The threshold values for the BPIX modules were programmed to the target values determined in the module testing. Afterwards the threshold and the noise were measured using the S-Curve method. An average threshold of 3,829 electrons was found, well above the noise level at 141 electrons. [Pg.132]

The results of the analog pulse height calibration are given in Fig. 8.19. The gain and pedestals are obtained from a linear fit. The tail in the pedestal distribution is due to mlsconfiguration of individual ROCs which was recovered at a later stage of detector calibration. [Pg.132]