Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Counter-indication level

If smoke is introduced with air pressure, adjust it to provide a smoke outlet velocity equal to the room air velocity at that point. Operate the particle counter with the sample tube at the normal work level and at a point remote from the smoke source. Verify that the counter indicates particle concentrations less than 200 particles of 0.5 ftm or greater. [Pg.183]

The dynamic nature of chemical equilibrium can be illustrated by placing equal masses of iodine crystals in two interconnected flasks, as shown in Figure 17.6a. The flask on the left contain iodine molecules made up entirely of the nonradioactive isotope 1-127. The flask on the right contain iodine molecules made up of the radioactive isotope 1-131. The radiation counters indicate the difference in the levels of radioactivity within each flask. [Pg.598]

You enter the control room and observe that the fission counter indicates a steady neutron level with no rods in motion. Which ONE condition below CANNOT be true ... [Pg.312]

Formation of the t2g band with a binding energy of 0.8 eV below the Fermi level in Fig. 21 indicates a change in Ru coordination for potentials above 0.4 V. The increased O/Ru ratio in Fig. 20 is therefore not only a consequence of enhanced counter ion adsorption, but rather a result of oxide/hydroxide formation. [Pg.103]

In halogenated solvents the results indicate that return can occur, even for the uncongested stilbenes. Unfortunately, its importance, as measured by the k i/kN ratio (Fig. 10), cannot be estimated. It must be noted that Bellucci s experiments prove only that return is possible, but do not demonstrate conclusively that it occurs in bromination, since reversibility is controlled by the relative energy levels of TS and TSN which can be affected by the reaction conditions. Now, these conditions are not the same for nucleophilic substitution on bromohydrins and for bromine addition in particular, the counter-ions, Br and Br3 respectively, can alter the lifetime of the intermediate and thus control its partitioning between return and nucleophilic attack. [Pg.284]

Some samplers incorporate an accumulated volume indicator such as a stroke-counter to permit correction for the above effects. However, time-of-day bias can occur if the exposure is not constant. A high contaminant level would be erroneously indicated if a high exposure occurred during the start of the sampling period rather than at the end. [Pg.492]

An expensive method is the use of nuclear radiation to obtain information on the level in an apparatus. The nuclear sensor is mounted at one side, and at the other side a scintillation counter is fixed near the surface of the apparatus. Both systems are sheathed with lead-screen shields to give protection from nuclear radiation. A continous level indicator using nuclear radiation is very complicated and therfeore seldom applied. [Pg.241]

The two outputs being separated by 90° are called quadrature outputs. A counter counts up all of the increment and decrement outputs to arrive at the total count that represents the present position. There are four states of the combination of A and B levels per the 360° of one cycle. The phase relationship between outputs A and B during a state change indicates the direction of motion (increment or decrement of the count). [Pg.491]

The distribution of chromium(III) in humans was analyzed using a whole-body scintillation scanner, whole-body counter, and plasma counting. Six individuals given an intravenous injection of 51chromium(III) as chromium trichloride had >50% of the blood plasma chromium(ni) distributed to various body organs within hours of administration. The liver and spleen contained the highest levels. After 3 months, the liver contained half of the total body burden of chromium. The study results indicated a three-compartment model for whole-body accumulation and clearance of chromium(III). The half-lives were 0.5-12 hours for the fast component, 1-14 days for the medium component, and 3-12 months for the slow component (Lim et al. 1983). [Pg.168]

Active/passive device Active devices require input of power, most often low-level (5-24 V) DC, to achieve their specified function, with the ctmse-quence that their output RF power level can exceed RF input powo-. Passive devices, on the other hand, effect some transformation of the input signal without use of any external power source, so that the output power is always less than or equal to the input power. In the following, simple components are specified as active or passive (note that the addition of an electronic control system, e.g., a motor drive to set the value of a variable attenuator, is not considered grounds for calling a component active). The active/passive distinction is made only for circuit components having an input and output, not for measurement instrumentation (e.g., a frequency counter), the output of which is a visual indication or computer bus-compatible signal. [Pg.355]


See other pages where Counter-indication level is mentioned: [Pg.134]    [Pg.147]    [Pg.2278]    [Pg.309]    [Pg.314]    [Pg.349]    [Pg.350]    [Pg.200]    [Pg.539]    [Pg.75]    [Pg.353]    [Pg.249]    [Pg.571]    [Pg.9]    [Pg.30]    [Pg.86]    [Pg.340]    [Pg.340]    [Pg.111]    [Pg.145]    [Pg.17]    [Pg.228]    [Pg.181]    [Pg.13]    [Pg.747]    [Pg.992]    [Pg.213]    [Pg.556]    [Pg.34]    [Pg.127]    [Pg.72]    [Pg.182]    [Pg.452]    [Pg.284]    [Pg.210]    [Pg.181]    [Pg.123]    [Pg.304]    [Pg.255]   
See also in sourсe #XX -- [ Pg.314 ]




SEARCH



Level indicators

© 2024 chempedia.info