroelectrochemistry, using an optically transparent electrode (OTE), was first demonstrated in 1964 using o-tolidine, a colourless compound that reversibly undergoes a 2-electron oxidation in acidic solution to form an intensely yellow coloured species (Eqn [1]). This system soon became a standard for testing spectro-electrochemical cells and new techniques. [Pg.995]

The effect of device movement and vibration must be considered in the design specification. The acceptable sound levels at various frequencies must be specified. A device must meet existing standards for electromagnetic interference and susceptibility. The location and size of system components other than the pump, such as the electronics module and energy transmission coils, must be carefully chosen. If wound coils are used for energy transmission, movement can cause a loss of coupling. [Pg.1514]


See other pages where Electronic Standards for Transmission is mentioned: [Pg.70]    [Pg.1699]    [Pg.70]    [Pg.1699]    [Pg.517]    [Pg.536]    [Pg.850]    [Pg.89]    [Pg.563]    [Pg.106]    [Pg.13]    [Pg.153]    [Pg.314]    [Pg.357]    [Pg.73]    [Pg.38]    [Pg.292]    [Pg.329]    [Pg.624]    [Pg.34]    [Pg.385]    [Pg.158]    [Pg.778]    [Pg.369]    [Pg.188]    [Pg.7]    [Pg.344]    [Pg.165]    [Pg.112]    [Pg.49]    [Pg.358]    [Pg.324]    [Pg.1058]    [Pg.47]   


SEARCH



Electron standards

Electronic standards

© 2024 chempedia.info