Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Constant matrix effect

Fig. 8. Exemplary data and graphic of a calibration - a constant matrix effect is evident and we include a blank sample for clarity. Fig. 8. Exemplary data and graphic of a calibration - a constant matrix effect is evident and we include a blank sample for clarity.
An important aspect of quenching in analysis is that the fluorescence exhibited by the analyte may be quenched by the molecules of some compound present in the sample, i.e. this is an example of a matrix effect. If the concentration of the quenching species is constant this may be allowed for by using suitable standards (i.e. containing the same concentration of quenching species), but difficulties occur when there is an unpredictable variation in the concentration of quenching species. [Pg.733]

The constant matrix in Group III is conveniently defined as all the sample except the elements being determined, but a word of caution is necessary. If these elements are present at appreciable concentration, then absorption and enhancement effects to which they give rise must of course be considered for, as was pointed out in Section 7.3, these effects do not reside in the matrix alone. [Pg.202]

The issue of selectivity is one that is often difficult to address. Initial method development is invariably carried out by using standards made up with pure solvents, i.e. free from any matrix effects. It is often only when real samples are analysed that the true extent of interference becomes apparent and the value of the method can be properly assessed. An added complication is that interferences , by their very nature, are not constant and a number of samples may have a combination of interferences that defy analysis by a method that is otherwise successful on a routine basis (another example of Murphy s law ). [Pg.269]

The gradual accumulation of structures III in the initial stages of the reaction could account for its auto-catalytic character until a "stationary-state" is reached in which the surface of the precipitated polymer available for such association with monomer remains practically constant. A structure such as III could also be responsible for the stereo-control of the propagation step. Indeed, the polymer formed at conversions below 2 per cent exhibits much lower stereoregularity than at higher conversions (see Figure 6). These various peculiar consequences connected with the arizing of structure III will be referred to as "matrix effect". [Pg.239]

The fact that X for both salts lies in the range 0.20-0.25 shows that water in the membrane is a less effective solvent for ions than is bulk water i.e. the low-dielectric-constant matrix polymer lies well within the range of the electrostatic fields around the ions. Our value of Xg, the molar distribution coefficient of sodium chloride between polymer and solution, is in good agreement with values obtained by direct measurement (1,5,10, 11,12). This is further evidence in favour of our theories and assumptions. [Pg.110]

Whereas TIMS suffers from the difficulties of time dependent mass fractionation in MC-ICP-MS, the mass fractionation is relatively constant. Its magnitude is often greater than that of TIMS and matrix effects are possible84. It is known that isotopic reference materials used for the calibration of isotope ratio measurements are available only as a pure metal or compound. Possible matrix effects during isotope ratio measurements, especially in solid-state... [Pg.230]

Instruments that have burners and require nebulisation of dilute aqueous sample solutions generally have low background noise in the signal. With graphite furnaces, incomplete atomisation of the solid sample at elevated temperatures can produce interfering absorptions. This matrix effect does not exist in an isolated state and thus cannot be eliminated by comparison with a reference beam. This is notably the case for solutions containing particles in suspension, ions that cannot be readily reduced and organic molecules, all of which create a constant absorbance in the interval covered by the monochromator. [Pg.264]

There are two major reasons why a traceability chain may be broken and trueness lost due to the introduction of bias insufficient commutability of a calibration material and non-specificity of a measurement procedure. The effect of these separate properties are often indiscriminately lumped together as matrix effect . Commutability refers to the ability of a material, here a calibrator, to show the same relationships between results from a set of procedures as given by routine samples [16, 17]. Analytical specificity refers to the ability of a measurement procedure to measure solely that quantity which it purports to examine [16, 18]. Discrepancies between results of a reference procedure and a routine procedure applied to routine samples are often caused by non-specificity of the routine procedure. The use of a set of human samples as a manufacturer s calibrator to eliminate so-called matrix effects should only be accepted if the relationship between the results from reference and routine procedures is sufficiently constant to allow explicit correction with consequent increased uncertainty of assigned values. [Pg.52]

Nowadays a wide variety of quantum-chemical programs are disposable, which permit to calculate with high accuracy the equilibrium geometry of the molecules and their energy of formation. Theoretical methods have been developed for analytical calculation of the first and second derivatives of energy [8,9], so that the force-constant matrix FHT and the harmonic frequencies can be extracted from the quantum-mechanical calculations. Since as a rule the molecular orbitals (MO) obtained by the quantum-mechanical methods are spread around the entire molecule, the corresponding quantum-mechanical force fields incorporate the important effects of the off-diagonal interactions. [Pg.342]

The ICP was a Perkin-Elmer 3000DV with an AS90 Autosampler, which has an instrument detection limit of about 1 ppb (for most elements) with a linear calibration up to 100 ppm (for most elements). Solid samples were prepared via microwave digestion in concentrated nitric and hydrochloric acids, then diluted to volume. The ICP was calibrated and verified with two independent, certified standard sets. Spikes and dilutions were done for each batch of samples to check for and/or mitigate any matrix effects. The ICP process ran a constant pump rate of 1.5 mL/min for all samples and standards during analysis. A 3 mL/min rinse and initial sample flush were used to switch between each sample and standard. The plasma was run at 1450 W with argon flow. Trace metal-grade (sub-ppb) acids and two independently NIST-certified calibration standard sets were used for calibration and method verification. [Pg.809]

Despite the advantages of an equilibrium, nonexhaustive extraction procedure, there are also disadvantages. Matrix effects can be a major disadvantage of a sample preparation method that is based on equilibration rather than exhaustive extraction [134], Changes in the sample matrix may affect quantitative results, due to alteration of the value of the distribution constant relative to that obtained in a pure aqueous sample [68,134],... [Pg.116]

The advantage to MHE is that sample matrix effects (which are mainly an issue only with solid samples) are eliminated since the entire amount of analyte is examined. This examination is done by performing consecutive analyses on the same sample vial. With the removal of each sample aliquot from the vial, the partition coefficient K will remain constant however, the total amount of analyte remaining in the sample will decline as each analysis is performed and more of the analyte is driven up into the vial headspace for removal and analysis. Chromatograms of each injection of sample show... [Pg.193]

When matrix effect exists, it is usually preferable to coeluate the analyte and its internal standard to better reduce the impact of matrix effect on quantitation. The more their chromatographic peaks overlap, the better the correction is. Since the concentration of the analyte varies while the amount of IS added is constant, a choice must be made as to match which part of a calibration curve. Usually, the segment between 1/3 and 1/2 of the ULOQ is most important because this segment is expected to cover the average Cma for most drugs and metabolites. This is probably why other researchers have proposed to use IS concentrations around 1/3 or 1/2 of the ULOQ of an analyte. [Pg.7]

Matrix effects can be complex and difficult to predict because most co-solutes may compete with the reactant of interest for reactions with radicals, becoming effective scavengers that reduce the sonochemical efficiency. Taylor et al. [30] have observed a significant inhibition of the sonolysis of polycyclic aromatic hydrocarbons (PAHs) in the presence of dissolved organic matter. Substrate concentration effects on the rate constants have also been reported. When the target molecules are volatile, they partition between... [Pg.216]

Approaches have been described for using standard additions without diluting to a constant volume. Multiplying measured responses by a ratio of total to initial volume accomplishes a correction for dilution. However, the matrix concentration is also diluted by the additions, creating a nonlinear matrix effect that may or may not be transformed into a linear effect by the volume correction. Kalivas [83] demonstrated the critical importance of maintaining constant volume. [Pg.155]


See other pages where Constant matrix effect is mentioned: [Pg.296]    [Pg.116]    [Pg.700]    [Pg.374]    [Pg.833]    [Pg.39]    [Pg.107]    [Pg.701]    [Pg.293]    [Pg.340]    [Pg.118]    [Pg.158]    [Pg.414]    [Pg.42]    [Pg.494]    [Pg.411]    [Pg.102]    [Pg.213]    [Pg.204]    [Pg.69]    [Pg.391]    [Pg.307]    [Pg.373]    [Pg.39]    [Pg.40]    [Pg.163]    [Pg.297]    [Pg.156]    [Pg.36]    [Pg.73]    [Pg.320]    [Pg.41]   
See also in sourсe #XX -- [ Pg.134 ]




SEARCH



Matrix effects

© 2024 chempedia.info